Creative Pc-100 Spot Check Monitor Instruction Manual

Creative Pc-100 Spot Check Monitor Instruction Manual

CREATIVE logoPC-100 Spot Check Monitor
Instruction Manual

PC-100 Spot Check Monitor

This Manual is written for the current Spot-Check Monitor.
The Manual describes, in accordance with the Spot-Check Monitor’s features and requirements, main structure, functions, specifications, correct methods for transportation, installation, usage, operation, repair, maintenance and storage, etc. as well as the safety procedures to protect both the user and equipment. Refer to the respective chapters for details.
The Manual is published in English and we have the ultimate right to explain the Manual.
Version of This User Manual: Ver1.1
All rights reserved.
Issued Date: November 18, 2021
Product expectancy life: 5 years (not warranty time) Manufacturing date: See label on device 3502-2600017
Marks in the Manual:
CREATIVE PC-100 Spot Check Monitor - icon 31 Caution: must be followed to avoid endangering the operator and the patient.
CREATIVE PC-100 Spot Check Monitor - icon 28  Attention: must be followed to avoid causing damage to the Spot-Check
monitor.
☞ Note: contains some important information and tips about operations and application.
Dear Customers,
Thank you for purchasing this quality product. Please read the following information very carefully before using this device.
Read these instructions carefully before using this Spot-Check Monitor. These instructions describe the operating procedures to be followed strictly. Failure to follow these instructions can cause monitoring abnormity, equipment damage and personal injury. The manufacturer is NOT responsible for the safety, reliability and performance issues and any monitoring abnormality, personal injury and equipment damage due to user’s negligence of the operation instructions. The manufacturer’s warranty service does not cover such faults.
For the user’s convenience, we provide the latest version of PC software for the Spot-Check Monitor, which be downloaded from our website (www.creative-sz.com). Please consult the manufacturer or local distributor for any issue about the software downloading.
Cautions:
CREATIVE PC-100 Spot Check Monitor - icon 31 This device is used for measurement. Do not use for self-diagnose and it is not a treatment device.
CREATIVE PC-100 Spot Check Monitor - icon 31 This device is applicable to adult.
CREATIVE PC-100 Spot Check Monitor - icon 31 This device is not designed to be used in movable vehicle (such as a car, airplane).
CREATIVE PC-100 Spot Check Monitor - icon 31 To avoid any harm during charging, it is recommended that the device shall not be in use while the battery is charging.
CREATIVE PC-100 Spot Check Monitor - icon 31 Check the device, accessories including cable and cuff before use. Stop use if any damage and aging is found.
CREATIVE PC-100 Spot Check Monitor - icon 31 Please check the monitor completely to verify that the accessories can function safely and normally.
CREATIVE PC-100 Spot Check Monitor - icon 31 Do NOT use the device under the flammable gas condition or in any environment that may lead to explosion.
CREATIVE PC-100 Spot Check Monitor - icon 31 The device is forbidden to be used together with other equipment not specified in this Manual.
CREATIVE PC-100 Spot Check Monitor - icon 31 All combinations of equipment must be in compliance with standard of IEC 60601-1-1 medical and electric system requirements.
CREATIVE PC-100 Spot Check Monitor - icon 31 Do not modify or disassemble this device and the cuff without authorization of the manufacturer.
CREATIVE PC-100 Spot Check Monitor - icon 31 Although biocompatibility tests have been performed on all the applied parts, some exceptional allergic patients may still have anaphylaxis. Do NOT apply to those who have anaphylaxis.
CREATIVE PC-100 Spot Check Monitor - icon 31 When disposing of the monitor and its accessories, the local law should be followed.
Contraindication: NO.
Attentions:
CREATIVE PC-100 Spot Check Monitor - icon 28 Please store and operate the device in specified temperature, humility and atmospheric pressure. The excessive ambient environment may affect the measuring result.
CREATIVE PC-100 Spot Check Monitor - icon 28 To avoid damaging device, if the monitor gets wet, please do not start it until it is air-dry.
CREATIVE PC-100 Spot Check Monitor - icon 28  ONLY monitor one person at a time.
Note:
☞ In this Manual, “long press” means press and hold key for about 3 seconds; “short press” means press key for less than 1 second.

OVERVIEW

1.1 Features

  • Blood pressure, oxygen saturation and pulse rate can be measured;
  • A wearable device, and convenient to measure;
  •  Small in size, light in weight, easy to carry and operate;
  •  It’s intended to use in homes;
  • The measurement result can be set to the host device via wireless communication.
  • The measured data can be displayed, stored and manged by APP software installed on the host device, such as smart phone, tablet, and computer, which can be Android and IOS system.
  • The device can option temperature probe to measure temperature, and/or option Blood Glucose Meter to measure blood glucose.

