Scican Bravo G4 Dental Sterilizer User Guide

Bravo G4 Dental Sterilizer

Product Information:

The BRAVOTM G4 is an autoclave designed for sterilizing
instruments and equipment. It features a touchscreen interface for
easy operation and offers various cycles for different types of
instruments. The autoclave has a front and side ventilation system
to ensure proper air circulation. It also has options for top and
front filling of water. The autoclave comes with a maintenance
schedule to ensure proper functioning and longevity.

Key Features:

  • Touchscreen interface
  • Front and side ventilation
  • Multiple cycle options
  • Top and front filling options
  • Maintenance schedule

Spare Parts and Accessories:

Part NumberDescription
97400145Door Gasket 17L/22L
97290160Bacteriological Filter
97290210Chamber Drain Filter
99951785Chamber Rack 17L
99951786Chamber Rack 22L
99950394Pouch Rack
99951317Tray 17L
99951318Tray 22L
97467066Tray Extractor

For more information and support, visit www.scican.com.

Product Usage Instructions:

  1. Before using the autoclave, consult the Operator’s Manual
    located on the USB key.
  2. Ensure the autoclave is powered on.
  3. Fill the autoclave with demineralized/distilled water. Never
    use tap water.
  4. Load the instruments or equipment according to the desired
    configurations.
  5. Close the unit door.
  6. Select a cycle using the touchscreen interface.
  7. Start the cycle to begin the sterilization process.
  8. Once the cycle is completed, open the door.
  9. Unload the sterilized instruments or equipment.
  10. If necessary, drain the autoclave manually.

Maintenance Schedule:

Daily:

  • Clean the door gasket and internal part of the door with a
    clean wet cloth. Check for wear and replace if required.
  • Completely empty the waste tank before refilling the water
    reservoir.

Weekly:

  • Clean the chamber and appropriate accessories with a wet cloth.
    If required, use a neutral detergent (without alkalines and
    bleaches). Carefully rinse with demineralized/distilled water and
    remove any residue.
  • Inspect and clean chamber anti-dust filter. Replace if
    necessary.
  • Check the bacteriological filter and replace as required.
  • Lubricate the door lock. Using a clean cloth, remove any
    residues and lubricate the inside of the bushing with a film of the
    silicone-based grease provided.

Every 3-6 Months:

  • Inspect and clean the chamber drain filter. Replace if
    required.
  • Ensure the area around the BRAVO G4 is dust-free.

Monthly:

  • Drain and clean the internal water reservoir as per the
    instructions outlined in the Operator’s Manual.
  • If applicable, clean the external water tank used for automatic
    filling.

Every Year or 1,000 Cycles:

  • Replace the door gasket.

For further assistance, contact +1 800-572-1211 or +1
800-667-7733.

BRAVOTM G4 Quick Reference Guide
Operator’s Manual (located on the USB key) should be consulted before using this autoclave.

+1 800-572-1211 +1 800-667-7733

BASIC OPERATION
1 Power unit ON.
A
3 Load configurations.

2 Fill with demineralized/distilled water.
NEVER USE TAP WATER

TOUCHSCREEN
Data
Settings

Cycles
Solid Solid, Non-Lumened Instruments
Hollow Hollow, Lumened Instruments
Rubber & Plastics Textiles, Hollow, Lumened, Solid

VENTILATION
Front & Side
0.78in / 20mm

0.78in 20mm

0.78in 20mm

Back

2.75in 70mm

4 Close unit door.

5 Run cycle.
Select a cycle.

132°C Hollow wrapped B

47′

132°C

1,90 bar

132°C Hollow wrapped B

Start a cycle.

132°C Hollow wrapped B
CYCLE COMPLETED
133,3°C 133,1°C 1,941 bar 1,929 bar
End of cycle.

FILLING OPTIONS
Top Filling
Front Filling

6 Open door.

7 Unload instruments.

8 If required, drain manually.

DRAINING OPTIONS

Manually

Automatic

A

MAINTENANCE SCHEDULE

Daily

Clean the door gasket and internal part of the door with a clean wet cloth. Check for wear and replace if required.
Completely empty the waste tank before refilling the water reservoir.

Weekly

Clean the chamber and appropriate accessories with a wet cloth. If required, use a neutral detergent (without alkalines and bleaches). Carefully rinse with demineralized/distilled water and remove any residue.
Inspect and clean chamber anti-dust filter. Replace if necessary.

Check the bacteriological filter and replace as required.
Lubricate the door lock. Using a clean cloth, remove any residues and lubricate the inside of the bushing with a film of the silicone-based grease provided.

Every
3 – 6 Months

Inspect and clean
the chamber drain filter.

Replace if required.

Ensure area around BRAVO G4 is dust free.

