Chauvin Arnoux F201 Clamp Multimeter User Manual

Chauvin Arnoux F201 Clamp Multimeter User Manual

CHAUVIN-ARNOUX-logo

CHAUVIN ARNOUX F201 Clamp Multimeter

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-product-image

F402 Clamp Multimeter

The F402 clamp multimeter is a versatile device designed for measuring AC/DC voltage, current, resistance, continuity, diode test, frequency, and starting current or over-current. It is capable of measuring up to 1000V CAT IV and 1500V CAT III.

Delivery Condition
The F402 clamp multimeter is delivered with the following items:

  • F402 clamp multimeter
  • Test leads
  • Battery (installed)
  • User manual

Keys
The F402 clamp multimeter has several keys:

  • CHAUVIN-ARNOUX-F201-Clamp-Multimeter-1 key: freezes the displayed value
  • REL key: used for relative measurements
  • CHAUVIN-ARNOUX-F201-Clamp-Multimeter-4 key: displays the maximum or minimum value measured
  • CHAUVIN-ARNOUX-F201-Clamp-Multimeter-5 key: used for frequency measurements

Usage Instructions
Before using the F402 clamp multimeter, it is important to read the user manual and observe the precautions for its use.

To measure voltage or current, follow these steps:

  1. Make sure the device is turned off and the test leads are connected properly
  2. Turn on the device and select the appropriate measurement category (IV, III, or II)
  3. Select AC or DC measurement mode using the dial
  4. Place the clamp around the wire or conductor to be measured
  5. Read the measured value on the display

To measure resistance, continuity, diode, or frequency, follow the instructions provided in the user manual.
Always use caution when working with electrical devices and follow proper safety procedures to avoid injury or damage to the device.
You have just acquired an F402 clamp multimeter and we thank you. For best results from your device:

  • read this user manual attentively,
  • observe the precautions for its use.
  • WARNING, risk of DANGER ! The operator should refer to this user’s manual whenever this danger symbol appears.
  • Application or withdrawal authorized on bare conductors carrying dangerous voltages. Type A current sensor as per IEC/EN 61010-2-032 or BS EN 61010-2-032.
    Battery.
  • Equipment protected throughout by double or reinforced insulation.
  • Earth.
  • The CE marking indicates compliance with the European Low Voltage Directive (2014/35/EU), Electromagnetic Compatibility Directive (2014/30/EU), and Restriction of  Hazardous Substances Directive (RoHS, 2011/65/EU and 2015/863/EU).
  • The UKCA marking certifies that the product is compliant with the requirements that apply in the United Kingdom, in particular as regards Low-Voltage Safety, Electromagnetic
  • Compatibility, and the Restriction of Hazardous Substances.
  • AC – Alternating current.
  • AC and DC – Alternating and direct current.
  • The rubbish bin with a line through it indicates that, in the European Union, the product must undergo selective disposal in compliance with Directive WEEE 2012/19/EU. This equipment must not be treated as household waste.

Definitions of the measurement categories

  • Measurement category IV corresponds to measurements taken at the source of low-voltage installations.
    Example: power feeders, meters and protection devices.
  • Measurement category III corresponds to measurements on building installations.
    Example: distribution panel, circuit-breakers, machines or fixed industrial devices.
  • Measurement category II corresponds to measurements taken on circuits directly connected to low-voltage installations. Example: power supply to domestic electrical appliances and portable tools.

PRECAUTIONS FOR USE

This device complies with safety standards IEC/EN 61010-1 or BS EN 61010-1 and IEC/EN 61010-2-032 or BS EN 61010-2-032 for voltages of 1 000 V in category IV and 1 500 V in category III, less than 2 000 m, indoors, with a degree of pollution not exceeding 2.
These safety instructions are intended to ensure the safety of persons and proper operation of the device. If the tester is used other than as specified in this data sheet, the protection provided by the device may be impaired.

  • The operator and/or the responsible authority must carefully read and clearly understand the various precautions to be taken in use.
  • If you use this instrument other than as specified, the protection it provides may be compromised, thereby endangering you.
  • Do not use the instrument in an explosive atmosphere or in the presence of flammable gases or fumes.
  • Do not use the instrument on networks of which the voltage or category exceeds those mentioned.
  • Do not exceed the rated maximum voltages and currents between terminals or with respect to earth.
  • Do not use the instrument if it appears to be damaged, incomplete, or not properly closed.
  • Before each use, check the condition of the insulation on the leads, housing, and accessories. Any element of which the insulation is deteriorated (even partially) must be set aside for repair or scrapped.
  • Use leads and accessories rated for voltages and categories at least equal to those of the instrument. If not, an accessory of a lower category lowers the category of the combined Clamp + accessory to that of the accessory.
  • Observe the environmental conditions of use.
  • Do not modify the instrument and do not replace components with “equivalents”. Repairs and adjustments must be done by approved qualified personnel.
  • Replace the batteries as soon as the symbolCHAUVIN-ARNOUX-F201-Clamp-Multimeter-8 appears on the display unit. Disconnect all cords before opening the battery compartment cover.
  • Use personal protective equipment when conditions require.
  • Keep your hands away from the unused terminals of the instrument.
  • When handling the test probes, crocodile clips, and clamp ammeters, keep your fingers behind the physical guard.
  • As a safety measure, and to avoid repeated overloads on the inputs of the device, we recommend performing configuration operations only when the device is disconnected from all dangerous voltages.

