General

1.1 Introduction
This user manual contains important information about the handling of the product. To ensure safety when using the product, read the user manual carefully and follow the safety instructions.
Invacare reserves the right to alter product specifications without further notice.
Before reading this document, make sure you have the latest version. You find the latest version as a PDF on the Invacare website.
If you find that the font size in the printed document is difficult to read, you can download the PDF version from the website. The PDF can then be scaled on-screen to a font size that is more comfortable for you.
In case of a serious incident with the product, you should inform the manufacturer and the competent authority in your country.

1.2 Compliance
Quality is fundamental to the company’s operation, working within the disciplines of ISO 13485.
This product features the CE mark, in compliance with the Medical Device Regulation 2017/745 Class I. The launch date of this product is stated in the CE declaration of conformity.
We are continuously working towards ensuring that the company’s impact on the environment, locally and globally, is reduced to a minimum.
We only use REACH-compliant materials and components.

1.3 Service Life
The expected service life of this product is three years when used daily and in accordance with the safety instructions, maintenance intervals, and correct use, stated in this manual.
The effective service life can vary according to frequency and intensity of use.
1.4 Warranty Information
We provide a manufacturer’s warranty for the product in accordance with our General Terms and Conditions of Business in the respective countries.
Warranty claims can only be made through the provider from whom the product was obtained.

1.5 Limitation of Liability
Invacare accepts no liability for damage arising from:

  • Non-compliance with the user manual
  • Incorrect use
  • Natural wear and tear
  • Incorrect assembly or set-up by the purchaser or a third party
  • Technical modifications
  • Unauthorized modifications and/or use of unsuitable spare parts

Safety

2.1 General Safety Information
Education, clinical judgment, and action-based planning based on vulnerability are fundamental factors in the prevention of pressure injuries.
A range of assessment scales can be used as a formal method of assessing risk from pressure injury development and should be used in conjunction with an informal assessment (informed nursing judgment). Informal assessment is considered to be of greater importance and clinical value.

WARNING!
Warning icon Risk of Serious Injury or Damage
Improper use of this product may cause injury or damage.
– If you are unable to understand the warnings, cautions, or instructions, contact a health care professional or provider before attempting to use this equipment.
– Do not use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as the user manual, service manual, or instruction sheet supplied with this product or optional equipment.

WARNING!
Warning icon Risk of Developing Pressure Injuries
Any object between the user and the pressure-reducing surface may result in the development of pressure injuries.
– Make sure that the support surface in contact with the user is kept free from crumbs and other food debris.
– Drip cables and other foreign objects must not become trapped between the user and the pressure-reducing surface.

CAUTION!
Warning icon Risk of Injury or Damage to Property
– Before use, check the product for any defects.
– Do not use a defective product.

2.2 Safety Information for Transport

  • Take care when handling the product to avoid any damage.
  • Avoid contact with jewelry, nails, abrasive surfaces, etc.
  • Avoid contact with walls, door frames, door catches or locks, etc.
  • Do not transport in roll cages unless completely protected from the sharp edges of the cage.

Product Overview

3.1 Product Description
The Invacare Matrx Contour Visco NG is a light anatomically contoured 100% viscoelastic (memory) foam cushion The cover is multi-stretch, water-resistant, and vapor-permeable.
It features an anti-slip base and carries a handle.
Certain sizes are available with an HR foam base layer.

3.2 Intended Use
This cushion is intended to be used in an armchair or wheelchair as part of an overall pressure injury prevention program of care.
There are no contraindications known.

3.3 Symbols on the Product

CE SYMBOLEuropean ConformityINVACARE Matrx Contour Visco NG - semblyMedical Device
EspenstrasseManufacturerINVACARE Matrx Contour Visco NG - sembly1Date of manufacture
microlife MT 1961 Digital Thermometer - ECEuropean representativemakita ML003G 1100LM 40V Max XGT Cordless LED Worklight - ICONRead user manual
INVACARE Matrx Contour Visco NG - sembly2Lot numberINVACARE Matrx Contour Visco NG - sembly3Maximum user weight
INVACARE Matrx Contour Visco NG - sembly4Machine wash (maximum temperature see label)INVACARE Matrx Contour Visco NG - sembly5Tumble dry low heat
INVACARE Matrx Contour Visco NG - sembly6Do not ironINVACARE Matrx Contour Visco NG - sembly7Do not dry clean
INVACARE Matrx Contour Visco NG - sembly8Do not bleachINVACARE Matrx Contour Visco NG - sembly9Avoid burning cigarette
INVACARE Matrx Contour Visco NG - sembly10Do not put near flameINVACARE Matrx Contour Visco NG - sembly11Do not pierce or cut
INVACARE Matrx Contour Visco NG - sembly12Orientation label (depending on the product)

