61061 Serenity XL Bariatric Low Air Loss Alternating Pressure Mattress System
Product Information
The product is the Selectis Alternating Pressure Pump and
Mattress, with model numbers 61061 and 61062.
Important Safeguards
- Ensure proper installation and operation of the product.
Environment Requirements
No specific information provided regarding environment
requirements.
Cleaning Guidelines
No specific information provided regarding cleaning
guidelines.
Maintenance
No specific information provided regarding maintenance.
Technical Specifications
Pump
- Item: Selec Serenity XL
- Power supply: 120V/60Hz, 1A
- Air output: 13 liter/min.
- Cycle: 10/15/20/25 min.
- Max firm mode Alarm: Low pressure/Power failure/Alternate
failure/Alternate/Static/Seat inflate mode - Size: 34cm(L) x 12.5cm(W) x 24.4cm(H)
- Weight: 3.5 kg
Mattress
- Item: Selec Serenity XL-61061 / Selec Serenity XL-61062
- Size: 80”(L) x 42”(W) x 10”(H) / 80”(L) x 48”(W) x
10”(H) - Top cover: Quilted bi-elastic cover with zipper
- Air cells: Nylon TPU (10” cell on cell structure with 6”
upper layer alternating and 4” lower layer static) - Base: Nylon / PVC 840D
- 20 air cells
- CPR valve for emergency procedures
- Transport function
- Quick connectors
- Pillow function
Product Usage Instructions
To use the Selectis Alternating Pressure Pump and Mattress,
follow these steps:
- Ensure the product is properly installed and connected to a
power supply. - Place the mattress on a suitable bed frame or surface.
- Ensure the top cover of the mattress is securely zipped.
- Connect the quick connectors of the mattress to the pump.
- Turn on the pump using the power switch.
- Adjust the cycle time according to your preference (10/15/20/25
min). - Select the desired mode: Max firm mode, Alarm mode, Alternate
mode, Static mode, or Seat inflate mode. - Monitor the pressure level and adjust as needed.
- For emergency procedures, use the CPR valve located on the
mattress.
If you encounter any issues or have questions, please contact
your local agent or dealer.
Tel: (888) 363-5563 Email:[email protected]
www.emeratldsupply.net
SELECTIS ALTERNATING PRESSURE PUMP AND MATTRESS USER MANUAL
Model 61061 & 61062
Important Safeguards
· · · · ·
· · · · · · · · · ·
1
Installa on & Opera ng Instruc ons
· Place pump on horizontal surface or hang it by means of the hook. · Open ma ress on the bed. Secure it with end flaps or straps. · Connect air feed tubes to air outlet on the ma ress and pump, ensuring that the
tubes are not “kinked” or twisted. Allow space for bed rails to drop freely. · Turn on the power. A er a beep sound, the pump will begin the opera on. · The pump will enter sta c mode ini ally when the power is turned on. The cycle
me is set at the 1st LED (10 min.), and the pressure level is set at the 3rd LED (30±2mmHg). First it will inflate the ma ress to 10-15±2 mmHg and then enter alternate mode. · During normal opera on, the pump will monitor pressure changes. If the pressure is below a user-defined pressure level, the pump will automa cally start to inflate the ma ress. The pump will stop when the user-defined pressure level is reached. · The first infla on of new ma ress: the alarm func on (visible and audible alarm) will ac vate if ma ress is not inflated completely within 40 minutes. · During pump is running, low pressure alarm will ac vate within 1 minute a er air tubes are disconnected or air cells are broken. · If there should be power failure, for example caused by power cord unplugged or power off when pump is working, the audible alarm will be ac vated and power failure indicator will flash. When power is back, pump will automa cally start working, do not need to press power bu on on the panel again. · If the ma ress stops alterna ng or alternate abnormally, the audible alarm will be ac vated and the alternate failure indicator will flash. · For nursing and caring convenience, press the max firm bu on to automa cally inflate the ma resses to the maximum level (65mmHg) for about 20 minutes. The pressure will return to alternate mode a er 20 minutes.
4
Environment Requirements
· Operaton Temperature: 50~95 (10~35) · Storage Temperature: 5~122 (-15~50) · Operaton Humidity: 20%~80% non-condensing · Storage Humidity: 10%~90% non-condensing
Cleaning Guidelines
· Can be cleaned by disinfectant solu on · Do not immerse the pump in water · Do not heat or steam in autoclave · The matress and cover are aerated un l they are fully dry
Maintenance
General
· Check the power cord and plug to see if there are abrasions or excessive wear. · Check the ma ress cover for signs of wear or damage. Ensure the ma ress cover and tubes
are connected together correctly. · Plug in the pump unit and check the airflow from the hose connec on port. The airflow should
alternate between ports every half-cycle me. · Check the air hoses to see if there are any kinks or breaks. For replacement, please contact
your local agent or dealer. · Make sure the ma ress tube is well connected. · Check the pump unit and make sure both power indicators are off when the switch is turned
off.
Low pressure
Examine if there is any air leakage between the pump and the ma ress connec ons or from the air ma ress tubes: · Check connectors between the air ma ress and pump. If there is any disconnec on, please
reconnect it. · Check the CPR Valves. Ensure their outlets are sealed. · Check the air-connec ng tubes. Ensure each single cell is not broken. · Set the pressure at Max firm. Keep the tubes fully inflated and inspect for air leakage. · Check if there is any air leakage from cells. Ensure no leakage occurs. If any leakage occurs,
please contact your local agent or dealer.
5
T
al Specifica ons
Pump
Ma ress
Item: Selec Serenity XL
Power supply: 120V/60Hz, 1A
Air output: 13 liter/min.
Pressure range: 20 mmHg 65 mmHg
Cycle 10/15/20/25 min. Max firm mode Alarm: Low pressure/Power failure/ Alternate failure Alternate/Sta c/Seat inflate mode Size: 34cm(L) x 12.5cm(W) x 24.4cm(H) Weight: 3.5 kg
Item: Selec Serenity XL-61061 Selec Serenity XL-61062
Size: 80”(L) x 42”(W) x 10”(H) 80”(L) x 48”(W) x 10”(H)
Top cover: Quilted bi-elas c cover with zipper Air cells: Nylon TPU (10” cell on cell structure with 6” upper layer alterna ng and 4” lower layer sta ) Base: Nylon / PVC 840D 20 air cells CPR valve for emergency procedures
Transport func on Quick connectors Pillow func n
6
2s 2
Tel:(888) 363-5563
Email: inf
aldsupply.net
www.emeraldsupply.net
7