Mtts Wm-wl-man-03 Wallaby Warmer User Guide

WM-WL-MAN-03 Wallaby Warmer

Wallaby Warmer

Quick Reference Guide

Refer to User Manual for comprehensive information

Modes of operation

Touch MODE Selection panel to enter MODE Selection screen

  • BABY MODE: Set temperature: 34oC – 37.5oC. In
    BABY MODE, the warmer provides stable control of the baby’s skin
    temperature by automatically adjusting the heater power to
    compensate for varying physiological and environmental
    conditions.
  • PREWARM MODE: Set temperature: not available.
    In PREWARM MODE, the warmer provides general alarm-free prewarming
    of the heater and environment above the warmer mattress.
  • MANUAL MODE: Set heater power: 0-100%. In
    MANUAL MODE, the warmer provides user-adjustable heater power and
    the option to monitor the baby’s skin temperature using skin
    sensor. If the patient skin temperature is more than 1oC above the
    set point, then the heater is OFF and the output value displayed is
    0 (zero).

Timers

  • APGAR Alarm at 1, 5 and 10 minutes. Switches
    off after 60 minutes.
  • PROCEDURE TIMERS Alarm at the end of preset
    period from 1 to 99 minutes.

Alarms

AlarmModeCauseEffectMute Button
HIGH SKIN TEMPBABYMeasured skin temperature higher than set temperature by
1oC
Heater power disabledSilences audible alarm only
LOW SKIN TEMPBABYMeasured skin temperature higher 38oCHeater power disabledSilences audible alarm only
SENSOR DISCONNECTBABY/MANUAL/PREWARM BABY/MANUAL PREWARM BABY/MANUAL PREWARM
ALL
Skin sensor unplugged or faultyHeater reduced to 25% of powerNot active
POWER FAILUREALLPower supply to Warmer failedWarmer disabledSilences audible alarm only
ERRORALLSoftware or hardware failureWarmer disabledSilences audible alarm only (refer to Service Manual for more
information on error codes)

When available, the mute alarm period is 10 min.

X-ray procedure

  1. Turn the main switch of the device off and silence the power
    failure alarm
  2. Wait until the device cools down
  3. Lift the overhead unit up
  4. Move overhead unit left or right
  5. After X-ray procedure is completed return the overhead unit to
    center position
  6. Make sure the overhead unit is locked in center position
  7. Turn the device on

Note: The action will activate POWER FAILURE
alarm.

Turning off the warmer

  1. Press and hold the OFF symbol for 3 seconds

V.1.03.EN

Product Usage Instructions

To use the Wallaby Warmer, select the desired mode of operation
by touching the MODE Selection panel. There are three modes
available:

  • BABY MODE: This mode is used to provide stable
    control of the baby’s skin temperature. The temperature can be set
    between 34oC – 37.5oC. The heater power is automatically adjusted
    to compensate for varying physiological and environmental
    conditions.
  • PREWARM MODE: This mode is used to provide
    general alarm-free prewarming of the heater and environment above
    the warmer mattress. The temperature cannot be set in this
    mode.
  • MANUAL MODE: This mode is used to provide
    user-adjustable heater power and the option to monitor the baby’s
    skin temperature using skin sensor. The heater power can be set
    between 0-100%. If the patient skin temperature is more than 1oC
    above the set point, then the heater is OFF and the output value
    displayed is 0 (zero).

The Wallaby Warmer also includes timers to help keep track of
time. The APGAR Alarm goes off at 1, 5, and 10 minutes. The
PROCEDURE TIMERS Alarm goes off at the end of a preset period from
1 to 99 minutes.

If an alarm goes off, refer to the table in the user manual to
determine the cause and effect of the alarm. The mute button can be
used to silence audible alarms only. If there is a software or
hardware failure, refer to the Service Manual for more information
on error codes.

During an X-ray procedure, follow the steps outlined in the user
manual to ensure the safety of the baby and the device. When
turning off the Wallaby Warmer, press and hold the OFF symbol for 3
seconds.

Wallaby
warmer

Quick Reference Guide

Refer to User Manual for comprehensive information

Modes of operation
Touch MODE Selection panel to enter MODE Selection screen

BABY MODE Set temperature: 34oC – 37.5oC

PREWARM MODE Set temperature: not available

MANUAL MODE Set heater power: 0-100%

In BABY MODE, the warmer provides stable control of the baby’s skin temperature by automatically adjusting the heater power to compensate for varying physiological and environmental conditions.

In PREWARM MODE, the warmer provides general alarm free prewarming of the heater and environment above the warmer mattress.

In MANUAL MODE, the warmer provides user-adjustable heater power and the option to monitor the baby’s skin temperature using skin sensor.

If the patient skin temperature is more than 1oC above the set point, then the heater is OFF and the output value displayed is 0 (zero).

Timers
APGAR Alarm at 1, 5 and 10 minutes. Switches off after 60 minutes.

PROCEDURE TIMERS Alarm at the end of preset period from 1 to 99 minutes.

Alarms

Alarm
HIGH SKIN TEMP
LOW SKIN TEMP
SENSOR DISCONNECT
POWER FAILURE

Mode
BABY
MANUAL
PREWARM BABY
MANUAL PREWARM BABY
MANUAL PREWARM ALL

Cause
Measured skin temperature higher than set temperature by 1oC
Measured skin temperature higher 38oC
Not active

Effect
Heater power disabled
Heater power disabled
Not active

Mute Button
Silences audible alarm only
Silences audible alarm only
Not active

Measured skin temperature lower than set temperature by 1oC
Not active Not active Skin sensor unplugged or faulty

Heater reduced to 25% of power
Not active Not active Heater reduced to 25% of power

Cancels alarm and restores heater power
Not active
Not active
Silences audible alarm only

Not active Not active Power supply to Warmer failed

Not active Not active Warmer disabled

Not active
Not active Silences audible alarm only

ERROR

ALL

Software or hardware failure

Warmer disabled Silences audible

(refer to Service Manual for more

alarm only

information on error codes)

When available, the mute alarm period is 10 min.
X-ray procedure
1. Turn the main switch of the device off and silence the power failure alarm 2. Wait until the device cools down 3. Lift the overhead unit up 4. Move overhead unit left or right 5. After X-ray procedure is completed return the overhead unit to center position 6. Make sure the overhead unit is locked in center postion 7. Turn the device on
Turning off the Warmer

2

1

3

The action will activate POWER FAILURE alarm.
V.1.03.EN

Press and hold the OFF symbol for 3 seconds
Wallaby Warmer. Quick Reference Guide.