Hayward Pool Rx Socket Dosing Pump Owner's Manual

Hayward Pool Rx Socket Dosing Pump Owner's Manual

HAYWARD Pool Rx Socket Dosing Pum logo

HAYWARD Pool Rx Socket Dosing Pum

HAYWARD Pool Rx Socket Dosing Pum productWARNINGS:

HAYWARD Pool Rx Socket Dosing Pum 15

STANDARD SUPPLIED ACCESSORIES

  • Probe Holder
  • Buffer solution HAYWARD Pool Rx Socket Dosing Pum 01
  • Measuring probe
  • probe loading collierHAYWARD Pool Rx Socket Dosing Pum 02

INSTALLATION

General rules

Install the equipment:

  • on the wall shelf .
  • Do not install over a tank of liquids that emanate fumes unless it is hermetically closed.
  • At maximum temperature of 45°C, in a ventilated place and easily accessible by an operator for periodical maintenance.HAYWARD Pool Rx Socket Dosing Pum 03

CONNECTIONS

  • A= ON/OFF switch
  • B=BNC connector of the probe
  • C= electric power cable 230V,50Hz HAYWARD Pool Rx Socket Dosing Pum 04
ELECTRIC CONNECTION

ATTENTION! Verify that the unit’s label values are compatible with main power supply.
Generally it is preferred to turn off the unit when the circulation pump is turned off, for this reason it is recommended to follow the electrical installation of the figure
To avoid damages to the circuit board do not install the power supply of the pump directly in parallel with filtration pump power supply but use a relay. The unit is switched on when the display is lit.HAYWARD Pool Rx Socket Dosing Pum 05

INSTALLATION SCHEME

  1. Analysis point of the probe
  2. 230Vac power supply in parallel with the recirculation pump
  3. Ground socket (optional)
  4. 230Vac electrolysis power supply (bipolar cable)HAYWARD Pool Rx Socket Dosing Pum 06

PERIODIC MAINTENANCE

Calibration of the unit        1 time every 3 months or in case of measure drift
The measuring probe is regarded as a replacing worn part in fact it undergoes a natural aging depending on its use, therefore it does not fall under the warranty.

ELECTRODE CLEANING

Dip the electrode into an acid solution like vinegar (max 2% acid 98% water). Wait five minutes and rinse it with water. It is advisable do not to leave the electrode dry; if it is stored for long time it must be covered with the plastic cap appositely filled with water to keep it with its tip wet

ELECTRODE STORAGE

Always keep the electrodes in the original stocking caps, the liquid inside is a transparent conservation liquid. Never leave the electrodes dry, they will be damaged/ slow response,

PROGRAMMING

  • Push and hold the SET button for 3 seconds, on the display the previously stored value will appear. Continue pushing the SET button to select the setpoint value in a range between 300 and 800mV. Once the desired value selected, wait for 5 seconds without pushing any button and the value will automatically be stored. HAYWARD Pool Rx Socket Dosing Pum 07

CALIBRATION

Attention:

  1. The electrode calibration procedure may require several minutes, for a very accurate reading, it is important to follow the operation sequence indicated here below.
  2. Be sure that the buffer solution used in calibration always matches the indicated value and that they’re not soiled.
  • Dip the probe in the Rx buffer solution at 468mV. Wait a few seconds before reading. HAYWARD Pool Rx Socket Dosing Pum 08
  • Push the CAL button for 3 seconds, the value 470 will start blinking on the display. Calibration is done automatically and the display stops blinking when finished.HAYWARD Pool Rx Socket Dosing Pum 09
    HAYWARD Pool Rx Socket Dosing Pum 10
  • If at the end of calibration, the display shows the message “rc”, push the Cal button to exit and return to reading. Repeat the procedure, if after the second try the writing “rc” reappears, the probe is not working correctly and should be changed.HAYWARD Pool Rx Socket Dosing Pum 11

FUNCTIONING AND VISUALIZATION

This regulator is a controller that visualizes on the display the value of the mV read from the Rx electrode. It powers the electrical outlet at 230V, 50Hz when the measure is lower than the desired custom’s setpoint
The functioning of the Regulator is ON-OFF type :the electrical outlet will not be powered if the measure is greater than the setpoint while it will be powered at 230V,50Hz if the measure is lower than the setpoint.
Functioning examples: Setpoint 730mV
Reading value 720mV, the Regulator will power the electrical outlet in continuous.
Reading value 740mV, the electrical outlet will not be powered.HAYWARD Pool Rx Socket Dosing Pum 12

DEFAULT SETTINGS

  • SETPOINT: Set = 730mV
  • PREARRANGEMENT: chlorinationHAYWARD Pool Rx Socket Dosing Pum 13

HAYWARD Pool Rx Socket Dosing Pum 16

DIMENSIONS

HAYWARD Pool Rx Socket Dosing Pum 14
HAYWARD Pool Rx Socket Dosing Pum 15

EXPLODED VIEW

HAYWARD Pool Rx Socket Dosing Pum 16

DescriptionCode
1Pump’s bracket111.100.018
2 3,5×16 TC+ screw600.030.024
3Cover100.010.206
4Gasket100.027.046
53×8 TC+ screw400.030.027
6Circuit1RR.001.645
7Box100.011.133
8Logo label100.006.691
9Front cover100.010.319
10Polycarbonate100.006.336
112,9×16 TC+ screw400.030.003
12Schuko socket100.050.119
13REG support100.010.836
14BNC100.050.006
15M12 cable clamp+nut1+100.050.062+00.050.063
16ON-OFF switch100.050.020

Documents / Resouces

Download manual
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.


Related Manuals