Galaxy Medical Centauri Gti-00022-03 Electrosurgical System User Manual

CENTAURI GTI-00022-03 Electrosurgical System

Product Information

The CENTAURITM Connect Model GTI-00022-03 is a component of the
CENTAURI System, an electrosurgical system used in conjunction with
cardiac ablation catheters to deliver energy to cardiac tissue. It
is used to connect a catheter to the CENTAURI Generator as well as
to an electroanatomic mapping (EAM) system and EP recording system
amplifiers. The CENTAURI Connect isolates the energy output of the
CENTAURI Generator from the EP recording and EAM systems. It should
not come in contact with the patient and therefore is not provided
sterile.

The CENTAURI Connect comes with a power cord and instrumentation
cables for powering the CENTAURI Connect and interconnecting
various instrumentation signal lines of the System. The
interconnect power and instrumentation cables have been specified
and tested for use with the CENTAURI System and should not be
substituted with any non-Galaxy Medical specified cables.

Product Usage Instructions

Before using the CENTAURITM Connect Model GTI-00022-03, please
refer to the CENTAURI Generator User Manual for full information
and instructions on the CENTAURI Generator.

WARNING: Do not substitute cables, equipment, or disposables
with any cables, equipment, or disposables not provided or
specified by Galaxy Medical, Inc. Doing so could potentially damage
the system or cause injury.

To use the CENTAURI Connect, follow these steps:

  1. Connect the catheter to the CENTAURI Generator (see the
    CENTAURI Generator User Manual).
  2. Connect the electroanatomic mapping (EAM) system and EP
    recording system amplifiers to the CENTAURI Connect.
  3. Ensure that the CENTAURI Connect is powered on by connecting it
    to AC mains using the power cord provided.
  4. Follow the instructions for use provided in the CENTAURI
    Generator User Manual for delivering energy to cardiac tissue.

NOTE: The CENTAURI Connect is not intended to come in contact
with the patient and should not be provided sterile.

CENTAURITM Connect Model GTI-00022-03
User Manual
LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

Copyright © 2022 Galvanize Therapeutics, Inc. All rights reserved. CENTAURI is a trademark of Galvanize Therapeutics, Inc. Patents pending in the United States and internationally. GALAXY
Medical is a trademark of Galvanize Therapeutics, Inc.

Page 2 of 25

LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

Table of Contents
TABLE OF CONTENTS
1. Introduction …………………………………………………………………………………………………………. 5 2. Intended Purpose and Training ………………………………………………………………………………. 8 3. Warnings and Precautions …………………………………………………………………………………….. 9 4. System Connections and Receptacles …………………………………………………………………….11 5. System Setup ……………………………………………………………………………………………………..13 6. System Operation ………………………………………………………………………………………………..15 7. Cleaning and Maintenance ……………………………………………………………………………………17 8. Technical Specifications………………………………………………………………………………………..19 Appendix A ……………………………………………………………………………………………………………….23

LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

Page 3 of 25

Introduction

List of Symbols
The following is a list of symbols used on the CENTAURI Connect and throughout this user manual.

Catalogue number

Defibrillator proof type CF applied part

Serial number

Electric shock hazard

Follow instructions for use

Foot switch

Manufacturer

Digital signal cable connection

Date of manufacture

Do not use if packaging is damaged

Alternating current
Power On (Connect to AC mains) Power Off (Disconnect from AC mains)
Potential equalization terminal
Fuse information

Keep dry during shipping
Humidity limits during transportation
Temperature limits during transportation Atmospheric pressure limits during shipping
Mapping system connections

Caution

Catheter connections

Warning

Electrogram output connection

WEEE compliant CE marking Importer

Medical device
Authorized representative in the European Community
Swiss Authorized representative

Page 4 of 25

LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

Introduction
1. Introduction
System Overview The CENTAURITM System is an electrosurgical system used in conjunction with cardiac ablation catheters to deliver energy to cardiac tissue. The CENTAURI System is comprised of the CENTAURI Generator (“Generator”), the CENTAURI Connect, and the IVY Cardiac Monitor (“Cardiac Monitor”). This user manual describes the use of the CENTAURI Connect (GTI00022-03). For full information and instructions on the CENTAURI Generator, see the CENTAURI Generator User Manual.