1.2 Application and Scope
The Spot-Check Monitor is applicable for measuring adult’s physiological parameters, such as non-invasive blood pressure (NIBP), functional oxygen saturation (SpO2), and pulse rate (PR).
1.3 Conformation
The Spot-Check Monitor consists of the main unit, functional accessories (cuff, SpO2 sensor) and the software package running on the host system.
1.4 Product Name and Model
Name: Spot-Check Monitor
Model and Configuration: (“√” means function is available, and “–” means function is not available).

ModelConfigurationRemarks
NIBPSp0²Pulse Rate
PC-101V
PC-102VVV
PC-104VVVPC-104 can receive the data from Blood Glucose Meter

1.5 Working Environment and Power Supply

  1. Operating temperature: 5℃~40℃; Operating humility: 15%~93% ; Atmospheric pressure: 70.0kPa~106.0kPa
  2. Power Supply:
    External power supply by AC adapter: Input: a.c. 100V~240V with 50/60Hz, current 0.5A; Output: d.c.5.0V current 1.2A
    Internal power supply by built-in battery: d.c. 3.7V (lithium battery) with capacity 1000mAh. When the device is charging, the icon ” CREATIVE PC-100 Spot Check Monitor - icon ” will be displayed on the screen (in scroll status), when the device is full charged, the icon turns to be ” CREATIVE PC-100 Spot Check Monitor - icon 1 “. Normally, it will take at least 4 hours to be fully charged to the capacity for use of 120 times of NIBP measurement.

APPEARENCE

CREATIVE PC-100 Spot Check Monitor - emonstration

  1. CREATIVE PC-100 Spot Check Monitor - icon 2 “: display screen.
  2. CREATIVE PC-100 Spot Check Monitor - icon 3 “: power switch. Long pressing to turn on/off the device; Short pressing to start/cancel NIBP measurement.
  3. CREATIVE PC-100 Spot Check Monitor - icon 4 “: general connector. It can be used to connect SpO2 probe, temperature probe or Blood Glucose Meter.
  4. CREATIVE PC-100 Spot Check Monitor - icon 5 “: external DC power input socket (Micro USB interface).
  5. Cuff: it is used for wrapping patient’s upper arm for NIBP measurement. It is self-bound to the device and can not be disassembled.
  6. Symbols and description
SymbolDescriptionSymbolDescription
CREATIVE PC-100 Spot Check Monitor - icon 6Type BF applied parts

CREATIVE PC-100 Spot Check Monitor - icon 12

Serial number
CREATIVE PC-100 Spot Check Monitor - icon 7“Attention!”–please refer to the User ManualCREATIVE PC-100 Spot Check Monitor - icon 13Class II device
CREATIVE PC-100 Spot Check Monitor - icon 1Battery is fully chargedCREATIVE PC-100 Spot Check Monitor - icon 14(rolling bar)Battery is charging
CREATIVE PC-100 Spot Check Monitor - icon 8Following WEEE regulations for disposalCREATIVE PC-100 Spot Check Monitor - icon 15Manufacturing date
CREATIVE PC-100 Spot Check Monitor - icon 9Wireless iconCREATIVE PC-100 Spot Check Monitor - icon 16No alarms
CREATIVE PC-100 Spot Check Monitor - icon 10Do not litter at willCREATIVE PC-100 Spot Check Monitor - icon 17Manufacturer (including address)
CREATIVE PC-100 Spot Check Monitor - icon 11Authorised representative in the European communityCREATIVE PC-100 Spot Check Monitor - icon 18UK Responsible Person

Note: the above signs and symbols may appear on your device.

INSTALLATION AND CONNECTION

The measurement result of Spot-Check Monitor can be ALSO displayed, stored and managed via the APP software installed in a host device (such as smart phone , tablet, computer etc.), so before making measurement, please install APP software firstly and make sure the APP software connects  to the Spot-Check Monitor via wireless successfully.
For example of the host device with Android system and iOS system, the APP software is installed in the following procedure:

 

  1. Scanning QR Code by smart phone, please focus the QR Code frame while scanning.
  2. If successfully scanned, the the scanning result, a web link for downloading the APP will be displayed on the smart phone.
  3. Open this web link by a web browser to download the APP software. Install this APP software if  successfully downloaded.

CREATIVE PC-100 Spot Check Monitor - qr codehttp://2bai.co/13921872

For terminal devices with the iOS system (such as iPhone, iPad), the user can also follow this procedure to download:

  1. On the App Store of the device, enter “Shenzhen Creative” into the search function.
    Note: if you use an iPad to search, please select “iPhone only” when searching.
  2. Once the search results are listed, select the result with @health icon “CREATIVE PC-100 Spot Check Monitor - icon 19“, then download the corresponding APP software.

Instruction for Measurement
☞ Make sure the APP software connects with the Spot-Check Monitor successfully.
☞ Refer to the manual of this APP software for detail operating.
You can also download this APP software from the following website: http://www.creative-sz.com/

 MEASUREMENT

4.1 Blood Pressure Measurement
Long pressing power switch ” CREATIVE PC-100 Spot Check Monitor - icon 3 “, the device displays software version number and UID, and 1.5 seconds later, it enters into the default screen, as shown in figure 4.1.CREATIVE PC-100 Spot Check Monitor - Default screenNote: If there is no signal or no operation for 3 minutes, then the device will power off automatically.
Safety instruction before making measurement:

  • Take off the sweater or the thick cloth to make the arm is uncovered or only one thin cloth left.
  •  Before making measurement, do not eat, drink, smoke, take exercise or have a bath.