Usage

4.1 Safety information

Warning icon WARNING!
Invacare advises that the patient is assessed by a trained healthcare professional to ensure the correct size and seating requirements. This is to achieve the best pressure and postural care and that position for the individual’s needs are met. – Always consult a healthcare professional before using the product.

Warning icon WARNING!
It is very important for the patient to reposition themselves, or to be repositioned, on a regular basis. This relieves pressure which helps prevent both tissue compression and potential pressure injury formation. Skin redness is one clinical indicator of skin breakdown.
– Regularly inspect/monitor if your skin develops any redness.
– Immediately contact a healthcare professional, if your skin develops redness.

INVACARE Matrx Contour Visco NG - sembly13 IMPORTANT!
Accidental cover damage allows fluid ingress and staining.
– Do not place sharp objects on the cover.
– When using patient transfer aids, ensure not to damage the cover. Check for sharp edges or burrs before use.
– Ensure that the cushion is not jammed or damaged by sharp edges when used on wheelchairs.
– Avoid burning cigarettes or open flames near the cushion.
– If pets are around, ensure that no claws puncture the cover.

INVACARE Matrx Contour Visco NG - sembly13 IMPORTANT!
The water-resistant cover allows fluids to roll off the surface, however with sustained periods of fluid contact the cover may absorb moisture. – Wipe away any fluid as soon as it is identified.

4.2 Using Cushion

  1. Remove all packaging before use.
  2. Place cushion on wheelchair or chair with the grey non-slip base facing down.INVACARE Matrx Contour Visco NG - Using Cushion
  3. Ensure that the cushion is placed in coordination with the orientation label, located on the left-hand side of the cushion.

Maintenance

5.1 Inspection
Check cushion (foam and cover) for strike-through (this may include fluid ingress, stains, rips, or damage) after the use of each patient and on a minimum monthly basis  depending on which occurs first) by a suitably qualified and competent person.
Check cushions

  1. Unzip the cover completely.
  2. Check for any staining on the interior foam.
  3. Replace any damaged cover or stained foam and dispose of it as per local authority procedure.

5.2 Cleaning and Disinfection
General Safety Information
CAUTION!
Risk of Contamination
– Take precautions for yourself and use appropriate protective equipment.

INVACARE Matrx Contour Visco NG - sembly13 IMPORTANT!
Wrong fluids or methods can harm or damage the product.
– All cleaning agents and disinfectants used must be effective, compatible with one another, and must protect the materials they are used to clean.
– Never use corrosive fluids (alkalines, acid, etc.), abrasive cleaning agents, or solvents (cellulose thinner, acetone, etc.). We recommend a mild detergent.
– Always make sure that the product is completely dried before taking it into use again.

For cleaning and disinfection in clinical or long-term care environments, follow your in-house procedures.

Cleaning Intervals
INVACARE Matrx Contour Visco NG - sembly13 IMPORTANT!
Regular cleaning and disinfection enhance smooth operation, increase the service life and prevent contamination.
Clean and disinfect the product
– regularly while it is in use,
– before and after any service procedure,
– when it has been in contact with anybody’s fluids,
– before using it for a new user.

Cleaning Instructions
Cleaning Cover
(Removal of contaminants such as dust and organic matter)

  1. Remove cover for laundering.
  2. Launder the cover using a diluted detergent solution (Instructions and maximum temperature on label).

Drying Cover

  1. Hang cover from a line or bar and drip dry in a clean indoor environment. or Tumble dry on a low heat setting.

INVACARE Matrx Contour Visco NG - sembly13 IMPORTANT!
– Tumble dry setting must not exceed 40 °C.
– Do not tumble dry for longer than 10 minutes.
– Dry thoroughly before use.

