CV-15142-UF You-Bend Two-Lumen Hemodialysis Catheter for High-Volume Infusions
Instructions
CV-15142-UF You-Bend Two-Lumen Hemodialysis Catheter for High-Volume Infusions
| 14 Fr. | 2 Lumen | 20 cm catheter length | .038 inch dia. spring-wire guide | |
Contents:
- 1: Two-Lumen Catheter‡: 14 Fr. (4.72 mm OD) x 20 cm Radiopaque Polyurethane with Blue FlexTip®, Extension Line Clamps
- 1: Spring-Wire Guide, Marked: .038″ (0.97 mm) dia. x 27-1/2″ (70 cm) (Straight Soft Tip on One End – “J” Tip on Other) with Arrow Advancer
- 1: Introducer Needle: 18 Ga. x 2-1/2″ (6.35 cm) XTW 1 : Syringe: 5 mL Luer-Slip
- 1: Tissue Dilator: 14 Fr. (4.7 mm) x 14 cm
- 2: Dust Cap: Non-Vented All components are CE 2797 unless otherwise noted.
All components are CE 2797 unless otherwise noted.
Warning: Read all package insert warnings, precautions, and instructions prior to use. Failure to do so may result in severe patient injury or death. www.teleflex.com/IFU
‡MR Conditional
Not made with natural rubber latex.
Fluid path components are non-pyrogenic.
Contraindications: The large-bore two-lumen catheter is not designed for long-term (≥ 30 days) hemodialysis or for use in patients with thrombosed vessels.
| Priming Volume* (mL) | Flow Rate (mL/min) | Venous Pressure Setting†† (mmHg) | |
| Distal ( 10 G a .) | 2.1 | 250 400 | 100 160 |
| Proximal ( 10 G a .) | 2.0 | 250 400 | 100 180 |
* Priming volumes are approximate and are done without accessories.
†† Blood flow rates are approximate and were determined by varying the venous pressure setting from 100 to 200 mmHg in an in vitro experiment using a blood analog solution.
EU Authorized Representative and Importer:
Teleflex Medical IDA Business and
Technology Park Dublin Road, Athlone Co. Westmeath, Ireland
Arrow International LLC
A subsidiary of Teleflex Incorporated
3015 Carrington Mill Blvd.
Morrisville, NC 27560 USA
www.teleflex.com
LBL057827 R00 (2021-05)![]()
![]()