Solinst 101 P2 Water Level Meters Installation Guide

101 P2 Water Level Meters

Product Information:

The Model 101 P2 Water Level Meter is a groundwater and surface
water monitoring instrument. It comes in two versions, Mk1 and Mk2.
The Mk1 version has a Molex connector from the faceplate
electronics with a 2 pin connection, and the Mk2 version has
push-release fittings on the circuit board to remove the tape
leads. The product comes with a replacement tape and a new grommet.

Tools and Materials Needed:

  • Phillips or Robertson Screwdriver
  • Wire Cutters
  • 3 Pin Molex Connector Housing
  • Jumper Cable (required only for connecting the new tape to Mk1
    Model 101 Water Level Meter with Molex connector)

Product Usage Instructions:

  1. Place the reel on a flat workbench with faceplate up. Undo the
    three screws from the faceplate, and slowly remove it from the
    reel.
  2. For the Mk1 Model 101 P2, disconnect the Tape Molex Connector
    from the Faceplate Molex Connector. For the Mk2 Model 101 P2, press
    down on the white terminals of the push-release fittings on the
    circuit board to remove the tape leads.
  3. Undo the two screws from the black plastic bridge holding the
    tape inside the reel hub, and remove the top piece of the
    bridge.
  4. For the Mk1 Model 101 P2, use wire cutters to cut the old Tape
    Molex Connector from the old tape. Pull the old tape through the
    grommet and remove it from the reel. For the Mk2 Model 101 P2 Water
    Level Meters, simply pull the old tape through the grommet and
    remove it from the reel.
  5. Feed the new tape through the grommet into the reel hub. Note:
    The replacement tape comes with a new grommet. The old grommet may
    be replaced with the new one, or left in if not damaged.
  6. For the Mk1 Model 101 P2, by hand, insert the connector pins
    into the new Tape Molex Connector housing. The negative connector
    pin is inserted into the terminal on the pointed side of the Tape
    Molex Connector housing and the positive pin into the middle
    terminal (the third terminal is left open). The negative pin is
    above the numbers on the tape.
  7. For the Mk2 Model 101 P2 Water Level Meters, cut the pins on
    the tape leads at the point shown in the photo. Use pliers to
    flatten the pins against the tape leads, so they fit into the
    terminals on the circuit board.
  8. Position the tape inside the reel hub with black plastic bridge
    over top and refasten the two screws to secure the tape to the
    reel.
  9. For the Mk1 Model 101 P2, use the Jumper Cable to connect from
    the new 3 Pin Tape Molex Connector to the 2 Pin Faceplate Molex
    Connector.

Connecting Polyethylene Replacement Tape to Reel

More Info | Instructions | Get Quote

Model 101 P2

Tools and Materials Needed
1. Model 101 P2 Replacement Tape Assembly · Comes with P2 Probe attached · Grommet · Jumper Cable (New Tape – 3 Pin to 2 Pin) (#110508) · 3 Pin Molex Connector Housing
2. Phillips or Robertson Screwdriver 3. Wire Cutters

3 Pin Molex Connector Housing

Jumper Cable

Note: The Jumper Cable is required if you are connecting the new tape to a Mk1 Model 101 Water Level Meter that has a Molex connector from the faceplate electronics with a 2 pin connection.

3. For the Mk2 Model 101 P2 Water Level Meters, press down on the white terminals of the push-release fittings on the circuit board to remove the tape leads.

Instructions
1. Place reel on a flat workbench with faceplate up. Undo the three screws from the faceplate, and slowly remove it from the reel.
2. For the Mk1 Model 101 P2, disconnect the Tape Molex Connector from the Faceplate Molex Connector.

Wires Connected to Faceplate
Faceplate Molex Connector

Tape Molex Connector

Tape Leads

Grommet
Black Plastic Bridge
Inside View of Mk1 101 P2 Reel Hub Showing Tape Connection Detail

Grommet
Black Plastic Bridge
Inside View of Mk2 101 P2 Reel Hub Showing Tape Connection Detail
4. Undo the two screws from the black plastic bridge holding the tape inside the reel hub, and remove the top piece of the bridge.
5. For the Mk1 Model 101 P2, use the wire cutters to cut the old Tape Molex Connector from the old tape. Pull the old tape through the grommet and remove it from the reel.
6. For the Mk2 Model 101 P2 Water Level Meters, simply pull the old tape through the grommet and remove it from the reel.
7. Feed the new tape through the grommet into the reel hub.
Note: The replacement tape comes with a new grommet. The old grommet may be replaced with the new one, or left in if not damaged.

(Page 1 of 2)

High Quality Groundwater and Surface Water Monitoring Instrumentation

8. For the Mk1 Model 101 P2, by hand, insert the connector pins into the new Tape Molex Connector housing. The negative connector pin is inserted into the terminal on the pointed side of the Tape Molex Connector housing and the positive pin into the middle terminal (the third terminal is left open). The negative pin is above the numbers on the tape (see diagram below).

3 Pin Molex Connector Housing

Negative Connector Pin

12.For the Mk2 Model 101 P2 Water Level Meters, press down on the white terminals on the circuit board and insert the tape leads. Release the terminals and the leads should be secured. The negative pin on the top tape lead (numbers facing up) is inserted into the terminal with a white square below it on the circuit board. See following images.
Negative pin inserted into terminal with white square below it on the circuit board.

2

1

102

2

1

9. For the Mk2 Model 101 P2 Water Level Meters, cut the pins on the tape leads at the point show in the photo below. Use pliers to flatten the pins against the tape leads, so they fit into the terminals on the circuit board (see Step 12).

White Square

10.Position the tape inside the reel hub with black plastic bridge over top and refasten the two screws to secure the tape to the reel.
11.For the Mk1 Model 101 P2, use the Jumper Cable to connect from the new 3 Pin Tape Molex Connector to the 2 Pin Faceplate Molex Connector.

Connect to Faceplate Molex
Connector

Connect to Tape Molex Connector

Jumper Cable (New Tape – 3 Pin to 2 Pin) (#110508)

Mk2 101 P2 Tape Connection
13.With the probe in a glass of tap water, turn the Meter `ON’. If the buzzer or light do not activate, check the tape connections.
14.Replace the faceplate on the reel and re-secure the three screws.
15.Slowly wind the tape onto the reel, holding to ensure no slack.