Application Specification Summary
The CENTAURI System is intended to be used by a healthcare professional within an Electrophysiology (EP) Laboratory to treat patients in need of conventional cardiac ablation procedures. A cardiac ablation catheter is connected to the CENTAURI System and positioned at the target ablation site within the heart. The user presses and holds the foot switch to deliver energy to the target tissue resulting in tissue ablation. The user then repositions the catheter at subsequent target ablation sites and the process is repeated until all targeted tissue is treated.

LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

Page 5 of 25

Introduction CENTAURI Connect (GTI-00022-03) CENTAURI Connect is used to connect a catheter to the CENTAURI Generator (see the CENTAURI Generator User Manual) as well as to an electroanatomic mapping (EAM) system and EP recording system amplifiers. CENTAURI Connect isolates the energy output of the CENTAURI Generator from the EP recording and EAM systems. The Connect is not intended to come in contact with the patient and therefore is not provided sterile.
Interconnect power and instrumentation cables The CENTAURI Connect is provided with a power cord and instrumentation cables for powering the CENTAURI Connect and for interconnecting various instrumentation signal lines of the System. The interconnect power and instrumentation cables have been specified and tested for use with the CENTAURI System and should not be substituted with any non-Galaxy Medical specified cables.

Page 6 of 25

LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

Introduction

Other Devices Used with the CENTAURI Connect
WARNING: Do not substitute cables, equipment, or disposables with any cables, equipment, or disposables not provided or specified by Galaxy Medical, Inc. Doing so could potentially damage the system or cause injury.
The CENTAURI Connect is designed to work with the devices listed in the following table.

Device Cardiac Ablation Catheter
Cardiac Ablation Catheter Cardiac Ablation Catheter
Electrophysiology Cable Interface Cable
Interface Cable
Interface Cable

Manufacturer Abbott
Biosense Webster Boston Scientific
Abbott (St. Jude Medical) Biosense Webster Boston Scientific Galaxy Medical

Manufacturer Reference
TactiCathTM Contact Force Ablation Catheter, Sensor Enabled A-TCSE-XX*
THERMOCOOL SMARTTOUCHTM Catheter D13360X** or D13270X**
INTELLANAV STABLEPOINTTM Ablation Catheter M004 ERFSDS9620 0 or M004 ERFSDS9620K2 0
Electrophysiology Cable 1804-S IBI-85809
CARTO3 System Interface Cable CR3434CT
INTELLANAV STABLEPOINT Catheter Cable M004 RARC03 0
Compatibility Cable, BW ThermoCool ST GTI-00038

*Applicable Abbott product codes include A-TCSE-D, A-TCSE-F, A-TCSE-J, A-TCSE-DD, A-TCSE-DF, A-TCSE-FF, A-TCSE-FJ, and A-TCSE-JJ.

**Applicable Biosense Webster product codes include D133601, D133602, D133603, D132701, D132702, D132703, D132704, D132705.

LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

Page 7 of 25

Introduction

Compatible Mapping Systems
The CENTAURI Connect is designed for use with the cardiac mapping systems listed in the following table.

System CARTO® 3 System EnSite Cardiac Mapping System RHYTHMIA HDxTM Mapping System

Manufacturer Biosense Webster Abbott Boston Scientific

While the foot switch is depressed, ablation catheter visualization and contact force measurement will be unavailable. In addition, the cardiac mapping system may display warnings (e.g., connection to the ablation catheter is lost). If warnings on the cardiac mapping system do not resolve when the foot switch is released, refer to the user manual of the respective mapping system.

When the CENTAURI System is used with a mapping system, an increase to contact force measurement may be observed within published manufacturer tolerances.