Operating procedure:

  1. When putting on the cuff, spread the cuff and wrap it around the upper arm evenly to appropriate tightness. The correct cuff position refers to figure 4.2. Note: locate the cuff in such a way that direction mark “▼” directs to the lower arm.CREATIVE PC-100 Spot Check Monitor - Cuff placement
  2. Adjust the teste’s position. Right position: the teste should stand, sit or lay down with calm condition and make the cuff and the teste’s heart on the same level to get accurate measurement.
  3. Short pressing power key to start NIBP measurement.
  4. When the measurement is finished, the measured result and the corresponding emoji will be displayed on the screen, refer to the following figure.CREATIVE PC-100 Spot Check Monitor - icon 21

Screen Description:
CREATIVE PC-100 Spot Check Monitor - icon “: battery capacity.
CREATIVE PC-100 Spot Check Monitor - icon 9 ”: wireless icon. Icon flashing means the host device is connecting to the device; Icon staying means the host device is successfully connected with the device.
“SYS 109 mmHg”: it displays the measured value and the unit of systolic pressure.
“DIA 84 mmHg”: it displays the measured value and the unit of diastolic pressure.
“PR 79 bpm”: it displays the measured value and the unit of pulse rate.
Emoji description:

  • CREATIVE PC-100 Spot Check Monitor - icon 22 : it means the measured result is normal, that is to say, the measured SYS <130 and DIA <85.
  • CREATIVE PC-100 Spot Check Monitor - icon 23 : it means the measured result is high normal, that is to say, the measured SYS is between 130 and 159, and DIA is between 85 and 99.
  • CREATIVE PC-100 Spot Check Monitor - icon 24 : it means the measured result is abnormal, that is to say, the measured SYS is ≥160 and DIA ≥100.

Safety Instructions for blood pressure measurement
CREATIVE PC-100 Spot Check Monitor - icon 29 During measurement, the testee should not talk, move the device, shake body or arm.
CREATIVE PC-100 Spot Check Monitor - icon 29 For the same testee, the testee should wait for 3 minutes for another NIBP measurement.
4.2 SpO2 Measurement (Optional)CREATIVE PC-100 Spot Check Monitor - measurement screenScreen description:
“%SpO2 99”: mark and value of oxygen saturation.
“PR 62″: mark and value of pulse rate.
“%PI 4.1”: mark and value of Perfusion index.
CREATIVE PC-100 Spot Check Monitor - icon 25 “: heart beat icon.
CREATIVE PC-100 Spot Check Monitor - icon 26 “: plethysmogram.
CREATIVE PC-100 Spot Check Monitor - icon 27  “: pulse bar graph.
Operation procedure:

  1. Connect the SpO2 probe to the connector on the sensor connector.
  2. When the device is powered on, insert one finger (index finger is preferred, the nail should be not too long) into the clip of the probe.
  3. Wait for 2 seconds, the device will auto begin to make measurement.
  4. The measured result will be displayed on the screen, as shown in figure 4.4. If the host device is connected to the device, then the measured result can be displayed on the screen of the host device by running APP software.
  5.  Remove the finger, the prompt “Finger out!” pops on the screen .

Note:

  1. PR and SpO2 over-limit indication are available for the device (if the PR and SpO2 options are selected).
  2. If NIBP and SpO2 are measured simultaneously, then the NIBP measured result will be prior to be displayed for 10 seconds.

4.3 Blood Glucose Measurement (Optional)
Note: there are three models of Blood Glucose Meter for optional: Persona Glucose Meter, Yicheng Glucose Meter and, Bioland Glucose Meter, please see the below description for details. Preparing:

  1. Wash the hands with soap and waiting for drying.
  2. Take out the glucose meter, blood-collecting needle (lancing device), blood-collecting pinhead (blood lancet) and test strip, install the pinhead to the blood-collecting needle.

Description for Yicheng Glucose Meter (optional)
Take out the blood-collecting needle and install the pinhead, select level 3 of the prick depth (there are 5 levels in total, the level 5 is deepest). Then pull the shuttle of the needle slightly. As shown in figure 4.5.CREATIVE PC-100 Spot Check Monitor -Blood-collecting needle and pinheadOperation procedure for Yicheng Glucose Meter:

  1. Connect the Yicheng Glucose Meter to the signal connector “ CREATIVE PC-100 Spot Check Monitor - icon 4 ” on the Spot-Check Monitor.
  2. Power on the Glucose Meter, then the “Calibrating” message blinks on the screen which reminds the user to check if the calibration code on device is the same as what on the packaging. If not, please long pressing the switch key of the Glucose Meter to shut off it, and change the calibration code.
  3. Taking a new test strip to insert into the testing port of the device when “Insert” message appears on the screen. Please do not touch the reaction area of the test strip and testing port.
  4. Disinfect the finger needed to test with alcohol and waiting for drying.
  5. Touch the pinhead on the finger and press the shuttle button of the needle. Do not use force to press the finger when collecting blood.
  6. Please put the blood sample on the reaction area of the test strip when “Receive” message appears on the screen. Please not that the reaction area should be full of blood at one time. It is forbidden to put blood on the reaction area repeatedly.
  7. The test result will be displayed in 5 seconds on the screen of APP software (or PC software) of the host device. The default unit is mmol/L.