DELIVERY CONDITION

The F402 clamp multimeter is delivered in its packaging box with:

  • 2 banana-banana leads, one red and one black
  • 2 test probes, one red and one black
  • 1 K thermocouple with banana terminations
  • 4 1.5 V batteries
  • 1 carrying bag
  • 1 multilingual getting started guide.

For accessories and spares, visit our web site: www.chauvin-arnoux.com

PRESENTATION

The F402 is a professional electrical measuring instrument that combines the following functions:

  • Current measurement,
  • Measurement of inrush current / over current (True-Inrush),
  • Voltage measurement,
  • Frequency measurement,
  • Continuity test with buzzer,
  • Resistance measurement,
  • Diode test,
  • Temperature measurement,

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-9

Figure 1: the F402 clamp multimeter

ItemDesignationSee §
1Jaws with centering marks (see connection principles)4.5 to 4.12
2Physical guard
3Switch2.1
4Function keys3
5Display unit2.3
6Terminals2.4
7Trigger

THE SWITCH
The switch has five positions. To access the CHAUVIN-ARNOUX-F201-Clamp-Multimeter-10  ,CHAUVIN-ARNOUX-F201-Clamp-Multimeter-11 , CHAUVIN-ARNOUX-F201-Clamp-Multimeter-12CHAUVIN-ARNOUX-F201-Clamp-Multimeter-13functions, set the switch to the desired function. Each  setting is confirmed by an audible signal. The functions are described in the table below:

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-14

ItemFunctionSee §
1OFF mode – Switches the clamp multimeter off4.3
2AC, DC voltage measurement (V)4.5
3AC, current measurement (A)4.9
4Temperature measurement (°C/°F)4.12
5Continuity test   CHAUVIN-ARNOUX-F201-Clamp-Multimeter-6
Resistance measurement Ω
Diode testCHAUVIN-ARNOUX-F201-Clamp-Multimeter-7
4.6

4.7

4.8

    

THE KEYS OF THE KEYPAD
Here are the five keys of the keypad:

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-15

Figure 3: the keys of the keypad

ItemFunctionSee §
1Storage of values, disabling of display
Compensation of the resistance of the leads in the continuity and ohmmeter function
3.1

4.6.1

2Selection of the type of measurement (AC, DC)3.2
3Activation or de-activation of the backlighting of the diplay unit3.3
4Activation or de-activation of the CHAUVIN-ARNOUX-F201-Clamp-Multimeter-4 mode Activation or de-activation of the INRUSH mode in A3.4
5Measurements of frequency (CHAUVIN-ARNOUX-F201-Clamp-Multimeter-5)3.5

THE DISPLAY UNIT
Here is the display unit of the clamp multimeter:

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-16

ItemFunctionSee §
1Display of the modes selected (keys)3
2Display of the measurement value and unit4.5 to 4.12
3Display of the CHAUVIN-ARNOUX-F201-Clamp-Multimeter-4 modes3.4
4Type of measurement (AC or DC)3.2
5Display of the selected modes (switch)4.5
6Spent battery indication6.2

THE SYMBOLS OF THE DISPLAY UNIT

SymbolDesignation
ACAlternating current or voltage
DCDirect current or voltage
CHAUVIN-ARNOUX-F201-Clamp-Multimeter-1Storage of the values and hold of the display
MaxMaximum RMS value
MinMinimum RMS value
VVolt
CHAUVIN-ARNOUX-F201-Clamp-Multimeter-5Hertz
AAmpere
Ohm
mMilli-prefix
kKilo-prefix
→ o ←Compensation of the resistance of the leads
CHAUVIN-ARNOUX-F201-Clamp-Multimeter-6Continuity test
  CHAUVIN-ARNOUX-F201-Clamp-Multimeter-7Diode test
  
CHAUVIN-ARNOUX-F201-Clamp-Multimeter-17Permanent display (automatic switching off de-activated)
 Spent battery indicator

MEASUREMENT CAPACITY EXCEEDED ( O.L)
The O.L (Over Load) symbol is displayed when the display capacity is exceeded.

THE TERMINALS
The terminals are used as follows:

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-18

ItemFunction
1Cold terminal (COM)
2Hot terminal (+)

THE KEYS

The keys of the keypad respond differently to short, long, and sustained presses.
The CHAUVIN-ARNOUX-F201-Clamp-Multimeter-4  and CHAUVIN-ARNOUX-F201-Clamp-Multimeter-5 keys provide new functions and allow the detection and acquisition of parameters complementary to the usual elementary measurements.
Each of these keys can be used independently of the others or in perfect complementarity with them: this makes navigation simple and intuitive for looking up all measurement results.
It is possible, for example, to look up in turn the MAX, MIN, etc. values of the RMS voltage only, then display relative values in parallel.
In this section, theCHAUVIN-ARNOUX-F201-Clamp-Multimeter-19 icon represents the possible positions of the switch for which the key concerned has some action.