Page 8 of 25

LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

Intended Purpose and Training
2. Intended Purpose and Training
Intended Purpose The CENTAURI System is intended for use in intracardiac ablation procedures with commercially available compatible ablation catheters for the treatment of atrial fibrillation. Refer to the CENTAURI Generator User Manual and the instructions for use (IFU) supplied with the cardiac ablation catheter to be used with the CENTAURI System.
Training The setup of the CENTAURI System shall be performed by qualified Galaxy Medical personnel. The CENTAURI System may be used only by medical personnel trained and experienced in the techniques of electrophysiology. Before using the CENTAURI System, review this User Manual thoroughly and completely. For further information, please contact Galaxy Medical, Inc.

LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

Page 9 of 25

Warnings and Precautions

3. Warnings and Precautions

Warnings and Precautions
THIS USER MANUAL SHOULD BE READ IN CONJUNCTION WITH THE IFU OF A COMPATIBLE CATHETER AND THE CENTAURI GENERATOR USER MANUAL BEFORE USING THE CENTAURI SYSTEM. FAILURE TO FOLLOW INSTRUCTIONS OR FAILURE TO HEED WARNINGS OR PRECAUTIONS MAY RESULT IN HARM TO THE PATIENT.
1. Hazardous electrical output. The CENTAURI System is for use only by qualified medical personnel trained in the use of this equipment.
2. Do not use the CENTAURI System if there are any signs of damage to any of its components. Visually inspect all accessories and verify cables are not damaged.
3. Do not substitute cables, equipment, or disposables with any cables, equipment, or disposables not provided or specified by Galaxy Medical, Inc. Doing so could potentially damage the system or cause injury.
4. Do NOT connect the catheter to the front panel of the CENTAURI Connect when powering on the CENTAURI Generator and preparing it for use. If the catheter is connected to the CENTAURI Connect while preparing the device for use, injury to the operator or the patient could happen.
5. Do not use sharp towel clips or metal instruments to attach cables to drapes. Sharp metal clips can damage electrical cables or provide an unwanted point of contact with the patient’s skin. Overlapping electrical wire around a metal clip creates an electrical transformer that can cause a hazard and may ignite drapes.
6. Cables should never be wrapped around metal instruments, as the current running through them can pass into the metal instrument and may result in harm to the patient and/or operator.
7. Do not use steam or heat sterilization to clean any system component. Do not soak any component in disinfectants or fluids. Do not allow liquid to enter into any of the electrical connections or the interior of any component.
8. To avoid the risk of electric shock, the CENTAURI System must only be connected directly to AC supply mains with protective earth.
9. Do not disassemble the system. Modification of this equipment is not permitted as serious injury to the operator or damage to the unit may result.
10. Electric shock hazard. Do not remove the cover of any equipment or any components from the equipment. Refer servicing to qualified personnel. There are no user-serviceable parts inside the CENTAURI Connect or any of the components of the CENTAURI Connect.
11. Position the connection cables of the ablation catheter in such a way that they do not touch either the patient or other cables. Keep active electrodes that are temporarily not in use at a safe distance from the patient.

Page 10 of 25

LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

Warnings and Precautions
12. Interference produced by the operation of the CENTAURI System may adversely influence the operation of other electronic equipment. For patients with cardiac pacemakers or other active implants, interference with the action of the active implant may occur, or the active implant may be damaged. In case of doubt, consult the manufacturer of the device. See the Technical Specifications section for Warnings related to Electromagnetic Compatibility.
Note: Any serious incident that occurs in relation to the device should be reported to the manufacturer and the competent authority of the Member State in which the user and/or patient is established.

LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

Page 11 of 25

System Connections and Receptacles
4. System Connections and Receptacles
CENTAURI Connect

Front

1. MAPPING: Receptacle for connecting the CENTAURI Connect to the compatible mapping system equipment.

2. EGM: Receptacle for electrophysiology cable to provide electrogram (EGM) signals to the mapping and EP recording systems

3.

CATHETER (Type CF Applied Part): Connection for compatible

catheters.

4.

INPUT: Connection to PEF output of the CENTAURI Generator

5.

Foot Switch: Receptacle for connecting the foot switch, which is used to

activate and deactivate energy delivery.