Disinfection Instructions
INVACARE Matrx Contour Visco NG - sembly13 IMPORTANT!
– Only use disinfectants and methods approved by your local infection control institution and follow your local infection control policy.

Disinfecting cover
(Reducing the number of microorganisms)
INVACARE Matrx Contour Visco NG - sembly13 IMPORTANT!
Failure in the disinfection process may result in the accumulation of reagents that could damage the polyurethane coating, react with the frame, or negate the biocompatibility results.
– Ensure that all cleaning agents, and disinfectants, are thoroughly rinsed off.
– Dry thoroughly before use.

  1. Wipe down the cover with a suitable detergent.
  2. Rinse the cover thoroughly with clean water using a single-use nonabrasive cloth.
  3. Dry the cover thoroughly.
Heavy Soilage
When badly soiled, we recommend cleaning with a diluted detergent solution at a maximum temperature in the washing machine.
– Clean up all spillages of bodily fluids i.e. blood, urine, feces, sputum, wound exudate, and all other bodily secretions as soon as possible using a suitable detergent.
– Large spillages of blood should be absorbed and removed with paper towels first, followed by as above.
When badly soiled, we recommend cleaning with a diluted detergent solution at a maximum temperature in the washing machine.
– Clean up all spillages of bodily fluids i.e. blood, urine, feces, sputum, wound exudate, and all other bodily secretions as soon as possible using a suitable detergent.
– Large spillages of blood should be absorbed and removed with paper towels first, followed by as above.

Replacing Cover

  1. Unzip the cover and remove it carefully from the foam core.
  2. Place the new cover onto the foam core.
    Ensure that the corners of the foam core are positioned correctly into the corners of the cover.
  3. Close the zipper. Where ever possible ensure the slider /sliders are positioned at the rear of the cushion.

Warning icon WARNING!
– In event of contamination, contact your hygiene specialist.
– Remove contaminated foams from use.

After Use

6.1 Storage

  • Store cushions in a dry environment.
  • Store cushions within a protective cover.
  •  Store items on a clean, dry surface free from sharp edges to avoid any possible damage.
  •  Never store other items on top of a cushion.
  • Do not store cushions next to radiators or other heating devices.
  • Protect cushions from direct sunlight.

6.2 Disposal
Be environmentally responsible and recycle this product through your recycling facility at its end of life.
Disassemble the product and its components, so the different materials can be separated and recycled individually.
The disposal and recycling of used products and packaging must comply with the laws and regulations for waste handling in each country. Contact your local waste management
company for information.

Technical Data

7.1 Dimensions and Weights

Total height [mm]100
Product weight [kg]1.5 – 2.5

User weight per size

Width [mm]Depth [mm]FirmnessUser Weight [kg]
380405Firm40 – 80
380430
380455
405405
405430
405455
Width [mm]Depth [mm]FirmnessUser Weight [kg]
430405Very Firm60 – 100
430430
430455
455405
455430
45545560 – 110
480405
480430
480455
50540560 – 120
505430
505455
35553060 – 100
40553060 – 120
45553060 – 130
50553060 – 140

7.2 Materials

Viscoelastic foam HR foam1Polyurethane foam components
CoverPU coated and Polyester fabric

1 Certain size only
All product components do not contain natural rubber Latex.

Notes

Belgium & Luxemburg:
Invacare nv
Autobaan 22
B-8210 Loppem
Tel: (32) (0)50 83 10 10
Fax: (32) (0)50 83 10 11
[email protected]
www.invacare.be
United Kingdom:
Invacare Limited
Pencoed Technology Park, Pencoed
Bridgend CF35 5AQ
Tel: (44) (0) 1656 776 222
Fax: (44) (0) 1656 776 220
[email protected]
www.invacare.co.uk
EspenstrasseInvacare UK Operations Limited
Pencoed Technology Park, Pencoed
Bridgend CF35 5AQ
UK
microlife MT 1961 Digital Thermometer - ECInvacare Portugal, Lda
Rua Estrada Velha 949
4465-784 Leça do Balio
PortugalINVACARE Matrx Contour Visco NG - sembly15Making Life’s Experiences Possible®