Cdc-25122-xcn1a Arrowg+ard Blue Two-lumen Hemodialysis Catheterization For High Volume Infusions Instructions

CDC-25122-XCN1A Arrowg+ard Blue Two-Lumen Hemodialysis Catheterization for High Volume Infusions

Product Information

  • Product Name: CDC-25122-XCN1A
  • Size: 12 2 Fr.
  • Lumen: Single lumen
  • Catheter Length: 20cm
  • Spring-Wire Guide Diameter: 0.035 inch

Contents

  • Catheter: 18 Ga. x 2-1/2 (6.35 cm) Radiopaque over 20 Ga. RW
    Introducer Needle
  • Raulerson Spring-Wire Introduction Syringe
  • Injection Needle: SafetyGlideTM1 22 Ga. x 1-1/2 (3.81 cm) and 5
    mL Luer-Slip
  • Hi-Lite OrangeTM Tint

Note: SafetyGlideTM1 is a trademark of Becton, Dickinson and
Company. Hi-Lite OrangeTM is a registered trademark of CareFusion
or one of its subsidiaries.

Contraindications

The Arrow large-bore two-lumen catheter is not designed for
long-term (30 days) hemodialysis or for use in patients with
thrombosed vessels. The Arrowg+ard Blue antimicrobial catheter is
contraindicated for patients with known hypersensitivity to
chlorhexidine, silver sulfadiazine, and/or sulfa drugs.

Note: Arrowg+ard Blue is a registered trademark of 3M
Company.

Priming Volume

DistalProximal
Priming Volume*1.60 mL1.50 mL

* Priming volumes are approximate and are done without
accessories.

Pressure and Flow Rate

The following graph illustrates the pressure and flow rate
relationship:

Pressure and Flow Rate Graph

Note: The graph shows the flow rate in mL/min and the
pressure in mmHg.

LBL068965 R00 (2023-04)

CDC-25122-XCN1A

12 2 Fr.

Lumen

20cm catheter length

.035 inch dia. spring-wire guide

Arrowg+ard Blue® Two-Lumen Hemodialysis Catheterization for High Volume Infusions

Contents:
1: Two-Lumen Catheter: 12 Fr. (4.06 mm OD) x 20 cm Radiopaque Polyurethane with Blue FlexTip®, Arrowg+ard® Antimicrobial Surface Treatment
1: Spring-Wire Guide, Marked: .035″ (0.89 mm) dia. x 23-5/8″ (60 cm) (Straight Soft Tip on One End – “J” Tip on Other) with Arrow® GlideWheelTM Wire Advancer
1: Catheter: 18 Ga. x 2-1/2″ (6.35 cm) Radiopaque over 20 Ga. RW Introducer Needle
1: Injection Needle: SafetyGlideTM1 25 Ga. x 1″ (2.54 cm) 1: Blunt Fill Needle: 18 Ga. x 1-1/2″ (3.81 cm) 1: Pressure Transduction Probe 1: Introducer Needle: Echogenic 18 Ga. x 2-1/2″ (6.35 cm) XTW and 5 mL Arrow®
Raulerson Spring-Wire Introduction Syringe 1: Injection Needle: SafetyGlideTM1 22 Ga. x 1-1/2″ (3.81 cm) and 5 mL Luer-Slip
Syringe 1: Syringe: 3 mL Luer-Lock 1: Tissue Dilator: 10 Fr. (3.5 mm) x 10.2 cm 1: Tissue Dilator: 12 Fr. (4.0 mm) x 14 cm 2: 10 mL Luer-Lock Pre-Filled Saline Syringe 1: 5 mL 1% Lidocaine HCl Solution and Alcohol Prep 2: 3 mL Applicator 2% CHG and 70% IPA ChloraPrep®2 One-Step Solution with
Hi-Lite OrangeTM Tint

1: SharpsAway® II Locking Disposal Cup 1: SharpsAway® Disposal Cup 1: Maximal Barrier DrapeTM with 4″ fenestration 1: Towel 1: Needle Holder 1: Safety Scalpel: #11 1: Checklist/CLIP Sheet 1: Sterile Procedure Sign 1: Medication Label: 1% Lidocaine 2: Gauze Pad: 2″ x 2″ (5 cm x 5 cm) 5: Gauze Pad: 4″ x 4″ (10 cm x 10 cm) 1: Surgical Apparel: Impervious Gown 1: Dressing: Tegaderm®3 CHG 3-1/2″ x 4-1/2″ (8.5 cm x 11.5 cm) 1: Transducer Cover: 14 cm x 147 cm 1: Surgical Apparel: Mask with Eye Shield 1: Surgical Apparel: Bouffant Cap 1: Suture: 3-0 Silk with Curved Needle 1: Tubing±: 20″ (50 cm) Extension 1: HemoHopper® Fluid Receptacle
1A trademark of Becton, Dickinson and Company. 2A registered trademark of CareFusion or one of its subsidiaries.

±Precaution: Contains Phthalate: DEHP. Warning: The results of certain animal experiments have shown phthalates to be potentially toxic to reproduction. Proceeding from the present state of scientific knowledge, risks for male premature infants cannot be excluded in the case of long-term exposure or application. Medical products containing phthalates should be used only temporarily with pregnant women, nursing mothers, babies and infants. Not made with natural rubber latex. Store between 20 ­ 25°C (68 ­ 77°F). Arrow International provides the enclosed medication label(s) for your convenience. Please ensure that these labels are applied to the correct syringe and corresponding medication.
Contraindications: The Arrow large-bore two-lumen catheter is not designed for long-term ( 30 days) hemodialysis or for use in patients with thrombosed vessels. The Arrowg+ard Blue antimicrobial catheter is contraindicated for patients with known hypersensitivity to chlorhexidine, silver sulfadiazine, and/ or sulfa drugs.

3A registered trademark of 3M Company.

Priming Volume*

Distal Proximal

1.60 mL 1.50 mL

* Priming volumes are approximate and are done without accessories.

Pressure (mmHg)

Dialysis Flow Rate 300

200

100 0
-100

VENOUS ARTERIAL

-200

-300

100

200

300

400

500

Flow (mL/min)

In vitro tests performed using 3:1 glycerine/saline at 37°C.