Page 12 of 25

LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

System Connections and Receptacles

Rear

6.

Digital Signal Connection: Receptacle for connecting the digital signal

cable between the Generator and CENTAURI Connect.

7. Power Switch: Switch for turning the CENTAURI Connect on and off

8.

Fuse Access Panel: Provides access to replaceable fuses

9. Power Cable: Receptacle to connect the Generator to an AC mains power outlet

10.

Potential Equalization Terminal: Connection that provides a common

electrical ground for other electronic devices used during the procedure

CENTAURI Generator Refer to the user manual provided with the CENTAURI Generator for a description of receptacles and connections.
Cardiac Monitor Refer to the user manual provided with the Cardiac Monitor for a description of receptacles and connections.

LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

Page 13 of 25

System Setup
5. System Setup
Prior to initial use of the CENTAURI System, the system should be unpacked and setup by qualified Galaxy Medical personnel according to the typical setup as shown in the System Connection Diagram.
WARNING: Do not use the CENTAURI System if there are any signs of damage to any of its components. Visually inspect all accessories and verify cables are not damaged. WARNING: Do not substitute cables, equipment, or disposables with any cables, equipment, or disposables not provided or specified by Galaxy Medical, Inc. Doing so could potentially damage the system or cause injury.

Page 14 of 25

LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

System Connection Diagram

System Setup

System Power-on Self-test 1. Complete the Pre-Procedure Power-On Sequence as specified in the System Operation section to confirm the system has been set up correctly.
Electrical Safety Inspection As applicable, follow any site and / or country specific requirements for equipment qualification for use within the facility prior to use. See the Electrical Safety Inspection section for recommended guidance and testing specific to on-site electrical safety testing.

LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

Page 15 of 25

System Operation

6. System Operation

Pre-Procedure Preparation and Power-On Sequence
Complete the following steps prior to preparing the patient for the cardiac ablation procedure and any time the system is turned off during a procedure.
WARNING: When powering on the CENTAURI Generator and preparing it for use, do NOT connect the catheter to the CENTAURI Connect. If the catheter is connected to the Generator while preparing the device for use, injury to the operator or the patient could happen.
WARNING: Do not use sharp towel clips or metal instruments to attach cables to drapes. Sharp metal clips can damage electrical cables or provide an unwanted point of contact with the patient’s skin. Overlapping electrical wire around a metal clip creates an electrical transformer that can cause a hazard and may ignite drapes.
WARNING: Cables should never be wrapped around metal instruments, as the current running through them can pass into the metal instrument and may result in harm to the patient and/or operator.

1. See the CENTAURI Generator User Manual for power on instructions of the Generator.

2. Power on the CENTAURI Connect module by turning the power switch on the rear

panel to the On illuminated.

position. Confirm the status light on the front panel is

3. Connect the PEF output of the CENTAURI Generator to the PEF input of the CENTAURI connect.

4. Connect the CENTAURI Connect digital signal output to the CENTAURI Generator.

5. Connect the footswitch to the front panel of the CENTAURI Connect.

6. Connect the cardiac ablation catheter to the CATHETER input of the CENTAURI Connect.

7. Connect the electrophysiology cable from the CENTUARI Connect EGM output to the input of the EP recording system being used.

8. Connect the CENTAURI Connect MAPPING output to the mapping system being used.

Page 16 of 25

LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

System Operation
Post Procedure Perform the following steps after the procedure is complete.
1. Disconnect the catheter from the CENTAURI Connect module. 2. Switch the power toggle on the rear of CENTAURI Connect to the Off position. 3. Refer to the Cleaning and Maintenance section for cleaning of the CENTAURI
Connect components.
For troubleshooting, see the Troubleshooting section in the CENTAURI Generator User Manual.

LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

Page 17 of 25

Cleaning and Maintenance

7. Cleaning and Maintenance

Cleaning
Clean components of the CENTAURI Connect per the instructions below.
CAUTION: Do not use steam or heat sterilization to clean any system component. Do not soak any component in disinfectants or fluids. Do not allow liquid to enter into any of the electrical connections or the interior of any component.