CREATIVE PC-100 Spot Check Monitor - Blood glucoseDescription for Smartest Persona Glucose Meter (optional)
By use of the optional link cable for Smartest Persona Glucose Meter, connect the Smartest Persona Glucose Meter to the signal connector “CREATIVE PC-100 Spot Check Monitor - icon 4 ” on the Spot-Check Monitor.
Appearance and key functions of the Smartest Persona Glucose Meter:

  1. Test strip slot: when the strip is inserted into the slot, the meter will automatically turn on.
  2.  LCD display.
  3. M key: power on/off, also for memory recalling mode.
  4. C key: Setting mode. Please refer to User Guide for “Smartest Persona Blood Glucose Monitoring System” for detailed function descriptions.
  5. Data interface: can be used to connect the Spot-Check Monitor for data transmitting.
  6. Battery compartment: insert 2 AAA size batteries with the correct polarities.
  7. Ejector: remove the used strip.

CREATIVE PC-100 Spot Check Monitor - Glucose MeterOperations for the Lancing Device and Blood Lancet

  1. Unscrew the lancing device by turning the end cap counter clockwise.
  2. Insert a new lancet firmly into the lancet holder.
  3. Twist off the protective tip of the lancet.
  4. Close the end cap of the lancing device. Slide into the locking position. Refer to figure 3.6B

CREATIVE PC-100 Spot Check Monitor - Blood LancetOperation procedure for Smartest Persona Glucose Meter:

  1. While the meter is off, insert a new test strip into the meter. The Meter will automatically turn on and a blinking blood icon will be displayed on the screen.
  2. Lance the finger and let a blood drop form.
  3. When the blood drop icon is still blinking on the meter, apply the blood drop to the front edge of the test strip. The meter will display the test result after 6 seconds.
  4. Remove the used strips by hand or by pushing the ejector and the meter will turn off and display “OFF” on the screen

CREATIVE PC-100 Spot Check Monitor - Testing instructionRefer to the provided user guide for the “Smartest Persona Blood Glucose Monitoring System” for further detailed instructions.
Safety Instruction for Blood Glucose Measurement
CREATIVE PC-100 Spot Check Monitor - icon 28 The provided test strips should be used with the corresponding Glucose Meter.
CREATIVE PC-100 Spot Check Monitor - icon 28 Do NOT clean or disinfect the finger with iodine.
CREATIVE PC-100 Spot Check Monitor - icon 28 The calibration code must be the same with that on the packaging.
CREATIVE PC-100 Spot Check Monitor - icon 28 The Smartest Persona Glucose Meter will automatically switch to stand-by mode if a test strip is not inserted for 1minute.
CREATIVE PC-100 Spot Check Monitor - icon 28 The test strip will draw blood at one end automatically.
CREATIVE PC-100 Spot Check Monitor - icon 28 Do NOT press or scrape the bleeding finger.
CREATIVE PC-100 Spot Check Monitor - icon 28 The test strip should be used as soon as possible after unpacking, and the unused strips should be kept in an airproof bottle.
CREATIVE PC-100 Spot Check Monitor - icon 28 Only take one measurement per minute.
☞ The blood-collect pinhead is a disposable item. It’s recommended to insert it back to the plastic cover and throw it into the specific dustbin
CREATIVE PC-100 Spot Check Monitor - icon 28 Description for Boland Blood Glucose Meter (optional) Appearance and key functions of the Bioland Blood Glucose Meter:CREATIVE PC-100 Spot Check Monitor - Strip ejectorOperation Procedure for Lancing Device and Blood Lancet

  1. Unscrew the lancing device cap.
  2. Put the needle into the needle holder until the bottom.
  3. Unscrew the needle cap and put it in the recycling dustbin.
  4. Cover the cap.
  5. Adjust the proper level (usually level 3).
  6. Aiming at the sterilized part for blood-taking, press the blue release key to finish blood-taking.
  7. Unscrew the device cap.
  8. Push the ejection button and discard the used lancet in appropriate container.
  9. Cover the device cap.

Refer to figure4.7B.