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-1HOLD KEY
This key is used to:

  • store and look up the last values acquired specific to each function (V, A, Ω, T°) according to the specific modes previously activated (CHAUVIN-ARNOUX-F201-Clamp-Multimeter-4, CHAUVIN-ARNOUX-F201-Clamp-Multimeter-5); the present display is then maintained while the detection and acquisition of new values continues;
  • perform automatic compensation of the resistance of the leads (see also § 4.6.1);
Successive presses on HOLDCHAUVIN-ARNOUX-F201-Clamp-Multimeter-1

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-19

… serve
short

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-10CHAUVIN-ARNOUX-F201-Clamp-Multimeter-11CHAUVIN-ARNOUX-F201-Clamp-Multimeter-12CHAUVIN-ARNOUX-F201-Clamp-Multimeter-13

to stored the results of the present measurements,
to hold the display of the last value displayed,
to return to normal display mode (the value of each new measurement is displayed)
sustainedCHAUVIN-ARNOUX-F201-Clamp-Multimeter-11to perform automatic compensation of the resistance of the leads (see § 4.6.1)

See also § 3.4.2 and § 3.5.2 for the action HOLD key with the action of the CHAUVIN-ARNOUX-F201-Clamp-Multimeter-4 key and with the action of the CHAUVIN-ARNOUX-F201-Clamp-Multimeter-5 key.

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-2 KEY (SECOND FUNCTION)
This key is used to select the type of measurement (AC, DC) and the second functions marked in yellow next to the relevant positions of the switch.
It can also be used, in the configuration mode, to modify the default value (see § 4.4).
Remark: the key is invalid in the CHAUVIN-ARNOUX-F201-Clamp-Multimeter-4 and CHAUVIN-ARNOUX-F201-Clamp-Multimeter-1 modes.

Successive presses onCHAUVIN-ARNOUX-F201-Clamp-Multimeter-2CHAUVIN-ARNOUX-F201-Clamp-Multimeter-19… serve
 

 CHAUVIN-ARNOUX-F201-Clamp-Multimeter-10

to select AC or DC. Depending on your choice, the screen displays AC or DC.
CHAUVIN-ARNOUX-F201-Clamp-Multimeter-11to cycle through the and diode test            modes and to return to the continuity test     .
CHAUVIN-ARNOUX-F201-Clamp-Multimeter-13to select °C or °F as the unit

KEY
This key is used to backligth the display unit.

Successive presses onCHAUVIN-ARNOUX-F201-Clamp-Multimeter-19… serve
 

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-10CHAUVIN-ARNOUX-F201-Clamp-Multimeter-11CHAUVIN-ARNOUX-F201-Clamp-Multimeter-12CHAUVIN-ARNOUX-F201-Clamp-Multimeter-13

to activate or de-activate the backlighting of the screen

Remark: the backlighting is switched off automatically at the end of 2 minutes.

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-4KEY
IN THE NORMAL MODE
This key activates detection of the MAX and MIN values of the measurements made. Max and Min are the extreme mean values in DC and the extreme RMS values in AC.
Remark: in this mode, the “automatic switching off” function of the device is automatically de-activated. The CHAUVIN-ARNOUX-F201-Clamp-Multimeter-17 symbol is displayed on the screen.

Successive presses onCHAUVIN-ARNOUX-F201-Clamp-Multimeter-4CHAUVIN-ARNOUX-F201-Clamp-Multimeter-19… serve
short 

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-10CHAUVIN-ARNOUX-F201-Clamp-Multimeter-12CHAUVIN-ARNOUX-F201-Clamp-Multimeter-13

 

  • to activate detection of the CHAUVIN-ARNOUX-F201-Clamp-Multimeter-4 values,
  •  to display the MAX or MIN value successively,
  • to return to display of the present measurement without exiting from the mode (the values already detected are not erased).

Remark: the MAX, MIN symbols are both displayed, but only the symbol of the quantity selected blinks.

Example: If MIN ha been selected, MIN blinks and MAX is lit steadily.

long (> 2 sec)

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-10CHAUVIN-ARNOUX-F201-Clamp-Multimeter-11CHAUVIN-ARNOUX-F201-Clamp-Multimeter-12CHAUVIN-ARNOUX-F201-Clamp-Multimeter-13

to exit from the CHAUVIN-ARNOUX-F201-Clamp-Multimeter-4 mode. The values previously recorded are then erased.
Remark: if the HOLD function is activated, it is not possible to exit from the MAX/

MIN mode. TheCHAUVIN-ARNOUX-F201-Clamp-Multimeter-1 HOLD function must first be de-activated.

THE CHAUVIN-ARNOUX-F201-Clamp-Multimeter-4 MODE + ACTIVATION OF THE HOLD MODE

Successive presses on CHAUVIN-ARNOUX-F201-Clamp-Multimeter-4CHAUVIN-ARNOUX-F201-Clamp-Multimeter-19… serve
 short

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-10CHAUVIN-ARNOUX-F201-Clamp-Multimeter-11CHAUVIN-ARNOUX-F201-Clamp-Multimeter-12CHAUVIN-ARNOUX-F201-Clamp-Multimeter-13

to display successively the CHAUVIN-ARNOUX-F201-Clamp-Multimeter-4values detected before the

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-1HOLD key was pressed.

Note: the HOLDCHAUVIN-ARNOUX-F201-Clamp-Multimeter-1 function does not interrupt the acquisition of new MAX, MIN values.

ACCESS TO THE TRUE-INRUSH MODE ( CHAUVIN-ARNOUX-F201-Clamp-Multimeter-4SET TO A )
This key allows measurement of the True-Inrush current (starting current, or over-current in steady-state operation) for AC or DC current only.

