Appa Sflex 10t Flexible Current Probe Meter User Manual

sFLEX 10T Flexible Current Probe Meter

sFLEX 10T User Manual

Product Information

The sFLEX 10T is a flexible current probe meter that is
available in TC, SC, JP, and RU. The meter comes with a limited
warranty of years. It has a lock and unlock feature for secure
usage, and it is protected by double or reinforced insulation. The
meter conforms to EU directives and has AC measurement
capabilities. The battery capacity indicator flashes at a slow rate
when the battery capacity is greater than 2.5V and flashes at a
fast rate when the battery capacity is less than 2.5V. The sFLEX
10T is not user-serviceable and should only be repaired or serviced
by qualified personnel.

Usage Instructions

  • Safety Information: To ensure safe operation
    and service of the meter, follow these instructions. Failure to
    observe warnings can result in severe injury.
  • Warning: Identify hazardous conditions and
    actions that could cause bodily harm or death. Always use proper
    terminals, switch position, and range for measurements. To reduce
    the risk of fire or electric shock, do not expose this product to
    rain or moisture. Verify the meter operation by measuring a known
    current. If in doubt, have the meter serviced. Do not apply more
    than the rated voltage/current as marked on the meter. To avoid
    false readings that can lead to electric shock and injury, replace
    the battery as soon as the low battery indicator appears. Do not
    use the meter around explosive gas or vapor. Do not use a flexible
    current sensor if the inner copper wire of the flexible cord is
    visible. De-energize the installation under test or wear suitable
    protective clothing during fitting or removal of the flexible
    current probe. Do not apply around or remove from uninsulated
    hazardous live conductors, which may render electric shock,
    electric burn, or arc flash.
  • Caution: Identify conditions and actions that
    could cause damage to the meter or equipment under test. Do not
    expose the meter to extremes in temperature of high humidity.
  • Power On/Off: To turn on the meter, slide the
    switch to the ON position. To turn off the meter, slide the switch
    to the OFF position. Do not touch the area above the slide switch
    during measurement.
  • Flexible AC Current Probe: Set the DMM at AC
    3.000V range and switch sFlex-T to the suitable range. Calculate
    the correct current according to AC current specifications.
  • Range: In order to get the best output signal,
    select the correct range according to tested current. Position 1 is
    for 30A – 300.0A, Position 2 is for 300A – 3000A. The slide switch
    is for selecting the output signal, with options of 100mV/1A,
    10mV/1A, and 1mV/1A.

sFLEX 10T
User Manual / /

EN Flexible Current Probe Meter TC SC JP RU

YEARS
LIMITED WARRANTY

sFLEX 10T

EN

! Read First
! Safety Information
To ensure safe operation and service of the meter, follow these instructions. Failure to observe warnings can result in severe Injury.
! Warning
Identify hazardous conditions and actions that could cause BODILY HARM or DEATH
– Individual protective equipment should be used if HAZARDOUS LIVE parts in the installation where measurement is to be carried out could be ACCESSIBLE.
– If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
– Always use proper terminals, switch position,and range for measurements.
– To reduce the risk of fire or electric shock, do Not expose this product to rain or moisture.
– Verify the Meter operation by measuring a known Current. If in doubt, have the Meter serviced.
– Do not apply more than the rated voltage / Current, as marked on Meter.
– To avoid false readings that can lead to electric shock and injury, replace battery as soon as low battery indicator.
– Do not use Meter around explosive gas or vapor. – Do not use a flexible current sensor if the inner
copper wire of the flexible cord is visible. – De-energize the installation under test or wear
suitable protective clothing during fitting a removal of the Flexible Current Probe. – Do not apply around or remove from UNINSU LATED HAZARDOUS LIVE conductors, which may render electric shock, electric burn, or arc flash.

1

sFLEX 10T

EN

! Caution
Identify conditions and actions that could cause DAMAGE the meter or equipment under test. Do not expose Meter to extremes in temperature of high humidity.

Symbols as marked on the meter and Instruction card
See instruction manual Do not apply to or remove from HAZARDOUS LIVE conductors. Equipment protected by double or reinforced insulation Battery Earth
Conforms to EU directives Do not discard this product or throw away
AC measurement

Maintenance
Do not attempt to repair this Clamp Meter. It contains no user-serviceable parts. Repair or servicing should only be performed by qualified personnel.
Cleaning
Periodically wipe the case with a dry cloth and detergent. Do not use abrasives or solvents.

2

sFLEX 10T

EN

Power on/off

OFF

Press

ON

Flexible AC current probe

CAT CAT

III IV

1000V 600V

10A

! Do not touch the area above the slide switch during measurement
Set DMM at AC 3.000V range and switch sFlex-T to the suitable range. Calculate the correct current according to “AC Current” specifications.
Lock
Unlock
3

sFLEX 10T

EN

In order to get correct unit and scale of the reading, connect Flex Clamp Meter with clamp meter or DMM which has Flex function. Set the clamp meter or DMM to Flex function and keep the range of Flex Clamp Meter at 3000A.
Range
In order to get the best output singal, select the correct range, according to tested current.