CREATIVE PC-100 Spot Check Monitor - figFigure 4.7B

Operation procedure for Bioland Blood Glucose Meter:

  1. Take a new test strip, and insert it into the Blood Glucose Meter.
  2.  Refer to below figure 4.7B, install the blood lancet into the lancing device to get the blood sample.
  3. When blood drop icon(“ CREATIVE PC-100 Spot Check Monitor - icon 30 ”)appears, and put the fingertip’s blood to the socking hole of the test trips, as shown in figure 4.7C.
  4. Five seconds later, the testing result will be displayed on the screen of Glucose Meter and Spot-check Monitor, as shown in figure 4.7D.

CREATIVE PC-100 Spot Check Monitor - Spot-check MonitorFigure 4.7C Blood Glucose measurement
Figure 4.7D

4.4 Temperature Measurement (Optional)
The infrared temperature probe is a delicate transducer. To operate please follow these steps and procedures. Failure to accurately operate may cause damage to the probes.
The infrared temperature probe Please place the infrared temperature probe in a stable ambient temperature for 30 minutes before taking a measurement.
Operation procedure:

  1. Connect the infrared temperature probe to the connector of “ CREATIVE PC-100 Spot Check Monitor - icon 4 ”.
  2. When the screen shows as figure 4.8B and the temperature unit “℃” is blinking, the user can begin to take the measurement.
  3. Insert the tip of the temperature probe into the earhole and press the measuring key to start the measurement. A short beep means the measurement has finished and the result will be

CREATIVE PC-100 Spot Check Monitor - infrared temperature probeNote:

  • The infrared temperature probe will switch to stand by automatically if there is no operation for 1 minute. If a further measurement is needed, press the measuring key and repeat step 2 and step 3.
  • Normal body temperature varies depending on the position/area the measurement is taken from.

The following table shows the varying temperature ranges of the different body positions.
CREATIVE PC-100 Spot Check Monitor - icon 31 This device meets requirements established in ASTM Standard (E1965-98).
CREATIVE PC-100 Spot Check Monitor - icon 31 Do NOT use the infrared temperature probe when the subject temperature and ambient temperature are outside the operating ranges specified by the manufacturer.
CREATIVE PC-100 Spot Check Monitor - icon 31 Performance of the device may be adversely affected when one or more of the following occur:
A. Operation outside of the manufacturer specified subject temperature range.
B. Operation outside of the manufacturer specified operating temperature and humidity ranges.
C. Storage outside of the manufacturer specified ambient temperature and humidity ranges.
D. Mechanical shock.
CREATIVE PC-100 Spot Check Monitor - icon 28 Manufacturer defined soiled or damaged infrared optical components.
Temperature varying range at different body positions:

Arm34.7 ~ 37.3 ℃
Oral35.5 ~ 37.5 ℃
Rectal36.6 ~ 38.0 ℃
Ear35.8 ~ 38.0 ℃

Safety Instruction for Temperature Measurement
CREATIVE PC-100 Spot Check Monitor - icon 31 This device meets requirements established in ASTM Standard (E1965-98).
CREATIVE PC-100 Spot Check Monitor - icon 31 Do NOT using the infrared temperature probe when the subject temperature and ambient temperature are outside the operating ranges specified by the manufacturer.
CREATIVE PC-100 Spot Check Monitor - icon 31 Performance of the device may be adversely affected when one or more of the following occur:
A. Operation outside of the manufacturer specified subject temperature range.
B. Operation outside of the manufacturer specified operating temperature and humidity ranges.
C. Storage outside of the manufacturer specified ambient temperature and humidity ranges.
D. Mechanical shock.
CREATIVE PC-100 Spot Check Monitor - icon 28 Manufacturer defined soiled or damaged infrared optical components.
CREATIVE PC-100 Spot Check Monitor - icon 28Do NOT take a measurement when the patient is moving.
CREATIVE PC-100 Spot Check Monitor - icon 28 Patients with tympanitis and otitis problems should NOT use this device.
CREATIVE PC-100 Spot Check Monitor - icon 28 When the infrared temperature probe is connected to the device, the probe will consecutively be at power-on status, therefore pressing the power on/off key on the temperature probe will not cause any effect.

TECHNICAL SPECIFICATIONS

5.1 Blood Pressure Measurement

  1. Blood pressure measuring range: Systolic pressure: 60mmHg~240mmHg Diastolic pressure: 30mmHg~180mmHg
  2. Blood pressure measuring accuracy: mean difference≤5mmHg, standard deviation≤8mmHg
  3. Pneumatic system error: ±3mmHg
  4. Pressure display resolution: 1mmHg

5.2 SpO2 Measurement (Optional)

  1. SpO2 measuring range: 35%~100%
  2. SpO2 measuring accuracy: Arms is not greater than 2% for SpO2 range from 70% to 100%
    Note: Arms is defined as root-mean-square value of deviation according to ISO 9919
  3. Resolution: 1%
  4. Low limit alarm value for SpO2: 90%.
    Note: when the measured SpO2 ≤ 90%, then the over-limit SpO2 value blinks.