CENTAURI Connect
1. Disconnect the unit from the wall outlet before cleaning. 2. Wipe the exterior surface of the unit with 70 % to 90 % isopropyl alcohol or an
equivalent alcohol-based wipe. 3. Allow all surfaces and connections to dry before reconnecting.

Cables 1. 2.
3.

Disconnect all cables from the system components.
Wipe down all cables with 70 % to 90 % isopropyl alcohol or an equivalent alcoholbased wipe.
Allow cables to dry before reconnecting.

Page 18 of 25

LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

Cleaning and Maintenance

Maintenance Do not perform maintenance or service when the CENTAURI Connect is in use with a patient.

Routine Maintenance The CENTAURI Connect does not require routine maintenance or servicing.

Electrical Safety Inspection: EN 62353
Follow test instructions in the appropriate area of the standard. The acceptable limits (normal condition) for the CENTAURI System can be found in Table 1 of Appendix A.

Fuse Replacement
The CENTAURI Connect has two fuses that can be replaced in the field by qualified personnel. Replace the fuses with the same value fuses (see table below). Replacing with incorrect fuse rating could damage the Connect or create a fire hazard. To replace the fuses, open the fuse access panel on the back of the Connect, remove fuse housing with fuses, and replace the fuses. Reinstall the fuse housing and close the fuse access panel.

Component CENTAURI Connect

Replacement Fuses T 1.0A H 250V

Service and Returns Should service or repair be necessary, contact a Galaxy Medical representative.
End of Life Disposition Do not dispose of CENTAURI System components in the unsorted municipal waste stream. Follow local regulations for proper disposal.

LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

Page 19 of 25

Technical Specifications

8. Technical Specifications

Input Power
Dimensions Weight Rear Controls Connections
Classifications

100 V – 240 V ~, 50 Hz – 60 Hz universal power supply, 2 A input power rating. The power cord is used for removing AC mains power from the unit. CENTAURI Connect: 48.3 cm wide x 17.8 cm high x 45.7 cm deep CENTAURI Connect: 7.5 kg Line Power On/Off AC Line power, foot switch, catheter connector (x3), mapping connector (x3), pulsed field generator input, and electrogram output Class 1, Defibrillator Proof – Type CF, IPX0. Applicable general test and electrical safety requirements for Class 1 protection of ANSI AAMI ES60601-1, EN 60601-1, and CAN/CSA C22.2 No. 60601-1. Safety requirements of high-frequency surgical equipment of ANSI AAMI IEC 60601-2-2 and EN 60601-2-2. Electromagnetic compatibility (EMC) requirements of ANSI AAMI IEC 60601-1-2 and EN 60601-1-2.

Environmental Conditions
Temperature Humidity (non-condensing) Atm. Pressure

Transport or Storage -29 °C to +60 °C 30 % to 85 % 600 hPa to 1060 hPa

Operating 15 °C to 40 °C 30 % to 70 % 700 hPa to 1060 hPa

Replacement Part Numbers
WARNING: Do not substitute cables, equipment, or disposables with other cables, equipment, or disposables not provided by Galaxy Medical, Inc. Doing so could potentially damage the system or cause injury.

Component CENTAURI Connect

Part Number GTI-00022-03

Page 20 of 25

LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

Technical Specifications
Electromagnetic Compatibility (EMC)
The CENTAURI System has been verified to the electromagnetic compatibility requirements of ANSI AAMI IEC 60601-1-2 and EN 60601-1-2. This equipment uses non-ionizing radiation for treatment. Interference produced by the operation of high-frequency surgical equipment, such as the CENTAURI Generator, may adversely influence the operation of other electronic medical equipment such as monitors and imaging systems and special precautions should be taken. Portable and mobile RF communications equipment (e.g., cellular phones) can affect electronic medical equipment. The CENTAURI System should be setup and put into service according to the EMC information provided in this section. The CENTAURI System is suitable for use in all establishments other than domestic (i.e., residential) and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. Note: The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment might not offer adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or re-orienting the equipment. If any of the essential performance characteristics are lost or degraded due to electromagnetic disturbances, the user may see any of the following conditions:
· Distortion and / or lock-up of the generator front panel display screen · Distortion and / or lock-up of the cardiac synchronization monitor

