Smart Splint Pan
Product Information
The Splint Bath is a device designed for cleaning and
maintaining splints. It features a water basin, a water sensor, and
a thermocouple inline filter. The unit operates at a temperature of
135°F/57°C and requires a cleaning solution consisting of vinegar
or coffee machine/commercial descaler. The unit also has a glass
surface that may accumulate calcium over time, which can be easily
cleaned with vinegar or a commercial glass cleaner. The unit is
compatible with tap water, but distilled water can be used in
hard-water environments with the provided proportions.
Product Usage Instructions
- Evacuate the water from the Splint Bath. Refer to the user
manual for the procedure. - Turn the unit off and add water and cleaning solution. Fill the
unit just above the water sensor with a 2:1 ratio of tap water to
cleaning solution. - Turn the unit back on and run it for 10 minutes at a
temperature of 135°F/57°C. - Evacuate the water from the Splint Bath. Refer to the user
manual for the procedure. - Turn the unit off, unplug it, and allow it to cool down before
proceeding. - Clean the inside water basin and the inline filter. Remove the
thermocouple inline filter found inside the water basin beneath the
grate to clean debris. See quick tips section for removal and
cleaning. Return the cleaned inline filter to the
thermocouple. - Refill the Splint Bath with tap water to the operating level.
See quick tips section for distilled water usage. - Run the Splint Bath for 5 minutes. Connect power, turn the unit
on, and prime it to ensure circulation. Refer to the quick tips
section for priming instructions. - If necessary, repeat the cleaning procedures.
Note: For technical service, please contact Strava Solutions via
email at [email protected] or by phone at
423.667.2325.
SPLINT BATH CARE CARD
Refer to the user manual at: www.stravasolutions.com
Cleaning Your Splint Bath
Frequency Two or more times monthly. Strava recommends each facility determine its cleaning frequency based on usage, water grade, visual inspection, and sanitation protocols that apply to your facility and/or JACHO.
1 EVACUATE WATER
See user manual procedure.
2 TURN UNIT OFF-ADD WATER &
CLEANING SOLUTION Fill unit just above the water sensor with a 2:1 ratio of tap water to cleaning solution.
Cleaning Solution = Vinegar -OR- Coffee Machine/ Commercial Descaler
3 TURN UNIT BACK ON
Run for 10 minutes @ 135oF/57oC.
4 EVACUATE WATER
See user manual for procedure.
5 TURN UNIT OFF
Unplug and allow unit cool down before proceeding.
6 CLEAN INSIDE WATER BASIN
Using a Scotchbrite® Scour Pad or cloth, wipe the inside tub with warm water. To remove scum, waterlines, mineral deposits, etc., add a mild detergent. Be sure to rinse throughly and wipe dry. Do NOT use bleach or extra abrasive products such as steel wool.
6 CLEAN INLINE FILTER
Remove the thermocouple inline filter found inside the water basin beneath the grate to clean debris. See “quick tips” section for removal and cleaning. Return the cleaned inline filter to the thermocouple.
8 REFILL WITH WATER
Fill unit to operating level with tap water. See “quick tips” section if you use distilled water.
9 RUN FOR 5 MIN.
Connect power, turn unit on, and prime unit (see “quick tips” section) to ensure circulation.
Repeat procedures if necessary
STRAVA SPLINT BATH CARE CARD
Cleaning The Glass
Over time, water vapor may cause calcium to build upon the glass and in the hinges. This is perfectly normal and easily cleaned with vinegar on a soft cloth or a commercial glass cleaner.
Replacing the UV-C Bulb
Annually Recommend changing the bulb annually even when if is still working.
Purchase Replacement bulbs can be purchased from a Strava dealer.
Replacement Bulb replacement is simple and can be done by anyone, however, we recommend a qualified service technician if your organization has one.
Check out the bulb replacement video at: www.stravasolutions.com
QUICK TIPS
DISTILLED WATER
Tap water is required for the unit to run correctly. Follow these proportions if you are in a hard-water environment and require distilled water.
1 Cup Tap Water
– PER 1 Gal.
Distilled
CLEAN INLINE FILTER
Note: there are two filter versions depending on the model date.
Loosen the filter with supplied Allen key. Wiggle the filter off of the thermocouple. Run under water and scrub out debris from the screen with a hard bristled brush utensil.
Fit the cleaned filter back on the thermocouple by wiggling it into place (screw heads facing out). Tighten the filter, ensuring that it is properly aligned and close to the hole.
PRIMING
Priming removes any trapped air allowing the pump to engage. Insert priming bulb into the priming port. Squeeze and hold until the bulb is fully inflated. Release from the port. Once the water bubbles, the unit is primed. Repeat if necessary.
FOR TECHNICAL SERVICE email: [email protected] tel: 423.667.2325
PART NUMBER 400.610 STRAVA SPLINT BATH CARE CARD ©Copyright 2022 Strava Solutions, LLC All rights reserved