Monthly

Drain & clean the internal water reservoir as per the instructions outlined in the Operator’s Manual.
If applicable, clean the external water tank used for automatic filling.

Every Year
or
1,000 Cycles

Replace the door gasket.

www.scican.com
97055284-1 | SD-622 NA EN R1. BRAVO is a trademark of SciCan Ltd. A Coltene Group Company. Manufactured for SciCan Ltd. 1440 Don Mills Rd. Toronto, ON, M3B 3P9 / Canada

To Release

Spare Parts and Accessories

97400145

Door Gasket 17L/22L

97290160

Bacteriological Filter

97290210

Chamber Drain Filter

99951785

Chamber Rack 17L

99951786

Chamber Rack 22L

99950394

Pouch Rack

99951317

Tray 17L

99951318

Tray 22L

97467066

Tray Extractor

IMPORTANT!
Users of this device must be sure to select the most appropriate sterilization cycle according to the recommendations of the leading infection control authorities and local regulatory guidelines / recommendations.

BRAVOMC G4 Guide de référence rapide
Veuillez lire le manuel de l’utilisateur (sur la clé USB) avant d’utiliser cet autoclave.

+1 800-572-1211 +1 800-667-7733

FONCTIONNEMENT DE BASE

1 Allumez l’appareil.

2 Remplissez-le d’eau déminéralisée / distillée.

A
3 Chargez les configurations.

NE JAMAIS UTILISER L’EAU
DU ROBINET

ÉCRAN TACTILE

Données
Info CYCLES ET TESTS
Paramètres

Cycles
Solides Instruments solides sans lumière
Creux Instruments creux à lumières
Caoutchouc et plastiques Textiles, instruments creux, à lumière, solides

VENTILATION

Devant et côté
0.78inp/o2/02m0mmm

Arrière

0.07.878pino 2200mmmm

0.78inpo 20 mmmm

2.75inpo 7700mmmm

4 Fermez la portière de l’appareil.

5 Exécutez le cycle.

CYCLES ET TESTS

132°C Enveloppé creux B

OPTIONS DE REMPLISSAGE.
Remplissage par le haut

132°C Enveloppé creux B
PRESSION

132°C Enveloppé creux B
CYCLE TERMINÉ

Sélectionnez un cycle.

Lancez un cycle.

Fin du cycle.

Remplissage par le devant

6 Ouvrez la portière.

7 Retirez les instruments.

8 Au besoin, faites la vidange manuelle.

OPTIONS DE VIDANGE

Manuelle

Automatique

A

CALENDRIER DE MAINTENANCE

Chaque jour

Nettoyez le joint de la portière et la partie intérieure de la portière avec un linge mouillé propre. Vérifiez s’il y a usure et remplacez-le au besoin.
Vidangez complètement le réservoir à matières avant de remplir le réservoir d’eau.

Vérifiez le filtre bactériologique et remplacez-le au besoin.
Lubrifiez le verrou de la portière. Enlevez les résidus avec un linge propre et lubrifiez l’intérieur du manchon avec une couche de la graisse de silicone fournie.

Chaque semaine

Nettoyez la chambre et les accessoires connexes à l’aide d’un linge mouillé. Au besoin, utilisez un détergent neutre (sans alcalin ni agent de blanchiment). Rincez-la soigneusement avec une eau déminéralisée / distillée et éliminez les résidus.
Inspectez et nettoyez le filtre anti-poussière de la chambre. Remplacez-le au besoin.

Assurez-vous que l’espace autour du BRAVO G4 est exempt de poussière.

Tous les 3 à 6 mois

Inspectez et nettoyez le filtre du drain
de la chambre.

Remplacez-le au besoin.

Chaque mois

Vidangez et nettoyez le réservoir d’eau interne conformément aux instructions données dans le manuel de l’utilisateur.
Le cas échéant, nettoyez le réservoir d’eau externe utilisé pour le remplissage automatique.

Chaque année ou tous les
1 000 cycles

Remplacez le joint de la portière.

www.scican.com
97055284-1 | SD-622 NA FR R1. BRAVO est une marque de commerce de SciCan ltée, une société du Groupe Coltene. Fabriqué pour SciCan Ltée., 1440 Rue Don Mills, Toronto, ON, M3B 3P9 / Canada

Pour désengager

Pièces de rechange et accessoires

97400145

Joint de portière 17 L / 22 L

97290160

Filtre bactériologique

97290210

Filtre du drain de la chambre

99951785

Panier de chambre 17 L

99951786

Panier de chambre 22 L

99950394

Panier à pochette

99951317

Plateau 17 L

99951318

Plateau 22 L

97467066

Extracteur de plateau

IMPORTANT!
L’utilisateur de l’appareil doit s’assurer de choisir le cycle de stérilisation le plus approprié en fonction des recommandations des autorités de prévention des infections et des directives / recommandations réglementaires locales.