Position 30A 300A
3000A

Best Measurement Range Ù30.00A
30.0A – 300.0A 300A – 3000A

Slide Switch 100mV/1A 10mV/1A 1mV/1A

4

sFLEX 10T

EN

Battery Capacity Indicator

Battery capacity > 2.5V, the battery capacity indicator flash at slow rate. Battery capacity < 2.5V, the battery capacity indicator flash at fast rate.
Replacing Batteries
Lock
Unlock
2
1
Unlock
! Caution
Remove Meter from conductor and turn off power before opening the battery cover.
5

sFLEX 10T

EN

General Specifications
Dimensions (W x H x D) : Approx. 120mm x 280mm x 25mm Weight(without Battery) : 170g Power Requirements : AAA Size Battery x 2 Battery Life : 160 hours ALKALINE Battery Low Batteries Indication : Voltage drops below operating voltage the battery indicator will flash. Maximum Conductor Size : coil size Safety Standard : EN61010-1, EN61010-2-032, EN61010-031 EN61326-1 CAT IV 600V, CAT III 1000V

Environmental Specifications
Indoor Use. Pollution Degree j 2 Maximum operating altitude : 2000m (6562 ft) Operating temperature & Relative Humidity : 0°C – 30°C, Ù80%RH 30°C – 40°C, Ù75%RH 40°C – 50°C, Ù45%RH Storage temperature : -20°C – 60°C, 0 to 80% RH (no batteries). Temperature coefficient : 0.2 x (Specified accuracy) / °C, < 18°C, > 28°C Shock vibration : Sinusoidal vibration MIL-PRF-28800F (5-55 Hz, 3g max.) Drop Protection : 4 feet drop to hardwood on concrete Floor

6

sFLEX 10T

EN

Electrical Specifications
Accuracy is given as ±(% of reading + counts of least significant digit) at 23°C ± 5°C, with relative humidity Less than 80% R.H. Accuracy is specified for a period of one year after calibration. Max. output voltage: 4.5Vpeak Position Error of Clamp: Accuracy and position error assumes centralized primary conductor at optimum position, no external electrical or magnetic field, and within operating temperature range.

Distance from Optimum (mm)

sFlex-10T 15 25 35

Error ±2.0% ±2.5% ±3.0%

AC Current

Range 30.00A 300.0A 3000A

Output Voltage 100mV/1A 10mV/1A 1mV/1A

Accuracy
±(3% of full scale) for
45Hz – 500Hz

Output Noise : < 5mV for each range. Max. output voltage : 4.5Vpeak

7

sFLEX 10T

EN

Limited Warranty
This Meter is warranted to the original purchaser against defects in material and workmanship for 3 years from the date of purchase. During this warranty period, manufacturer will, at its option, replace or repair the defective unit, subject to verification of the defect or malfunction.

This warranty does not cover fuses, disposable batteries, or damage from abuse, neglect, accident, unauthorized repair, alteration, contamination, or abnormal conditions of operation or handling. Any implied warranties arising out of the sale of this product, including but not limited to implied warranties of merchantability and fitness for a particular purpose, are limited to the above.

The manufacturer shall not be liable for loss of use of the instrument or other incidental or consequential damages, expenses, or economic loss, or for any claim or claims for such damage, expense or economic loss. Some states or countries laws vary, so the above limitations or exclusions may not apply to you.

8

sFLEX 10T

TC

!
!

!
·
·
· ·
·
· ·
· ·
·
·

!

9

sFLEX 10T

TC

AC

/

10

sFLEX 10T

TC

CAT CAT

III IV

1000V 600V

10A

!
DMM AC .V sFlex-T “”

11

sFLEX 10T

TC

DMM DMM Flex A

30A 300A 3000A

Ù30.00A
30.0A – 300.0A 300A – 3000A

100mV/1A 10mV/1A 1mV/1A

>.V <.V
12

sFLEX 10T

TC

2
1
!

(xx) : mm x mm x mm () : g : AAA x : : : : EN-, EN–, EN- EN- CAT IV V, CAT III V
13

sFLEX 10T

TC

: m ( ft) : °C – °C, %RH °C – °C, %RH °C – °C, %RH : -°C – °C % RH () : . x () / °C< °C, > °C : MIL-PRF-F (- Hz g ) :

°C ± °C % R.H. ±(%+ )

: .V :

(mm)

sFlex-T 15 25 35

±2.0% ±2.5% ±3.0%

30.00A 300.0A 3000A

100mV/1A 10mV/1A
1mV/1A

: < mV : .V

±(%) Hz – Hz

14

sFLEX 10T

TC

/

15

sFLEX 10T

SC

!
!