Successive presses on CHAUVIN-ARNOUX-F201-Clamp-Multimeter-4CHAUVIN-ARNOUX-F201-Clamp-Multimeter-19… serves
long (> 2 sec)CHAUVIN-ARNOUX-F201-Clamp-Multimeter-12

 

  • to enter in the True-INRUSH mode
  • “Inrh” is displayed for 3 s (the backlighting blinks).
  • the triggering threshold is displayed for 5 s (the backlighting is steady).
  • “——- ” is displayed and the “A” symbol flashes.
  • after detection and acquisition, the inrush current measurement is displayed, after the calculations stage “—— ” (backlighting off)
  • Remark: the A symbol flashes to indicate “surveillance” of the signal.
  • to exit from the True-INRUSH mode (return to simple current measurement).
short (< 2 sec)
Note: a short press is functional only if an True-Inrush value has been detected.
CHAUVIN-ARNOUX-F201-Clamp-Multimeter-12
  • to display the PEAK+ value of the current,
  • to display the PEAK- value of the current,
  • to display the RMS True-Inrush current.

Remark: the A symbol is displayed steadily this sequence.

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-5KEY
This key is used to display the frequency measurements of a signal. Remark: this button is not functional in DC.

THE Hz FUNCTION IN THE NORMAL MODEL

Successive presses on   CHAUVIN-ARNOUX-F201-Clamp-Multimeter-5 … serves
shortCHAUVIN-ARNOUX-F201-Clamp-Multimeter-10CHAUVIN-ARNOUX-F201-Clamp-Multimeter-12
  •  to display:
  • the frequency of the signal measured,
  • the present voltage (V) or current (A) measurement.

THE CHAUVIN-ARNOUX-F201-Clamp-Multimeter-5 FUNCTION + ACTIVATION OF THE HOLD MODE

Successive presses on CHAUVIN-ARNOUX-F201-Clamp-Multimeter-5 … serves
shortCHAUVIN-ARNOUX-F201-Clamp-Multimeter-10CHAUVIN-ARNOUX-F201-Clamp-Multimeter-12
  •  to store the frequency,
  •  to display successively the stored frequency, then the voltage or the current,

USE

COMMISSIONING
Insert the batteries supplied with the device as follows:

  1. Using a screwdriver, unscrew the screw of the battery compartment cover (item 1) on the back of the housing and open it ;
  2. Place the 4 batteries in the compartment (item 2), taking care to get the polarities right ;
  3. Close the battery compartment cover and screw it to the housing.

 

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-20

Figure 6: the battery compartment cover

STARTING UP THE CLAMP MULTIMETER
The switch is set to OFF. Turn the switch to the function of your choice. The whole display lights (all symbols) for a few seconds (see § 2.3), then the screen of the function chosen is displayed. The clamp multimeter is then ready to make measurements.

SWITCHING THE CLAMP MULTIMETER
The clamp multimeter can be switched off either manually, by setting the switch to OFF, or automatically, after ten minutes with no action on the switch and/or the keys. Thirty (30) seconds before the device is switched off, an audible signal sounds intermittently. To re-activate the device, press any key or turn the switch.

CONFIGURATION
As a safety measure, and to avoid repeated overloads on the inputs of the device, we recommend performing configuration operations only when the device is disconnected from all dangerous voltages.

PROGRAMMING OF THE MAXIMUM RESISTANCE ALLOWED FOR A CONTINUITY
To program the maximum resistance allowed for a continuity:

  1.  From the OFF position, hold theCHAUVIN-ARNOUX-F201-Clamp-Multimeter-2 key down while turning the switch toCHAUVIN-ARNOUX-F201-Clamp-Multimeter-11 , until the “full screen” display ends and a beep is emitted, to enter the configuration mode. The display unit indicates the value below which the buzzer is activated and the symbol is displayedCHAUVIN-ARNOUX-F201-Clamp-Multimeter-6.
    The value stored by default is 40 Ω. The possible values lie between 1 Ω and 999 Ω.
  2. To change the threshold, press the CHAUVIN-ARNOUX-F201-Clamp-Multimeter-2key. The right-hand digit flashes: each press on the to the next digit, apply a long press (> 2 s) to theCHAUVIN-ARNOUX-F201-Clamp-Multimeter-2 key.

To exit from the programming mode, turn the switch to another setting. The detection threshold chosen is stored (emission of a double beep).

DE-ACTIVATION OF AUTOMATIC SWITCHING OFF (AUTO POWER OFF)
To de-activate automatic switching off:

  1. In the OFF position, hold the HOLD CHAUVIN-ARNOUX-F201-Clamp-Multimeter-1key down while turning the switch to CHAUVIN-ARNOUX-F201-Clamp-Multimeter-10, until the “full screen” display ends and a beep  is emitted, to enter the configuration mode. The CHAUVIN-ARNOUX-F201-Clamp-Multimeter-17 symbol is displayed.
  2. When the HOLD CHAUVIN-ARNOUX-F201-Clamp-Multimeter-1key is released, the device is in the voltmeter function in the normal mode.
  3. The return to Auto Power OFF takes place when the clamp is switched back on.

PROGRAMMING OF THE CURRENT THRESHOLD FOR THE TRUE INRUSH MEASUREMENT
To program the triggering current threshold of the True INRUSH measurement:

  1. In the OFF position, hold theCHAUVIN-ARNOUX-F201-Clamp-Multimeter-4 key down while turning the switch to CHAUVIN-ARNOUX-F201-Clamp-Multimeter-12 , until the “full screen” display ends and a beep is emitted, to enter the configuration mode. The display unit indicates the percentage overshoot to apply to the measured current to determine the measurement triggering threshold.
    The value stored by default is 10 %, representing 110 % of the established current measured. The possible values are 5 %, 10 %, 20 %, 50 %, 70 %, 100 %, 150 %, and 200 %.
  2. To change the threshold, press the keyCHAUVIN-ARNOUX-F201-Clamp-Multimeter-2. The value flashes: each press on the  CHAUVIN-ARNOUX-F201-Clamp-Multimeter-2 key displays the next value. To  record the chosen threshold, apply a long press (> 2 s) on the CHAUVIN-ARNOUX-F201-Clamp-Multimeter-2 key. A confirmation beep is emitted.
    To exit from the programming mode, turn the switch to another setting. The chosen threshold is stored (emission of a double beep).
    Note: The starting current measurement triggering threshold is fixed at 1 % of the least sensitive range. This threshold is not adjustable.

