Titan 5 Star Dentalez Lubricated Prophy Angle Instruction Manual

DENTALEZ Lubricated Prophy Angle

Product Information: TITAN 5 STAR LUBRICATED PROPHY ANGLE

The TITAN 5 STAR LUBRICATED PROPHY ANGLE is a non-sterile dental
instrument used by trained dental professionals for various dental
procedures. It is designed for drilling and preparing dental
cavities for restoration, such as fillings, disking, cavity and
crown preps, polishing, post and pin drilling, and pin setting.

It is important to note that this device should not be used for
oral surgery. The device should only be used by or on the order of
a licensed physician, as restricted by U.S. Federal law.

The TITAN 5 STAR LUBRICATED PROPHY ANGLE does not have internal
air or water cooling capabilities. External water for cooling may
be required during use.

Product Usage Instructions

Installation and Operation

PROPHY ANGLE INSTALLATION FOR TITAN 3 OR TITAN T
SYSTEMS:

  1. Thread the angle onto the motor-to-angle adapter by hand (Star
    Titan 3 Part #264073 or Titan T Part #263970). Ensure it is seated
    completely.
  2. If the angle does not seat on the adapter, remove it and spin
    the motor-to-angle adapter drive gear. Repeat step one.
  3. Use the angle disconnect wrench (Star Part #203048) to tighten
    securely by fitting it into the mating indentations on the angle
    neck.

PROPHY CUP INSTALLATION:

The prophy angle can use either a screw-on or a snap-on type
polishing cup.

  • For snap-on type cups, use the screw-to-button adapter (Star
    Part #258950).
  • To attach the screw-to-button adapter or polishing cup, thread
    it clockwise until snug into the hollow cartridge shaft of the
    prophy angle.

PROPHY CUP REMOVAL:

To remove the prophy cup, unthread the adapter or polishing cup
counterclockwise. If the screw-to-button adapter is difficult to
disengage, remove the prophy cup and use a small screwdriver to
loosen it.

Safety Instructions

Before use, verify concentric rotation by operating the
handpiece outside the oral cavity in a safe place.

Lubricating

  1. Lubricate with DentaLube II (Star Part #262539) once a day or
    before each sterilization procedure.
  2. Remove the prophy cup.
  3. Place a drop of DentaLube II down the center of the hollow
    cartridge shaft, allowing a few seconds for the lubricant to drain
    down.
  4. Place a drop of DentaLube II at the base of the angle, allowing
    a few seconds for the lubricant to spread.

Following these instructions and observing all warnings and
cautions provided in the user manual will ensure safe and effective
usage of the TITAN 5 STAR LUBRICATED PROPHY ANGLE.

TITAN 5 STAR
LUBRICATED PROPHY ANGLE NON-STERILE MD
DENTALEZ, Inc., StarDental ®Division 1816 Colonial Village Lane Lancaster, PA 17601 USA
058952 Rev U Date of Issue: 2022-11-01 ©2022 DENTALEZ, Inc.

DO NOT USE FOR ORAL SURGERY Before operating the handpiece, carefully read and follow all of these instructions. Save
these instructions for future reference and observe all warnings and cautions.
INDICATIONS FOR USE
Used intraorally by trained dental professionals for drilling and preparation of dental cavities for restoration, such as fillings, disking, cavity and crown preps, polishing, post and pin drilling and pin setting.
Caution: U.S. Federal law restricts this device to sale by or on the order of a licensed physician.
WARNING Under normal operating conditions, air is exhausted from the back of the head of this handpiece. If air is directed into an open soft tissue wound or beneath the mucosa or dermis, injury to the patient could result from air emphysema of air embolism.
WARNING This angle has no internal air or water cooling capabilities. External water for cooling maybe required.