5.3 Pulse Rate Measurement (Optional)

  1. Pulse Rate measuring range: 30bpm~240bpm
  2.  Pulse rate measuring accuracy: ±2bpm or ±2% of reading, whichever is greater
  3.  Resolution: 1bpm

Note: when the measured PR ≥ 120bpm or PR ≤50bpm, then the over-limit PR value blinks.
5.4 Temperature Measurement (Optional)

  1. Measuring range: 32.0℃~43.0℃
  2. Measuring accuracy: ±0.2℃ is for TEMP range from 36.0℃ to 39.0℃, and ±0.3℃ is for the rest; ±0.4℉is for TEMP range from 96.8℉to 102.2℉, and ±0.5℉ is for the rest.
  3. Response time: ≤5s

5.5 Blood Glucose Measurement (Optional)

  1. Technique: Amperometric, glucose oxidase
  2. Measuring range: 1.1mmol/L~33.3mmol/L (20~600mg/dL)
  3. Measuring time: 6 seconds
    Refer to the Blood Glucose Meter User Manual for detailed specification.

5.6 Wireless Communication

  1. Emit and receiving frequency: 2402MHz~ 2480MHz
  2. The Channel number: 40
  3. Bandwidth: 2 MHz
  4. Modulation method: GFSK
  5. Effective radiated power : <1mW

5.7 Classification

  1. Protection against electric shock: Class II equipment and internally powered equipment.
  2. The degree of protection against electric shock: Type B applied parts
  3.  The degree of protection against harmful ingress of liquid: The equipment is IPX1 with protection against ingress of liquid.
  4.  Electric-magnetic compatibility: Group I, Class B.
  5. Mode of operation: Continuous

5.8 Dimension and Weight
Dimension: 140.5mm(L) ×60.0mm(W) × 24.5mm(H) Gross Weight: 260g

MAINTENANCE AND SERVICE

The Spot-Check Monitor is designed to serve for 5 years, it should be properly maintained to ensure its maximum performance and long service life. In addition to the warranty period, the manufacturer also offers long-term service for customers. It is important that the user read and follow the operating instructions, important information and maintenance measures.
6.1 Technical Maintenance
6.1.1 Daily Examination
Before using the device, the following checks should be carried out:

  • Check the device for any mechanical damage;
  • Inspect the exposed parts and the connected parts for all the cables, and the accessories;
  • Examine all the functions of the device that are likely to be used for measuring, and ensure that it is in good working condition.

In case of any indication of damage about the function of the device is detected and proven, it is not allowed to use. Please contact the local dealer or the manufacturer.
6.1.2 Battery Maintenance
CREATIVE PC-100 Spot Check Monitor - icon 31 Only use the provided AC adapter to charge the built-in battery, if other charger is used, the device may be damaged.
CREATIVE PC-100 Spot Check Monitor - icon 28 Whether the monitor is on or off, the built-in battery will be charged as long as the monitor is connected to the provided AC adapter and AC power is on. When the battery is full, it will stop charging for protecting the battery from damage.
CREATIVE PC-100 Spot Check Monitor - icon 28 In order to maintain battery power and prolong battery lifetime, please charge the battery routinely. Generally, charge the battery every 3 months if the device has not been used for more than 3 months.
6.2 Cleaning and Disinfection of the Main Unit
CREATIVE PC-100 Spot Check Monitor - icon 31 Switch off the device and disconnect the power adapter before cleaning.
CREATIVE PC-100 Spot Check Monitor - icon 28 Do NOT let any cleanser and disinfectant flow into the connector or any parts of the device.
CREATIVE PC-100 Spot Check Monitor - icon 28 Do NOT put any parts of the monitor or its accessories in the liquid.
CREATIVE PC-100 Spot Check Monitor - icon 28 Do NOT pour the disinfectant on its surface while disinfecting.
6.3 Cleaning and Disinfection of Accessories
CREATIVE PC-100 Spot Check Monitor - icon 28 The accessories can not be entirely immerged into water, liquid or cleanser.
CREATIVE PC-100 Spot Check Monitor - icon 28 Wipe the accessories clean with a mild cloth if it is dirtied.
CREATIVE PC-100 Spot Check Monitor - icon 28 Wipe the accessories clean and keep it in the packaging for maintenance after using.
6.4 Storage
If the device will not be used for long period of time, wipe it clean and keep it in the package, which shall be kept in a dry good ventilation place free from dust and corrosive  gases. Storage environment: Ambient temperature: -20°C~60°C
Relative humidity: ≤93%
Atmospheric pressure: 53.0kPa~106.0kPa

TROUBLESHOOTING

TroublePossible reasonSolution
Can not turn on the deviceThe built-in battery is exhaustedRecharge it by connecting the power supply adapter
Incompatible accessories are misconnected to this deviceRemove the misconnected parts and try again.
No blood pressure resultThe cuff is wrapped around the arm incorrectlyWrap the cuff around the arm correctly
No Sp02 resultSp02 probe is not inserted to the sensor connector or the finger is out of probePlug Sp02 probe to the sensor connector and check if the finger is inserted into the probe properly.
Can not connect to the host
device
1.Bluetooth version is not compatible;
2.Low battery
1.Use the mobile phone with Bluetooth Ver 4.0 and try again.
2.Recharging the battery

 ERROR MESSAGE INTERPRETING

Error CodeDescription
ERR 01Fail to inflate pressure to 30mmHg within 7 seconds (The cuff is not well-wrapped)
ERR 02Cuff pressure is over 295mmHg (Overpressure protection)
ERR 03No valid pulse is detected
ERR 04Excessive motion artifact
ERR 05The measurement fails
ERR 06Air leakage

Note: the above error codes appears when the NIBP measurement fails.

