Rebotec Tre-1860010 Fixi (small) And Fox (junior) Folding Pediatric Anterior Walker Instructions

Rebotec Tre-1860010 Fixi (small) And Fox (junior) Folding Pediatric Anterior Walker Instructions

Rebotec LOGOMaintenance Instructions
Rollators and Walkers Rebotec TRE1860010 Fixi Small and Fox Junio Folding Pediatric Anterior WalkerImages show a selection of rollators and walkers
Important advisory!
Store the maintenance inspection reports carefully. This will allow you to keep an
overview of the product’s maintenance status.read this guide

Maintenance of the rollator and walker

1.1 Maintenance / inspection requirements
In this maintenance instructions pamphlet, you’ll find information and advisories on the safe and professional maintenance and/or repair work on rollator and walker.
The products are Class I medical products as to the 93/42/EEC and the German MPG/Medical Products Act.
As to the requirements of the German MPBetreibV Medical Products Ordinance, the products must be maintained and the hygiene requirements must be adhered to. Check your local requirements.
The product’s lifetime is in large part affected by how it is used. The frequency of use and manner of use reduces the lifetime as does improper handling and inadequate maintenance.
In order to ensure safe use, a visual inspection and function check must be done at least once per year. Shorter time intervals between maintenance may be necessary when the frequency of use or the condition of the product exists due to safety reasons.
The operator has to carry out the maintenance on his own or must instruct a competent person (specialized supplier, medical supply store, or a medical-maintenance service) if he has not had the needed expertise. Inspection results must be documented and confirmed by a signature. The product is recommended with an inspection sticker that states the next inspection. The product operator is responsible for assuring regularly-conducted inspections.

1.2 Maintenance scope
The product’s inspection extends from the general condition to the individual components, function, and safety.

warning - 1 Necessary repair works must be done without delay. If immediate repair work is not possible, then it is in the patient’s interest and health to put the product out of service.

1.3 Maintenance advisories
When making repairs, use only original spare parts from REBOTEC.
Do not undertake any modifications or conversions to the product on your own. These will adversely impact the product’s safety and function. REBOTEC accepts no liability in such cases.
When performing maintenance and repairs, please use the product’s User Instructions booklet as an aid and follow the instructions concerning the proper use, re-use, lifetime, guarantee, care, and maintenance.
After completing maintenance or repair work on the product, follow the requirements for hygiene for re-use.

1.4 Maintenance reports and overview
In the appendix, you will find a form for documenting maintenance. You can make copies of the form. This allows you to keep an overview of the product’s maintenance status.

Rebotec TRE1860010 Fixi Small and Fox Junio Folding Pediatric Anterior Walker - ICONMaintenance protocol (single report) It is a form for proof of a single maintenance session.
Maintenance overview (total report) It is a form for proof of executed maintenance sessions.

1.5 Disposal
For disposing of the product, follow all the advisories in the product’s User Instructions booklet. Disposal must be done at a recycling company. Contact your local recycling company for more information.

1.6 Transport and storage
For transporting or storing the product follow the instructions of the particular user manual.
Advisory on other documents
At our www.rebotec.de website, you can download the User Instructions booklets, forms, and other information.

Maintenance protocol No.:

Product information / Rollator, Walker (medical product class I)
REBB02.11.149-17-06

Product
Serial no.Year ofUse
Manufacturer
User
Location of

Maintenance- General Condition

PChecksScope of workRemarksFinding*
1.Visual inspectionGeneral damage, wear, and corrosion
2.Connection elementsCompleteness, seat, function (screws, nuts, etc.)
3.FunctionsFolding mechanism
Height adjustment
Brake (permanent lock/normal brake)

Maintenance- Component checking

PArticleScope of workRemarksFinding
1.rollator frameDamages
2.grip bowFunction, fastening
3.swivel castorsFunction, damage, wear
4.brake wheelsFunction, damage, wear
5.brake leversFunction, damage, wear
6.hand switchFunction, damage, wear
7.wood seatStability, damage, wear
8.cushionDamage
9.accessoryFunction, wear
10.YANO-supportFunction, fastening
11.YANO-handleFunction, damage, wear
12.Castor brakesFunction, damage, wear
13.YANO-accessoryFunction, damage, wear

* Findings: I=inspected; E=exchanged’ R=repaired, i.o. = in order; n.i.o. = not in order

The following parts were replaced: (Use only original parts from the manufacturer)

Qty. Article description 

Maintenance results / summary

Rebotec TRE1860010 Fixi Small and Fox Junio Folding Pediatric Anterior Walker - Maintaince

Maintenance overview
Product information / Rollator, Walker (medical product class I) Rebotec TRE1860010 Fixi Small and Fox Junio Folding Pediatric Anterior Walker - ICON
REB02.11.149-17-06

Product
Serial no.Year manuf.Use start
Manufacturer
User
Location of useInventory no.
1st Maintenance DateMaintenance companyReportSignature
Remarks:
2nd Maintenance DateMaintenance companyReportSignature
Remarks:
3rd Maintenance DateMaintenance companyReportSignature
Remarks:
4th Maintenance DateMaintenance companyReportSignature
Remarks:
5th Maintenance DateMaintenance companyReportSignature
Remarks:

REBOTEC / WA-No: 02.11.152/2017-06