DENTALEZ 205554 Rubber Dam Radio Lucent Holder

RADIOLUCENT

Will not shadow X-rays.

FUNCTIONAL DESIGN

Will not warp or become brittle. Unaffected by chloroform, germicides or irrigating solutions.

WRAPPED STERILIZATION

A standard sterilization packaging material must be used. Ensure that the pack is large enough to contain the Visi-Frame without stressing the seals

GRAVITY TYPE AUTOCLAVE 

  • Minimum of 15 minutes at 132° C: Dry time: minimum 30 minutes
  • Minimum of 30 minutes at 121° C: Dry time: minimum 40 minutes

PRE-VACUUM AUTOCLAVE 

  • Minimum of 4 minutes at 132° C Dry time: minimum 30 minutes

NOTE: Do not exceed 135° C
Dry using the sterilization dry cycle and allow to cool to room temperature in bag prior to use. If moisture
is present, increase your sterilizer dry cycle appropriately