Fkg Race Evo Sequences Taper User Manual

Fkg Race Evo Sequences Taper User Manual

FKG RACE EVO Sequences Taper

FKG RACE EVO Sequences Taper

Description

Taper (1) and ISO diameter (2) identification
Description

Depth marks (in millimetres)
Available on 21 / 25 / 31 mm instruments

Example on 25 / 31 mm instruments
Description

Endo stop – ISO length / active part

Description

SequenceGlide pathShaping
RACE EVO 4%RE1
15/.04
RE2
25/.04
RE3
30/.04
RACE EVO 6%RE1
15/.04
RE2
25/.04
RE3
25/.06

General information

Parameters for use :
Recommended speed range : 800-1000 rpm continuous rotation.

Torque : 1.5 Ncm.

Indications for use :
RACE® EVO endodontic instruments are intended for use in medical or hospital facilities, by qualified health professionals.

RACE® EVO instruments are designed for the glide path and shaping of root canals.

Contraindications :
Nickel-Titanium instruments must not be used on individuals with a known allergic sensitivity to these metals.

Adverse reactions : None known.

Notice and precautions :

  • Take several radiographs from different angles to determine the anatomy of the root canals (length, width and curvature).
  • The number of canals to be treated with a RACE® EVO instrument is limited to a maximum of 8 root canals.
  • Irrigate the canal thoroughly and frequently after the RACE® EVO instrument is extracted.
  • Regularly inspect the RACE® EVO instrument during use, and discard if it shows any signs of wear (for example : straightening).
  • If the RACE® EVO instrument is not progressing easily, remove it from the canal, clean it, inspect its cutting edges, then irrigate the canal and recapitulate with an ISO 010 hand file.
  • Always inspect the instrument(s) before use, and discard (it) them if there are any visible defect(s).
  • When an instrument reaches the end of its life, please dispose of it in accordance with the applicable laws and regulations.
  • Visit the FKG Dentaire SA website (www.fkg.ch) for all information relating to the applicable cleaning and sterilization methods.

Symbols

SymbolSterilized product (irradiation)
SymbolSterilizable in a steam sterilizer (autoclave) at the specified temperature
SymbolTo be used in continuous rotation at the recommended speed
SymbolWarning : refer to the user manual
SymbolBatch number
SymbolExpiry date
SymbolDo not use if package is damaged
SymbolNickel-Titanium instrument
SymbolReference number
SymbolManufacturer

Customer Support

FKG Dentaire Sàrl
Le Crêt-du-Locle 4, 2322 Le Crêt-du-Locle, Switzerland
Tel. +41 32 924 22 44, [email protected], www.fkg.ch