Warnings
· The CENTAURI System is intended for use by healthcare professionals only. This device may cause radio interference and disrupt the operation of nearby equipment. It may be necessary to take mitigation measures, such as reorienting or relocating the CENTAURI System or shielding the location. Interference with electronic medical equipment such as monitors and imaging systems is usually resolved or minimized by rearranging the cables such that the unit cables do not overlap with the cables from the monitoring equipment.
· Use of the device adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to confirm that they are operating properly.
· Use of accessories and cables other than those specified and provided by Galaxy Medical, Inc. could result in increased electromagnetic emissions or decreased electromagnetic immunity of the device and result in improper operation.
· Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm to any part of the device including cables specified by the manufacturer. Otherwise, degradation of the performance of the device could result.

CENTAURI Connect Cables
Cable AC Power Cable Interconnect Cable, High Voltage Interconnect Cable, Digital Signal Contact Force Interconnect Compatibility Cable

Maximum Length 2.5 m 0.2 m 0.2 m 1.5 m

LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

Page 21 of 25

Technical Specifications

Guidance and Manufacturer’s declaration ­ Electromagnetic Emissions

The CENTAURI System is suitable for use in the electromagnetic environment specified below. The user should ensure that it is used in such an environment.

Emissions Test

Compliance

Electromagnetic Environment – Guidance

RF emissions CISPR 11 (EN55011)

Group 1

When Generator is in the STANDBY state, radio frequency [RF] energy is only used for its internal function. Therefore, its RF emissions are very low and are not likely to cause interference in nearby electronic equipment.
When Generator is delivering energy, the device must emit electromagnetic energy to perform its intended function. Nearby electronic equipment may be affected.

RF emissions CISPR 11 (EN55011)
Harmonic emissions EN 61000-3-2
Voltage fluctuations / flicker emissions EN 61000-3-3

Class A Class A Complies

The CENTAURI System is suitable for use in all establishments other than domestic (i.e., residential) and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

Conducted emissions may exceed limits specified in CISPR 11 by up to 14.2 dB when the system is connected to a compatible mapping system.

Guidance and Manufacturer’s declaration ­ Electromagnetic Immunity The information provided within this section is applicable to the CENTAURI System.

The CENTAURI System has been verified and found to be in compliance with the immunity standards at the test levels specified below. It is suitable for use in the electromagnetic environment specified below. For maintaining the Essential Performance of the device with regards to electromagnetic disturbances, the user must ensure it is used in such an environment.

Immunity test

Level Required

Level Tested

Electromagnetic environment ­ guidance

EN 61000-4-2 Electrostatic discharge (ESD)

±8 kV contact ±15 kV air

± 8 kV contact ± 15 kV air

Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.

EN 61000-4-3 Radiated, radiofrequency, electromagnetic field immunity

80 MHz – 2700 MHz, 3 V/m, 80 % 1 kHz AM

80 MHz – 2700 MHz, 3 V/m*, 80 % 1 kHz AM

Various per Table 9 of ANSI AAMI IEC 60601-1-2 and EN 60601-1-2.

As per Table 9 of ANSI AAMI IEC 60601-1-2 and EN 60601-1-2.

Radio frequency and electromagnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

EN 61000-4-4
Electrical fast Transient/burst

For AC / DC power ports: ±2 kV
±1 kV for signal ports

±2 kV for AC power ports

Equipment does not have

DC power ports therefore no Mains power quality should be that

test.

of a typical commercial or hospital

Equipment does not have environment.

cables > 3 m connected to

signal ports

Page 22 of 25

LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

Technical Specifications

The CENTAURI System has been verified and found to be in compliance with the immunity standards at the test levels specified below. It is suitable for use in the electromagnetic environment specified below. For maintaining the Essential Performance of the device with regards to electromagnetic disturbances, the user must ensure it is used in such an environment.