 PACKING LIST

ItemDescriptionQuantityCheck
1Spot-Check MonitorOne pieceOK
2User ManualOne pieceOK
3AC power adapter (charger)One pieceOK
4SpO2 probeOne pieceOptional
5Charging cableOne pieceOK
6HandbagOne pieceOK

Quality Certificate
Name: Spot-Check Monitor
Model:……….
Date:…………
QA:…………………..
This product has been inspected in accordance with the standards specified in the User Manual. Shenzhen Creative Industry Co., Ltd
Warranty Clause

  1. We offer one year free repair for the main unit and 6-month for the accessories since purchasing. The invoice date is the starting date.
  2. Make sure that the Warranty is sealed. Please present both the Warranty and the Invoice when you are requesting free repair.
  3. Please send the device to the specified place for repair.
  4. We will not offer free repair if one of the following condition exists:
    Disassemble or refit the device arbitrarily;
    Incorrect operation;
    Drop, immerse and collision of the device;
    Incorrect repair by unauthorized personnel.
  5. The cost of repair beyond warranty will be taken by the charge rate, and we offer 3-month repair warranty for such malfunction.

Warranty

Device Information:
NameModel
Serial Number:
DateShop
User Information:
NamePost code
Tel:
Add:
Repair Record
DateRepairing ItemRepairer

Appendix
Classification of Blood Pressure Level
The below form is applicable to adult with age over 18 years.

Blood Pressure LevelSYS (mmHg)DIA (mmHg)
Optimal blood pressure<120<80
Normal blood pressure<130<85
High normal blood pressure130-13985-89
Hypertension Grade 1140-15990-99
Hypertension Grade 2160-179100-109
Hypertension Grade 3180110

CREATIVE PC-100 Spot Check Monitor - icon 31EMC Compliance
Note:
Warnings:

  • The instrument conforms to the requirements of IEC60601- 1 – 2 , EN 60601-1-2 and ISO 80601-2-61standards for electromagnetic compatibility.
  • The user shall install and use the EMC information provided in the random file.
  •  Portable and mobile RF communication equipment may affect the performance of the instrument, avoid strong electromagnetic interference when using, such as close to the mobile phone, microwave oven, etc .
  •  The guidance and manufacturer’s declaration are detailed in the table below .
  • The instrument should not be close to or stacked with other equipment. If it must to be close to or stacked, it should be observed and verified to be able to operate normally under its configuration.
  • In addition to the cables sold by the instrument manufacturer as spare parts for internal components, the use of other accessories and cables may result in increased emission or reduced immunity.
Guidance and manufacturer’s declaration-electromagnetic emission
The Spot-Check Monitor is intended for use in the electromagnetic environment specified below. The customer or the use of the Spot-Check Monitor should assure that it is used in such an environment.
Emissions testComplianceElectromagnetic environment-guidance
Conducted emissions CISPR 11Group 1 Class BThe Spot-Check Monitor uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Radiated emissions CISPR 11The Spot-Check Monitor suitable for use in all establishments, including domestic establishments and those directly network that supplies buildings used for domestic purposes.
Harmonic emissions IEC61000-3-2Class A
Voltage fluctuations/flicker emissions IEC61000-3-3Complies

Guidance and manufacturer’s declaration-electromagnetic emission
The Spot-Check Monitor is intended for use in the electromagnetic environment specified below. The customer or the user

of the Spot-Check Monitor should assure that it is used in such an environment.
Immunity testIEC60601 test levelCompliance levelElectromagnetic environment -guidance
Electrostatic discharge(ESD) IEC61000-4-2±8 kV contact ±2 kV, ±4 kV, +8 kV, ± 15 kV air±8 kV contact
+1 kV, ±4 kV, ±8 kV, ± 15 kV air
Floors should be wood, concrete or ceramic tile. if floors are covered with synthetic material, the relative humidity should be at least 30%
Electrical fast transient/ burst IEC61000-4-4±2kV for power Supply lines
±1kV for Input a.c. Power Ports
±2kV for power Supply lines
±1kV for Input a.c. Power Ports
N/A
Surge IEC 61000-4-5±0.5 kV, 1kV line (s) to line(s)
±0.5 kV, ± 1 kV, ±2kV line(s) to earth
±0.5 kV, 1kV line (s) to line(s) ±-0.5 kV, ± 1 kV,  line(s) to earthN/A
Voltage dips, short interruptions and voltage variations on power supply input lines
IEC61000-4-11
<5% UT
(>95% dip in UT) for 0.5 cycle
<40% UT
(60% dip in UT) for 5 cycles
<70% UT
(30% dip in UT) for 25 cycles
<5% UT
(>95% dip in UT) for 5 s
<5% UT
< (>95% dip in UT) for 0.5 cycle <40% UT
(60% dip in UT) for 5 cycles <70% UT
(30% dip in UT) for 25 cycles <5% UT
(>95% dip in 1_ I I for 5 s
N/A
Power frequency( 501-1z/60Hz) magnetic field IEC61000-4-830A/m30A/mPower frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.