CHANGE OF TEMPERATURE MEASUREMENT UNIT
To program the measurement unit, °C or °F:

  1.  In the OFF position, hold the keyCHAUVIN-ARNOUX-F201-Clamp-Multimeter-2 down while turning the switch to CHAUVIN-ARNOUX-F201-Clamp-Multimeter-13 , until the “full screen” display ends and a beep  is emitted, to enter the configuration mode. The display unit indicates the existing unit (°C or °F). The default unit is °C.
  2. Pressing the keyCHAUVIN-ARNOUX-F201-Clamp-Multimeter-2 toggles between °C and °F.
    When the desired unit is displayed, turn the switch to another setting. The unit chosen is stoored (emission of a double beep).

DEFAULT CONFIGURATION
To reset the clamp to its default parameters (factory configuration):
In the OFF position, hold the key CHAUVIN-ARNOUX-F201-Clamp-Multimeter-2down while turning the switch to CHAUVIN-ARNOUX-F201-Clamp-Multimeter-12 , until the “full screen” display ends and a beep is  emitted, to enter the configuration mode. The “rSt” symbol is displayed.
After 2 s, the clamp emits a double beep, then all of the symbols of the screen are displayed until theCHAUVIN-ARNOUX-F201-Clamp-Multimeter-2 default parameters are then restored:

  • Continuity detection threshold = 40 Ω
  • True Inrush triggering threshold = 10 %
  • Temperature measurement unit = °C
  • Adapter function scale factor = 1

VOLTAGE MEASUREMENT (V)
To measure a voltage, proceed as follows:

  1. Set the switch to CHAUVIN-ARNOUX-F201-Clamp-Multimeter-10,
  2. Connect the black lead to the COM terminal and the red lead to “+”,
  3. Place the test probes or the crocodile clips on the terminals of the circuit to be measured. The device selects AC or DC automatically according to which measured value is larger. The AC or DC symbol lights in blinking mode.
    To select AC or DC manually, press the yellow key to reach the desired choice. The symbol corresponding to the choice made then lights in fixed mode.

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-21

The measured value is displayed on the screen.

 CONTINUITY TEST
Warning: Before performing the test, make sure that the circuit is off an any capacitors have been discharged.

  1. Set the switch toCHAUVIN-ARNOUX-F201-Clamp-Multimeter-11 , the symbol is displayed.CHAUVIN-ARNOUX-F201-Clamp-Multimeter-6
  2. Connect the black lead to the “COM” terminal and the red lead to “+”.
  3. Place the test probes or the crocodile clips on the terminals of the circuit or component to be tested.

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-22

An audible signal is emitted if there is continuity, and the measured value is displayed on the screen.

AUTOMATIC COMPENSATION OF THE RESISTANCE OF THE LEADS
Warning: before the compensation is executed, the CHAUVIN-ARNOUX-F201-Clamp-Multimeter-4 and CHAUVIN-ARNOUX-F201-Clamp-Multimeter-1 modes must be de-activated.
To perform automatic compensation of the resistance of the leads, proceed as follows:

  1. Short-circuit the leads connected to the device.
  2. Hold the CHAUVIN-ARNOUX-F201-Clamp-Multimeter-1 key down until the display unit indicates the lowest value. The device measures the resistance of the leads.
  3. Release the CHAUVIN-ARNOUX-F201-Clamp-Multimeter-1 key. The correction and the →0← symbol are displayed. The value displayed is stored.
    Remark: The correction value is stored only if it is ≤ 2 Ω. Above 2 Ω, the value displayed blinks and is not stored.

RESISTANCE MEASUREMENT Ω
Warning: Before making a resistance measurement, make sure that the circuit is cold and any capacitors have been discharged.

  1. Set the switch CHAUVIN-ARNOUX-F201-Clamp-Multimeter-11to and press theCHAUVIN-ARNOUX-F201-Clamp-Multimeter-2 key.
    The Ω symbol is displayed.
  2. Connect the black lead to the “COM” terminal and the red lead to “+”.
  3. Place the test probes or the crocodile clips on the terminals of the circuit or component to be measured.

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-23

The measured value is displayed on the screen.
Remark: to measure low resistance values, first carry out the compensation of the resistance of the leads (see § 4.6.1)

DIODE TEST
Warning: Before performing the diode test, make sure that the circuit is cold and any capacitors have been discharged.

  1. Set the switchCHAUVIN-ARNOUX-F201-Clamp-Multimeter-11 to and press theCHAUVIN-ARNOUX-F201-Clamp-Multimeter-2 key twice. TheCHAUVIN-ARNOUX-F201-Clamp-Multimeter-7 symbol is displayed.
  2. Connect the black lead to the “COM” terminal and the red lead to “+”.
  3. Place the test probes or the crocodile clips on the terminals of the component to be tested.

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-24

The measured value is displayed on the screen.

