Gastec 102l Hexane Detector Tube Instruction Manual

Gastec 102l Hexane Detector Tube Instruction Manual

March 28, 2023

Detector tube No. 102L, Hexane, Instruction Manual Change
Gastec Corporation

Number of pump strokes and correction factor for will change, and also associated measuring range will change.

  • Details of the change to the Instruction Manual:

Before the change:

Measuring Range

4 – 50 ppm50 – 1200 ppm

Number of Pump Strokes

5

1

Correction Factor1/12

1

After the change:

Measuring Range

3.5 – 9 ppm*¹9 – 25 ppm25 – 50 ppm50 – 1200 ppm
Number of Pump Strokes842

1

Correction Factor

0.070.180.5

1

*1) If this tube is used with 8 pump strokes, it should be used in an environment with an absolute humidity of 7 mg/L (at 20-25℃, RH 30%) or less.

  • Contact Information

GASTEC CORPORATION
8-8-6 Fukayanaka, Ayase-City, Kanagawa, 252-1195, JAPAN
Tel: +81(0)467-79-3910 Fax: +81(0)467-79-3979