Teleflex Ca-45854-p1a Pressure Injectable Arrowg Plus Ard Blue Plus Four-lumen Cvc Instruction Manual

Teleflex Ca-45854-p1a Pressure Injectable Arrowg Plus Ard Blue Plus Four-lumen Cvc Instruction Manual

Teleflex -logo

CA-45854-P1A Pressure Injectable Arrowg Plus Ard Blue Plus Four-Lumen CVC

Teleflex -icon

Pressure Injectable Arrowg+ard Blue Plus® Four-Lumen CVC

Contents:

1: Arrowg+ard Blue Plusel Four-Lumen Catheter: 8.5 Fr. (3.0 mm OD) x 20 cm, Pressure Injectable
1: Safety Scalpel: #11
1: Checklist/CLIP Sheet
1: Spring-Wire Guide, Marked. 032″ (0 81 mm) dia. x 23-5/8″ (60 cm) (Straight
1: Sterile Procedure Sign Soft Tip on One End – “J” Tip on Other) with Mow Advancer 2: Gauze Pad: 2″ x 2″ (5 cm x 5 cm)
1: Catheter: 18 Ga. x 2-1/2″ (6.35 cm) Radiopaque over 20 Ga. RW Introducer 5: Gauze Pad: 4″ x 4″ (10 cm x 10 cm) Needle
1: Dressing: Tegaderm03 10 cm x 12 cm
1: Pressure Transduction Probe 1: Suture: 3-0 Silk with Curved Needle
1: Introducer Needle: Echogenic 18 Ga. x 2-112″ (6.35 cm) XTW and 5 mL Arrowe
1: HemoHopper0 Fluid Receptacle Raulerson Spring-Wire Introduction Syringe ‘Licensed under US Patent No. 7,329,412.
1: Injection Needle: Eclipse Tm2 22 Ga. x 1-1/2″ (3.81 cm) RW and 5 mL Luer-Slip 2A trademark of Becton, Dickinson and Company. Syringe 3A registered trademark of 3M Company.
1: Injection Needle: EclipseT112 25 Ga. x 1″ (2.54 cm) RW and 3 mL Luer-Lock Syringe
1: Syringe: 10 mL Luer-Lock
1: Tissue Dilator: 10 Fr. (3.5 mm) x 10.2 cm
1: SecondSiteul Adjustable Hub: Fastener
1: SecondSiteul Adjustable Hub: Catheter Clamp
1: SharpsAway€ Disposal Cup
1: SharpsAwaye II Locking Disposal Cup
1: Drape: 41″ x 55″ with 4″ fenestration
1: Needle Holder

Not made with natural rubber latex.
Store below 25°C (77°F). Avoid excessive heat above 40°C (104°F). Contraindications: The Arrowg+ard Blue Plus antimicrobial catheter is contraindicated for patients with known hypersensitivity to chlorhexidine, silver sulfadiazine and/or sulfa drugs.

 

 

Teleflex -icon1

 

 Priming
Volume*
(mL)
Gravity
Flow Ratet
(mUhr)
MAX Pressure
Injection
Flow Rate”
(mUsec)
Distal (16 Ga.)0.4022835
Medial 1 (14 Ga.)0.50431610
Medial 2 (18 Ga.)0.361079not rated
Proximal (18 Ga.)0.391162not rated

* Priming volumes are approximate and are done without accessories.
† Flow rate values are approximate and are determined using deionized water at 100 cm head height.
** Pressure injection flow rates are determined using media of 11.8 centipoise (cP) viscosity, with 152 cm administration set tubing.

LBL053811 R 00 (2021-06)
Teleflex -icon2Teleflex -logo