CURRENT MEASUREMENT (A)
The jaws are opened by pressing the trigger on the body of the device. The arrow on the jaws of the clamp (see the diagram below) must point in the presumed direction of flow of the current, from the generator to the load. Make sure that the jaws have closed correctly.
Remark: the measurement results are optimal when the conductor is centered in the jaws (aligned with the centering marks).

AC MEASUREMENT
For an AC current measurement, proceed as follows:

  1. Set the switch to CHAUVIN-ARNOUX-F201-Clamp-Multimeter-12
  2. Encircle only the conductor concerned with the clamp.

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-25

The measured value is displayed on the screen.

 STARTING CURRENT OR OVER-CURRENT (TRUE INRUSH) MEASUREMENT
To measure a starting current or over current, proceed as follows:

  1. Set the switch to CHAUVIN-ARNOUX-F201-Clamp-Multimeter-12 , then apply the clamp around the single conductor concerned.
  2. Effect a long press on the CHAUVIN-ARNOUX-F201-Clamp-Multimeter-4 key. The InRh symbol is displayed, then the triggering threshold. The clamp then awaits detection of the True-Inrush current. “——” is displayed and the “A” symbol flashes.
  3. After detection and acquisition for 100 ms, the RMS value of the True-Inrush current is displayed, along with the PEAK+/PEAK- values subsequently.
  4. A long press on the CHAUVIN-ARNOUX-F201-Clamp-Multimeter-4 key or a change of function leads to exiting from the True-Inrush mode.

Remark: the triggering threshold in A is 20 A if the initial current is zero (starting of installation), it is that set in the configuration (see § 4.4) for an established current (overload in a installation).

FREQUENCY MEASUREMENT (CHAUVIN-ARNOUX-F201-Clamp-Multimeter-5)
The frequency measurement is available in V and A for AC quantities. The measurement is based on a count of the passages of the signal through zero (positive-going edges).

FREQUENCY MEASUREMENT IN VOLTAGE
To measure the frequency in voltage, proceed as follows:

  1. Set the switch toCHAUVIN-ARNOUX-F201-Clamp-Multimeter-10 and press the CHAUVIN-ARNOUX-F201-Clamp-Multimeter-5 key. The CHAUVIN-ARNOUX-F201-Clamp-Multimeter-5 symbol is displayed.
  2. Select AC by pressing the yellow CHAUVIN-ARNOUX-F201-Clamp-Multimeter-2key until the desired choice is reached.
  3. Connect the black lead to the “COM” terminal and the red lead to “+”.
  4. Place the test probes or the crocodile clips on the terminals of the circuit to be measured.

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-26

The measured value is displayed on the screen.

FREQUENCY MEASUREMENT IN CURRENT

  1. Set the switch to CHAUVIN-ARNOUX-F201-Clamp-Multimeter-12 and press the CHAUVIN-ARNOUX-F201-Clamp-Multimeter-5 key. The “CHAUVIN-ARNOUX-F201-Clamp-Multimeter-5” symbol is displayed.
  2. Select AC by pressing the yellow key CHAUVIN-ARNOUX-F201-Clamp-Multimeter-2until the desired choice is reached.
  3. Encircle only the conductor concerned with the clamp.

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-27

The measured value is displayed on the screen.

TEMPERATURE MEASUREMENT
MEASUREMENT WITHOUT EXTENAL SENSOR

  • Set the switch toCHAUVIN-ARNOUX-F201-Clamp-Multimeter-13 .

The temperature displayed (blinking) is the internal temperature of the device, equal to the ambient temperature after a sufficiently
long thermal stabilization time (at least one hour).

MEASUREMENT WITH EXTERNAL SENSOR
The device measures the temperature using a K thermocouple.

  • Connect the K thermocouple to the + and COM input terminals of the device.
  • Set the switch toCHAUVIN-ARNOUX-F201-Clamp-Multimeter-13 .
  • Place the K thermocouple on the element or zone to be measured, which must not be at a dangerous voltage

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-28

The temperature is displayed on the screen.
To change the unit, °F or °C, press the keyCHAUVIN-ARNOUX-F201-Clamp-Multimeter-2.

Remarks:

  • If the external sensor is defective, the temperature displayed blinks.
  • If there are large variations of the environment of the device, the measurement must be preceded by a stabilization time.

CHARACTERISTICS

REFERENCE CONDITIONS

Quantities of influenceReference conditions
Temperature23°C ± 2°C
Relative humidity45 % to 75 %
Supply voltage6.0 V ± 0.5 V
Frequency range of the applied signal45 – 65 Hz
Sine wavepure
Peak factor of the applied alternating signal√ 2
Position of the conductor in the clampcentred
Adjacent conductorsnone
Alternating magnetic fieldnone
Electric fieldnone

CHARACTERISTICS UNDER THE REFERENCE CONDITIONS
The uncertainties are expressed in ± (x % of the reading (R) + y points (pt)).

DC VOLTAGE MEASUREMENT

Measurement range0.00 V to 99.99 V100.0 V to 999.9 V1 000 V to 1 700 V
Specified measurement range0 to 1 600 V
Uncertaintiesfrom 0.00 V to 9.99 V ± (1 % R + 10 pts) from 10.00 V to 99.99 V ± (1 % R + 3 pts)± (1 % R + 4 pts)
Resolution0.01 V0.1 V1 V
Input impedance10 MΩ

AC VOLTAGE MEASUREMENT

Measurement range0.15 V to 99.99 V100.0 V to 999.9 V1 000 V to 1 200 V RMS 1 700 V peak (1)
Specified measurement range (2)0 to 1 100 VAC / 1 600 V peak
Uncertaintiesfrom 0.15 V to 9.99 V ± (1 % R + 10 pts) from 10.00 V to 99.99 V ± (1 % R + 3 pts)± (1 % R + 4 pts)
Resolution0.01 V0.1 V1 V
Input impedance10 MΩ

Note

  1. The display indicates “OL” above 1 700 V.
    Above 1 200 V RMS, a repetitive beep indicates that the voltage being measured is greater than the safety voltage for which the device is guaranteed.
    Bandwidth in AC = 3 kHz.
  2. Any value between zero and the min. threshold of the measurement range (0.15 V) is forced to “——” on the display.