Immunity test

Level Required

Level Tested

Electromagnetic environment ­ guidance

EN 61000-4-5 Surge

For AC / DC power ports:
±1 kV differential mode ±2 kV common mode 1.2/50 µs

For AC power port:

±1 kV differential mode

±2 kV common mode 1.2/50 µs

Mains power quality should be that of a typical commercial or hospital

environment.

Equipment does not have

DC power ports therefore no

test.

EN 61000-4-6
Conducted, radiofrequency immunity

For AC / DC power ports, signal ports, and patient ports:
0.15 MHz – 80 MHz, 3 Vrms, 80 % 1 kHz AM, 6 Vrms in ISM/Amateur bands

For AC power, signal, and patient ports:

0.15 MHz – 80 MHz, 3 Vrms*, Radio frequency and

80 % 1 kHz AM, 6 Vrms in electromagnetic fields should be at

ISM/Amateur bands

levels characteristic of a typical

location in a typical commercial or

Equipment does not have hospital environment.

DC power ports therefore no

test.

EN 61000-4-8
Power frequency magnetic field

30 A/m

30 A/m

Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

EN 61000-4-11

Mains power quality should be that

Voltage dips, short interruptions, and voltage variations on power supply input lines,
UT = AC

>95 % dip in UT for 0.5 cycle
60 % dip in UT for 5 cycles
30 % dip in UT for 25 cycles
>95 % dip in UT for 250 cycles

>95 % dip in UT for 0.5 cycle
60 % dip in UT for 5 cycles
30 % dip in UT for 25 cycles
>95 % dip in UT for 250 cycles

of a typical commercial or hospital environment. If the user of the CENTAURI System requires continued operation during mains power interruptions, it is recommended that the CENTAURI System is powered from an

230V/50Hz

uninterruptible power supply.

EN 61000-4-39
Radiated fields in close proximity, magnetic field

134.2 kHz, 2.1 kHz pulse mod, 65 A/m
13.56 MHz, 50 kHz pulse mod, 7.5 A/m

134.2 kHz, 2.1 kHz pulse mod, 65 A/m
13.56 MHz, 50 kHz pulse mod, 7.5 A/m

Radiated magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

* The level tested exceeds the test level required by the standard. Testing was performed at a higher severity level to meet the requirements of other tests.

LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

Page 23 of 25

Technical Specifications

Appendix A

Electrical Safety Tests (Optional)
Only qualified individuals should perform electrical safety testing. Measurements should be made per local electrical safety standards.

Note: These tests are not required for planned preventative maintenance Table 1. Acceptable Safety Test Limits (normal condition) per EN 62353
Measurement Protective earth resistance (resistance between Mains Plug protective earth connector and protectively earthed accessible conductive parts) Earth (or Equipment) leakage current (NFPA99/120VAC operation)*
Earth (or Equipment) leakage current (all other AC operating voltages)** Enclosure (or Touch) leakage current
Patient (or Applied Part) leakage current
* Applicable only for United States. ** Applicable for all geographies outside of the United States.

Limit
0.3 ohm
0.3 mA (300 A) 0.5 mA (500 A) 0.1 mA (100 A) 0.01 mA (10 A)

Page 24 of 25

LBL-00085-001 Rev D, User Manual, CENTAURITM Connect, Model GTI-00022-03, 2022-06

ICON (LR) Limited South County Business Park Leopardstown, Dublin 18 D18 X5R3, Ireland
MedEnvoy Switzerland Gotthardstrasse 28 6302 Zug Switzerland
Galvanize Therapeutics, Inc. 1531 Industrial Road San Carlos, CA 94070, USA +1 (650) 268-4252
MedEnvoy Prinses Margrietplantsoen 33 – Suite 123 2595 AM, The Hague The Netherlands
MedEnvoy Switzerland Gotthardstrasse 28 6302 Zug Switzerland

Documents / Resouces

Download manual
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.


Related Manuals