Bante Iodide Ion Selective Electrode User Guide

Bante Iodide Ion Selective Electrode User Guide

BANTE logo

Iodide
Ion Selective Electrode I User Guide

This ion-selective electrode is designed for the detection and analysis of the iodide ion in an aqueous solution and is suitable for laboratory applications.

BANTE Iodide Ion Selective Electrode-

Required Equipment and Solutions

  • An ion meter
  • Volumetric flasks and beakers
  • Distilled or deionized water:
    To prepare the standard solutions or rinse the electrode between measurements.
  • Ionic strength adjuster (order code: ISA-I):
    To keep a constant background ionic strength and adjust the pH.
  • Iodide standard solution 0.1M:
    To prepare this standard solution, half fill a 1-liter volumetric flask with distilled water and add 14.99 grams of analytical grade sodium iodide (NaI) reagent. Swirl the volumetric flask gently to dissolve the reagent and fill it to the mark with distilled water. Cap and upend the volumetric flask several times to mix the solution.
  • Iodide standard solution 1000 ppm:
    To prepare this standard solution, half fill a 1-liter volumetric flask with distilled water and add 1.18 grams of analytical grade sodium iodide (NaI) reagent. Swirl the volumetric flask gently to dissolve the reagent and fill it to the mark with distilled water. Cap and upend the volumetric flask several times to mix the solution.

BANTE Iodide Ion Selective Electrode-fig1

Iodide ISE

Prior to Use

  1. Remove the protective cap from the bottom of the electrode.
  2. Soak the electrode in 100 ppm standard solution for about 10 minutes.

Measurement and Calibration Hints

  • Do not use this electrode to measure the strongly acidic or alkaline samples, strong detergents, and organic solvents, these solutions will cause permanent damage to the electrode.
  • For better accuracy, we recommend adding the ionic strength adjuster (ISA) to all of the standards and samples.
    A typical addition would be 2 ml ISA to 100 ml of standard and sample solutions.
  • During the calibration and measurement, ensure that all standard and sample solutions are at the same temperature.
  • The calibration should be from the lowest concentration standard to avoid cross-contamination.
  • The sample solution must fall in the pH range of 2 to 12. If necessary, add the ionic strength adjuster.
  • Stir the standard and sample solutions at a uniform rate that will promote the accuracy of measurement.

1. Calibrate the meter according to the manufacturer’s instructions.
2. Rinse the electrode with distilled water and blot dry.
3. Place the electrode into the sample and record the stable reading.

Electrode Maintenance

  • Rinse the electrode thoroughly with distilled water after use, wipe clean with lint-free tissue, then replace the protective cap and store the electrode in a dry, cool, and well-ventilated area.
  • Never touch or scratch the ion-sensitive membrane on the bottom of the electrode.
  • If the electrode response becomes sluggish, soak the electrode in 100 ppm standard solution for at least 1 hour.

Specifications

ModelISE-I
Concentration Range0.06 to 127000 ppm
Slope54 to 59 mV/decade
pH Range2 to 12
InterferencesS²-, CN-
Operating Temperature5 to 50°C (41 to 122°F)
Electrode Dimensions150 (L) × 12 (Dia.) mm (5.9 × 0.47″)
Cable Length1 m (3.3 ft)
ConnectorBNC
Body TypeEpoxy

Optional Accessories

Order CodeDescription
ION-IIodide standard solution 1000 ppm, 480 ml
ISA-IIonic strength adjuster, 480 ml

Bante Instruments Inc.
BANTE -icon[email protected]
BANTE -icon1www.bante-china.com 

References

Documents / Resouces

Download manual
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.


Related Manuals