CONTENTS
FEATURES INSTALLATION AND OPERATION SAFETY LUBRICATING CLEANING AND STERILIZATION TROUBLESHOOTING CARTRIDGE REPLACEMENT REPLACEMENT PART NUMBERS WARRANTY

FEATURES
Rated for use with Star Motors and Attachments Titan Five Star Prophy Angle: Star Titan 3 Part #265803 Controlled Speed Reduces Prophy Paste `Spin-off’ Field Replaceable Cartridge Quick, Easy Cleaning Compatible with Titan 3 and Titan T Systems Steam Sterilization: 135° C

135°C

Sterilizable up to 135°C

Motor
5K 20K

Accessories
MTA Torque Multiplier & MTA

Ratio
1:1 4:1

Max Allowable RPM
5,000
5,000

INSTALLATION AND OPERATION

PROPHY ANGLE INSTALLATION FOR TITAN 3 OR TITAN T SYSTEMS
MOTOR-TO-ANGLE

1. For the Titan 3 or Titan T systems thread angle onto motor-to-

ADAPTER DRIVE GEAR

angle adapter (FIG. 1), Star Titan 3 Part #264073 or Titan T Part #

263970, by hand. Seat completely.

FIG. 1

CAUTION When angle does not seat on adapter, gears are not meshed correctly. Remove angle; spin motor-to-angle adapter drive gear and repeat step one.

2. Fit angle disconnect wrench (FIG. 2), Star Part #203048, furnished with every angle, to mating indentations on angle neck; tighten securely.

MATING INDENTATIONS

PROPHY CUP INSTALLATION
The prophy angle uses either a screw-on or a snap-on type polishing cup. When using a snap-on type cup, use the screw-to-button adapter, Star Part #258950. Thread the screw-to-button adapter (FIG. 3), or polishing cup, with a clockwise rotation until snug into the hollow cartridge shaft of the prophy angle.

MOTORTO-ANGLE ADAPTER
FIG. 2

TITAN PROPHY ANGLE
DISCONNECT WRENCH

CAUTION Operate the Titan motor in the forward direction only. If the Titan motor is operated in reverse direction, the prophy cup

will separate from the motor.

SCREW-TO-

PROPHY CUP REMOVAL

FIG. 3 BUTTON

ADAPTER

Unthread the adapter, or polishing cup, counterclockwise. If the

screw-to-button adapter is difficult to disengage, remove the prophy cup, and use a

small screwdriver to loosen the screw-to-button adapter.

SAFETY INSTRUCTIONS
PLEASE OBSERVE SAFETY INSTRUCTIONS
Before use, verify concentric rotation by operating the handpiece outside the oral cavity in a safe place.
LUBRICATING
1. Lubricate with DentaLube II, Star Part #262539 once a day or before each sterilization procedure.
2. Remove prophy cup. 3. Place a drop of DentaLube II down the center of the hollow cartridge shaft (allow a few
seconds for the lubricant to drain down the shaft). (See FIG. 4) 4. Place a drop of DentaLube II at the base of the angle. Allow a few seconds for the
lubricant to drain and reach the bronze bushing. (See FIG. 5)

RED TIP APPLICATOR

RED TIP APPLICATOR

CLEANING AND STERILIZATION

CAUTION

It is the responsibility of the user to use only FDA compliant sterilizing equipment and accessories that are capable of meeting the manufacturers recommendations below.

CAUTION

Do not exceed 135°C

Do not use:

· dry heat sterilizer · cold sterilizers · ultrasonic cleaners

·immersion techniques · sterilizer bags containing disinfectants

· detergents

· surface wipes ·surface sprays

· chemical disinfectors · hand soaps · surgical milk

LIMITS ON REPROCESSING

Repeated processing has minimal effect on these Adapters. End of life is normally determined by wear and damage due to use.

INSTRUCTIONS

WARNING

The angle must be cleaned and sterilized before first use and after each patient use.

CONTAINMENT AND It is recommended that angles are reprocessed as soon as is reasonably TRANSPORTATION: practical following use.

PREPARATION FOR Wear puncture resistant gloves to prepare the prophy angle for sterilization.

CLEANING:

Remove prophy cup or screw-to-button adapter from prophy angle. Remove

angel from motor-to-angle adaptor.

CLEANING AUTOMATED:
CLEANING MANUAL:

Washer-disinfectors are not recommended.

1. Clean and inspect once a day.
2. Clean all external surfaces (FIG.6) with isopropyl alcohol and a lint-free cloth or brush.