!
·
·
· ·
·
· ·
· ·
·
·

!

16

sFLEX 10T

SC

AC

/

17

sFLEX 10T

SC

CAT CAT

III IV

1000V 600V

10A

!
DMM AC .V sFlex-T “”

18

sFLEX 10T

SC

DMM DMM Flex A

30A 300A 3000A

Ù30.00A
30.0A – 300.0A 300A – 3000A

100mV/1A 10mV/1A 1mV/1A

>.V <.V
19

sFLEX 10T

SC

2
1
!

(xx) : mm x mm x mm () : g : AAA x : : : EN-, EN–, EN- EN- CAT IV V, CAT III V
20

sFLEX 10T

SC

: m ( ft) : °C – °C, %RH °C – °C, %RH °C – °C, %RH : -°C – °C % RH () : . x () / °C< °C, > °C : MIL-PRF-F (- Hz g ) :

°C ± °C % R.H. ±(%+ )

: .V :

(mm)

sFlex-T 15 25 35

±2.0% ±2.5% ±3.0%

30.00A 300.0A 3000A

100mV/1A 10mV/1A
1mV/1A

: < mV : .V

±(%) Hz – Hz

21

sFLEX 10T

SC

/

22

sFLEX 10T

JP

!
!

!

·
· · · · ·/ · · · · ·
23

sFLEX 10T

JP

!

EU
AC

24

sFLEX 10T

JP

/

AC
!
DMACVsFlex-T AC

25

CAT CAT

III IV

1000V 600V

10A

sFLEX 10T

JP

DMM DMM A

30A 300A 3000A

Ù30.00A
30.0A – 300.0A 300A – 3000A
26

100mV/1A 10mV/1A 1mV/1A

sFLEX 10T

JP

.V .V

2
1

!

27

sFLEX 10T

JP

×× : mm x mm x mm : g : AAA× : : : EN-, EN–, EN- EN- CAT IV V, CAT III V

: m ( ft) : °C – °C, %RH °C – °C, %RH °C – °C, %RH : -RH : . x () / °C, < °C, > °C : MIL-PRF-F – Hz g) :

28

sFLEX 10T

JP

±R.H.±

: .Vpeak :

(mm)

sFlex-T 15 25 35

±2.0% ±2.5% ±3.0%

AC

30.00A 300.0A 3000A

100mV/1A 10mV/1A
1mV/1A

: mV : .Vpeak

Hz- Hz ±

29

sFLEX 10T

JP

30

sFLEX 10T

RU

!
!
, .
!
, .
– .
– , ,


– , .

– , , ,
– .
– ,

– , , .

31

sFLEX 10T

RU

!
, .

. , ..

AC

. .

,

32

sFLEX 10T

RU

/

CAT CAT

III IV

1000V 600V

10A

AC
!
3000V AC sFlex-T . « AC».

33

sFLEX 10T

RU

, . Flex 3000 .

, .

30A 300A
3000A

Ù30.00A
30.0A – 300.0A 300A – 3000A
34

. 100mV/1A 10mV/1A
1mV/1A

sFLEX 10T

RU

> 2.5V, . < 2.5V, .

2
1

!
, .
35

sFLEX 10T

RU

( x x ) : . 120mm x 280mm x 25mm ( ) : 170g : AAA x 2 : 160 : , . : : EN61010-1, EN61010-2-032, EN61010-031 EN61326-1 CAT IV 600V, CAT III 1000V

. j 2 : 2000m (6562 ft) : 0°C – 30°C, Ù8 0% 30°C – 40°C, Ù7 5% 40°C – 50°C, Ù4 5% : -20°C – 60°C, 0 80% RH ( ). : 0.2 x (. ) / °C, < 18°C, > 28°C : MIL-PRF-28800F (5-55 Hz, 3g .) : 4

36

sFLEX 10T

RU

± (% + ) 23°C ± 5°C ( 80%). . . . .: 4.5V : , , , .

(mm)

sFlex-10T 15 25 35

±2.0% ±2.5% ±3.0%

AC

30.00A 300.0A 3000A

. . 100mV/1A 10mV/1A
1mV/1A

±(3% . )
45Hz – 500Hz

: < 5mV . . . . : 4.5V

37

sFLEX 10T

RU

3 . , , .

, , , . , , , .

, , , , , . , .

38

www.appatech.com
APAC MGL APPA Corporation
[email protected] Flat 4-1, 4/F, No. 35, Section 3 Minquan East Road, Taipei, Taiwan Tel: +886 2-2508-0877
: : : : : : : : : [email protected] : —
400-099-1987 [email protected]
700020050