AC CURRENT MEASUREMENT

Measurement range (2)0.25 A to 99.99 A100.0 A to 999.9 A1 000 A (1 500 A peak) (1)
Specified measurement range0 to 100 % of the measurement range
Uncertainties± (1 % R + 10 pts)± (1 % R + 3 pts)
Resolution0.01 A0.1 A1 A

Note

  • The display indicates “OL” above 1 500 A. Bandwidth in AC = 1 kHz.
  • Any value between zero and the min. threshold of the measurement range (0.25 A) is forced to “——” on the display.

TRUE-INRUSH MEASUREMENT

Measurement range10 A to 1 000 AAC10 A to 1 500 ADC
Specified measurement range0 to 100 % of the measurement range
Uncertainties± (5 % R + 5 pts)
Resolution1 A

Specific characteristics in PEAK mode in True-Inrush (from 10 Hz to 1 kHz in AC):

  • Uncertainties: add ± (1.5 % R + 0.5 A) to the values in the tables above.
  • PEAK capture time: 1 ms min. to 1.5 ms max.

CONTINUITY MEASUREMENT

Measurement range0.0 Ω to 999.9 Ω
Open-circuit voltage≤ 3.6 V
Measurement current550 µA
Uncertainties± (1 % R + 5 pts)
Buzzer triggering thresholdAdjustable from 1 Ω to 999 Ω (40 Ω is the default)

RESISTANCE MEASUREMENT

Measurement range (1)0.0 Ω to 99.9 Ω100.0 Ω to 999.9 Ω1 000 Ω to 9999 Ω10.00 kΩ to 99.99 kΩ
Specified measurement range1 to 100 % of the measurement range0 to 100 % of the measurement range
Uncertainties± (1% R + 10 pts)± (1 % R + 5 pts)
Resolution0.1 Ω1 Ω10 Ω
Open-circuit voltage≤ 3.6 V
Measurement current550 µA100 µA10 µA

Note

  • Above the maximum display value, the display unit indicates “OL”.
    The “-” and “+” signs are not managed.

Specific characteristics in MAX-MIN mode:

  • Uncertainties: add 1 % R to the values of the table above.
  • Capture time of the extrema: approximately 100 ms.

DIODE TEST

Measurement range0.000 V to 3.199 VDC
Specified measurement range1 to 100 % of the measurement range
Uncertainties± (1 % R + 10 pts)
Resolution0.001 V
Measurement current0.55 mA
Indication: junction reversed or open-circuitDisplay of “OL” when the measured voltage > 3,199 V

Note: The “-” sign is disabled for the diode test function.

FREQUENCY MEASUREMENT
Characteristics in voltage

Measurement range (1)5.0 Hz to 999.9 Hz1 000 Hz to 9 999 Hz10.00 kHz to 19.99 kHz
Specified measurement range1 to 100 % of the measurement range0 to 100 % of the measurement range
Uncertainties± (0.4 % R + 1 pt)
Resolution0.1 Hz1 Hz10 Hz

Characteristics in current

Measurement range (1)5.0 Hz to 999.9 Hz
Specified measurement range1 to 100 % of the measurement range
Uncertainties± (0.4 % R + 1 pt)
Resolution0.1 Hz

Note

  • If the level of the signal is too low (U < 3 V or I < 3 A) or if the frequency is less than 5 Hz, the device cannot determine
    the frequency and displays dashes “——“.

Specific characteristics in CHAUVIN-ARNOUX-F201-Clamp-Multimeter-4 mode (from 10 Hz to 1 kHz in voltage and from 10 Hz to 1 kHz in current):

  • Uncertainties: add 1 % R to the values of the table above.
  • Capture time of the extrema: approximately 100ms.

TEMPERATURE MEASUREMENT

FunctionExternal temperature
Type of sensorK thermocouple
Measurement range-60.0°C to +999.9°C
-76.0°F to +1 831.8°F
+1 000°C to +1 200°C
+1 832°F to +2 192°F
Etendue de mesure spécifiée1 to 100 % of the measurement range0 to 100 % of the measurement range
Uncertainties (1)1% R ±3°C
1% R ±5.4°F
1% R ±3°C
1% R ±5.4°F
Resolution0.1°C
0.1°F
1°C
1°F

Note

  • The stated external temperature measurement accuracy does not take the accuracy of the K thermocouple into account.
  • Use of the thermal time constant (0.7 min/°C):
    If there is a sudden variation of the temperature of the clamp, by 10°C for example, the clamp will be at 99 % (cnst=5) of the final temperature after 0.7 min/°Cx5-35 min (to which must be added the constant of the external sensor).

Specific characteristics in CHAUVIN-ARNOUX-F201-Clamp-Multimeter-4 mode:

  • Uncertainties: add 1 % R to the values of the table above.
  • Capture time of the extrema: approximately 100 ms.