Old Style: -No Markings on End Cap -Unthread End Cap in CounterClockwise Direction

New Style: -c34o and L
Marked On End Cap -Unthread End Cap in Clockwise Direction

3. Use the plastic wrench to unthread the end cap by turning clockwise. (See FIG.7):
4. Remove the cartridge and inspect as follows:

FIG. 6
UNTHREAD END CAP IN CLOCKWISE DIRECTION

5. If prophy paste is not found on the cartridge, internal cleaning is not necessary. Simply wipe off with lint-free cloth.

PLASTIC
WRENCH FIG. 7

6. If prophy paste is found on the cartridge, clean off the paste manually. The gear assembly and end cap may be ultrasonically cleaned (FIG.8).

CARTRIDGE

FIG. 8

HOUSING

GEAR ASSEMBLY

StarDental® Instrument Solutions

StarDental® Instrument Solutions

FIG. 4

FIG. 5

NOTICE

BRONZE BUSHING O-RING
END CAP
Do not ultrasonically clean any other components.

CLEANING MANUAL CONTINUED:

7. When prophy paste is consistently found in the cartridge, install a new bronze bushing, or a new cartridge, Star Part #258948, which includes the bronze bushing. Replacing the bronze bushing renews the seal and protects the gear and bearing from abrasive paste.
8. Wipe the inside of the housing with a lint-free cloth or cotton swab.
Note: Do not immerse head in solvent.

9. After cleaning, apply one drop of DentaLube II, Star Part #262539, to the inside diameter of the O-ring and to the gear as shown. (See FIG. 9)

FIG. 9

10. To reassemble, insert gear assembly into bronze bushing and guide shaft through the hole in O-ring. Install cartridge into prophy head angle.

11. Turn end cap counterclockwise and hand tighten to secure. Use wrench to fully seat ­ 30° rotation maximum.

NOTICE
LUBRICATION: DISINFECTION: MAINTENANCE:
INSPECTION & FUNCTION TESTING: PACKAGING:
WRAPPED INSTRUMENT STERILIZATION:
DRYING:
STORAGE: ADDITIONAL INFORMATION: MANUFACTURER CONTACT:

Do not overtighten as poor performance and damage to angle may result

Lubricate – See `Lubrication’ instruction section.

Use steam sterilization exclusively for complete sterility as outlined below.

Angle should be routinely maintained, checked for corrosion, worn parts, etc. and should be replaced on an as needed basis.

See Installation, Operating Recommendations, and/or Troubleshooting sections of this manual.

Single: A standard sterilization packaging material must be used. Ensure that the pack is large enough to contain the item without stressing the seals.

Gravity Type Autoclave: Autoclave:
· Minimum of 15 minutes at 132° C minutes at 132° C · Minimum of 30 minutes at 121° C

Pre-Vacuum · Minimum of 4

Do not exceed 135° C

Dry the prophy angle for 30 minutes minimum using the sterilization dry cycle and allow the prophy angle to cool to room temperature in bag prior to use. If moisture is present, increase your sterilizer dry cycle appropriately.

Store in bag after sterilization until ready to use.

Do not exceed the sterilizer manufacturer’s guidelines for load capacity. Use only paper or paper/plastic bags.

See first or last panels of this manual.

TROUBLESHOOTING
PROBLEM: LOW POWER
Cause: Air leak at motor Correction: Replace gasket;
refer to Titan Motor Instruction Manual. Cause: Debris build-up Correction: Refer to Cleaning and Lubricating section. Cause: Faulty angle Correction: Replace cartridge.

Note: If a problem persists, contact dealer for further assistance; or, contact Star technical service at: 1-866-DTE-INFO (1-866-383-4636)

CARTRIDGE REPLACEMENT
1. Remove the end cap using the plastic wrench to turn it clockwise. (See illustration under Cleaning and Lubricating.)
Note: Do not use pliers on the bronze bushing.
2. Pull bronze bushing to remove cartridge from prophy head as shown below in (FIG. 10). If cartridge is not removable by hand, thread a prophy cup or a screw-to-button adapter in the cartridge and tug on it.