Table 3
Guidance and manufacturer’s declaration – electromagnetic immunity
The Spot-Check Monitor is intended for use in the electromagnetic environment specified below. The customer or the user

of The Spot-Check Monitor should assure that it is used in such an electromagnetic environment.
Immunity testIEC60601 test levelCompliance levelElectromagnetic environment -guidance
Conducted RF IEC61000-4-60,15MHz–80MHz 3 V RMS outside the ISM band, 6 V RMS in the ISM0,15MHz–80MHz 3 V RMS outside the ISM band, 6 V RMS in the ISMPortable and mobile RF communications equipment should be used no closer to any part of The Spot-Check Monitor, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance d=1.2 P
 d=1.2 P 80MHz to 800MHz
Radiated RF IEC61000-4-380 MHz to 2.7 GHz 3V/m80 MHz to 2.7 GHz 3V/md=2.3 P 800MHz to 2.5GHz
Where P is the maximum output power rating of the transmitter in watts (W) according  to  the  transmitter manufacturer and is the recommended separation distance in metres (m). b Field    strengths     from     fixed     RFtransmitters,    as    determined    by    an
 electromagnetic site survey ,a should be less than the compliance level in each frequency range .
b Interference may occur in the vicinity of equipment marked with the following symbol.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a: Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, and electromagnetic site survey should be considered If the measured field strength in the location in which The Spot-Check Monitor is used exceeds the applicable RF compliance level above, The Spot-Check Monitor should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating The Spot-Check Monitor.
b: Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.

Table 4

Frequency Range and Level: RF wireless communication equipment
Test Frequency (MHz)ModulationMinimum immunity Level (V/m)immunity Level Applied (V/m)
385**Pulse Modulation: 18 Hz2727
450*FM + 5 Hz deviation: 1 kHz sine
**Pulse Modulation: 18 Hz
2828
710
745
780
**Pulse Modulation: 217 Hz99
810
870
930
**Pulse Modulation: 18 Hz2828
1720
1845
1970
**Pulse Modulation: 217 Hz2828
2450**Pulse Modulation: 217 Hz2828
5240
5500
5785
**Pulse Modulation: 217 Hz99
ATTENTION:
If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.

a) For some services, only the uplink frequencies are included
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual modulation, it would be worst case.
Table 5

Recommended separation distances between portable and mobile RF communication the equipment
The Spot-Check Monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of The Spot-Check Monitor can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the

Spot-Check Monitor as recommended below, according to the maximum output power of the communications equipment.

Rated maximum output power of transmitter W(Watts)Separation distance according to frequency of transmitter M(Meters)
150kHz to 80MHz d=1.2 CREATIVE PC-100 Spot Check Monitor - icon 3280MHz to 800MHz d=1.2CREATIVE PC-100 Spot Check Monitor - icon 3280MHz to 2,5GHz d=2.3 CREATIVE PC-100 Spot Check Monitor - icon 32
0,01N/A0.120.23
0,1N/A0.380.73
1N/A1.22.3
10N/A3.87.3
100N/A1223
For transmitters rated at a maximum output power not listed above, the recommended separation distance in metres (m) can

be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 : At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 : These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Patent
State Intellectual Property Office of the P.R.C. patented and certificated Creative Spot-Check Monitor on March 26th, 2014. Patent Number: ZL 2013 2 0615696.X

Espenstrasse Shenzhen Creative Industry Co., Ltd.
Floor 5, BLD 9, Baiwangxin High-Tech Industrial Park, Songbai Road, Xili Street, Nanshan District, 518110 Shenzhen, P. R. China
Tel: +86-755-2643 3514
E-mail: [email protected]
Fax: +86-755-2643 0930
Website: www.creative-sz.com
CREATIVE PC-100 Spot Check Monitor - icon 33 Shanghai International Holding Corp. GmbH (Europe) Eiffestrasse 80, 20537 Hamburg Germany
CREATIVE PC-100 Spot Check Monitor - icon 34 Etheria Medical Ltd
The Old Brush Factory Unit 2d Whickham Industrial Estate, Swalwell, Newcastle Upon Tyne, United Kingdom, NE16 3DA
Tel: +44-191-4889922
Fax: +44-191-4889922

References

Documents / Resouces

Download manual
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.


Related Manuals