ENVIRONMENTAL CONDITIONS

Environmental conditionsin usein storage
Temperature– 20°C to + 55°C– 40°C to + 70°C
Relative humidity (RH)≤ 90 % to 55°C≤ 90 % up to 70°C

CHARACTERISTICS OF CONSTRUCTION

HousingRigid polycarbonate shell with moulded elastomer covering
JawsPolycarbonate Opening: 48 mm
Clamping diameter: 48 mm
ScreenLCD display unit Blue backlighting
Dimension: 41 x 48 mm
DimensionH-272 x W-92 x D-41 mm
Weight600 g (with the batteries)

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-29

POWER SUPPLY

Batteries4 x 1.5 V LR6
Mean life> 350 hours (without backlighting)
Duration of operation before automatic switching offAfter 10 minutes without action on the switch and/or keys

COMPLIANCE WITH INTERNATIONAL STANDARDS

Electric safetyCompliant with standards IEC/EN 61010-1 or BS EN 61010-1,
IEC/EN 61010-2-032 or BS EN 61010-2-032: 1 000 V CAT IV and 1 500 V CAT III
Electromagnetic compatibilityCompliant with standard IEC/EN 61326-1 or BS EN 61326-1
Classification: residential environment
Mechanical strengthFree fall: 2 m (in accordance with standard IEC 68-2-32)
Level of protection of the housingHousing: IP 54 (per standard IEC 60529) Jaws: IP 40

VARIATIONS IN THE DOMAIN OF USE

Quantity of influenceRange influenceQuantity influencedInfluence
TypicalMAX
Temperature– 20 … + 55°CVAC VDC
A*
Ω

CHAUVIN-ARNOUX-F201-Clamp-Multimeter-7

T°C


0.1 % R / 10°C
1 % R / 10°C*

(0.2 % R +1°C) / 10°C
0.1 % R / 10°C
0.5 % R / 10°C + 2 pts
1.5 % R / 10°C + 2 pts* 0.1 % R / 10°C + 2 pts (0.3 % R + 2°C) / 10°C
Humidity10 % … 90 %HRΩV A ≤ 1 pt

0.2 % R
0.1 % R + 1 pt
0.1 % R + 2 pts
0.3 % R + 2 pts
CHAUVIN-ARNOUX-F201-Clamp-Multimeter-7 
Frequency10 Hz … 1 kHz
1 kHz … 3 kHz
10 Hz … 400 Hz
400 Hz … 2 kHz
V
A
1 % R + 1 pt
8 % R + 1 pt
1 % R + 1 pt
4 % R + 1 pt
1 % R + 1 pt
9 % R + 1 pt
1 % R + 1 pt
5 % R + 1 pt
Position of the conductor in the jaws (f ≤ 400 Hz)Any position on the internal perimeter of the jawsA1,5 % R3 % R + 1 pt
Adjacent conductor carrying a current of 150 A DC or RMSConductor touching the external perimeter of the jawsA42 dB35 dB
Conductor enclosed by the clamp0-500 ADC or RMSV< 1 pt1 pt
Application of a voltage of the clamp0-1 600 VDC or RMSA< 1 pt1 pt
Peak factor1.4 to 3.5 limited to 1 500 A peak 1 400 V peakA (AC)
V (AC)
1 % R
1 % R
3 % R + 1 pt
3 % R + 1 pt

Note * in temperature: Influence specified until 1 000 ADC

MAINTENANCE

The instrument has no parts that can be replaced by personnel who are not trained and approved. Any non-approved repair or other work, or replacement of a part by an “equivalent”, may severely compromise safety.

CLEANING

  • Disconnect everything connected to the device and set the switch to OFF.
  • Use a soft cloth moistened with soapy water. Rinse with a damp cloth and dry quickly using a dry cloth or forced air.
  • Dry perfectly before putting back into use.

REPLACEMENT OF THE BATTERIES
The symbol CHAUVIN-ARNOUX-F201-Clamp-Multimeter-8 indicates that the batteries are spent. When this symbol appears on the display unit, the batteries must be  replaced. The measurements and specifications are no longer guaranteed. To replace the batteries, proceed as follows:

  • Disconnect the measurement leads from the input terminals,
  • Set the switch to OFF,
  • Use a screwdriver to unscrew the screw securing the battery compartment cover to the back of the housing and open the cover (see § 4.1),
  • Replace all of the batteries (see § 4.1),
  • Close the cover and screw it to the housing.

WARRANTY

Except as otherwise stated, our warranty is valid for 3 years starting from the date on which the equipment was sold. The extract from our General Conditions of Sale is available on our website. www.chauvin-arnoux.com/en/general-terms-of-sale The warranty does not apply in the following cases:

  • Inappropriate use of the equipment or use with incompatible equipment;
  • Modifications made to the equipment without the explicit permission of the manufacturer’s technical staff;
  • Work done on the device by a person not approved by the manufacturer;
  • Adaptation to a particular application not anticipated in the definition of the equipment or not indicated in the user’s manual;
  • Damage caused by shocks, falls, or floods.

FRANCE
Chauvin Arnoux
12-16 rue Sarah Bernhardt
92600 Asnières-sur-Seine
Tél : +33 1 44 85 44 85
Fax : +33 1 46 27 73 89
[email protected]
www.chauvin-arnoux.com

INTERNATIONAL
Chauvin Arnoux
Tél : +33 1 44 85 44 38
Fax : +33 1 46 27 95 69
Our international contacts
www.chauvin-arnoux.com/contacts

References

Documents / Resouces

Download manual
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.


Related Manuals