FIG. 10

CARTRIDGE

BRONZE BUSHING

GEAR ASSEMBLY
3. Separate gear assembly from the bronze bushing by pulling apart as shown above. 4. If gears are worn, replace cartridge, Star Part #258948. If not, simply wipe clean
with a lint free cloth. 5. Wipe the inside of the housing with a lint free cloth or cotton swab, barely
dampened with solvent. Note: Do not immerse head in solvent. 6. Guide the shaft of the gear assembly through the center of the bronze bushing.
Install the assembled cartridge into the angle housing. 7. Position end cap and turn counterclockwise. Hand tighten to secure. Use wrench to
fully seat ­ 30° maximum.
NOTICE Do not overtighten as poor performance and damage to angle may result.

REPLACEMENT PART NUMBERS

ALWAYS INSIST ON STAR PARTS FOR BEST PERFORMANCE.

DESCRIPTION

ITEM NUMBER

Titan 3 Five Star Prophy Angle Cartridge Lubricated ……………………………………. 258948 (Includes cartridge and bronze bushing with O-ring) Transmission Gear Assembly Lubricated…………………………………………………….. 259008 Prophy Angle End Caps (Package of 2) Lubricated ………………………………………. 265805 Screw-to-Button Adapter …………………………………………………………………………… 258950 Angle Disconnect Wrench…………………………………………………………………………. 203048 O-Rings (Package of 3) ……………………………………………………………………………. 258951 DentaLube II …………………………………………………………………………………………… 262539 Titan 3 Motor-to-Angle Adapter ………………………………………………………………….. 264073 Titan T Motor-to-Angle Adapter ………………………………………………………………….. 263970

IMPORTANT NOTICE ABOUT RETURNING PRODUCT TO DENTALEZ INC.

Anyone wishing to return any DENTALEZ Inc. product to a DENTALEZ facility MUST obtain the proper RETURN AUTHORIZATION NUMBER from the Customer Service Department. The RA Number must appear clearly marked on both the outside and inside of the returned product carton. When calling for a return authorization number you will be asked to provide the following information:

To Obtain A Return Authorization Number Call 1-866-DTE-INFO

1. Product Name, Model Number or Part Number 2. Serial Number of Product 3. Dealer Return Purchase Order Number 4. Reason For Return 5. Copy of Invoice Showing Purchase

Items will be evaluated. Return Authorization Number is not a guarantee for credit. Your cooperation with this procedure will help us expedite processing of any returns.

FOR NEW PRODUCTS, MANUALS AND TECHNICAL INFORMATION, CALL 1-866-DTE-INFO OR VISIT OUR WEB SITE @ WWW.DENTALEZ.COM

LIMITED WARRANTY

O-Rings and Seals are specifically excluded from this warranty.

DENTALEZ warrants the Titan Lubricated Five Star Prophy Angle to be free of defects in material and workmanship, under normal usage, under the following terms:

Star Products: Lubricated Five Star Prophy Angle

Warranty Period*: 3 Months from date of purchase

Please note the following additional terms of our warranty and return policy:
· Warranty covers manufacturing defects only and does not cover defects resulting from abuse, accidents, misuse, improper handling, cleaning, care or maintenance or the failure to follow or observe our operating/use, maintenance and/or installation instructions. Damage resulting from the use of chemicals, cleaners, disinfectants or sterilization is not covered under this warranty nor is normal wear and tear. Failure to use authorized parts or an authorized repair facility voids this warranty.
· Liability is limited to repair or replacement of the defective part(s) or products at our sole discretion. All other liabilities, in particular liability for damages, including, without limitation, special, exemplary, consequential or incidental damages are excluded
· THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. NO EMPLOYEE, REPRESENTATIVE OR DEALER IS AUTHORIZED TO CHANGE THIS WARRANTY IN ANY WAY OR TO GRANT ANY OTHER WARRANTY.
*Provided conditions in warranty are met.

WARRANTY REGISTRATION
Please visit the website below to register your Star product.This will ensure proper registration and prompt and accurate service should any of your purchases need evaluation or repair during the warranty period.
Sta r Pro du c ts: W W W.D EN TAL EZ.c o m /wa rra n t y