Apera Labsen 851-1 Viscous Ph Electrode Electrochemical Sensors User Manual

LabSen 851-1 Viscous pH Electrode Electrochemical Sensors

Product Information:

The LabSen 851-1 Viscous pH Electrode is a premium pH electrode
with manufacturing technology and key components imported from
Switzerland. It is specifically designed for use with cosmetics and
viscous samples. The electrode features an impact-resistant
membrane to prevent breakage during normal use. It also has a
conical low resistance membrane, Protelyte reference solution, and
a junction of 3 ceramic cores, making it suitable for measuring
viscous samples. The electrode’s blue gel inner solution does not
flow and avoids bubble formation. Additionally, the silver ion trap
reference prevents contamination of the junction when testing
samples containing sulfides and proteins.

Technical Data:

  • Measuring Range: Not specified
  • Temperature Range: Not specified
  • Shaft Material: Not specified
  • Membrane Shape: Conical
  • Reference Junction: Silver Ion Trap, Ceramic x 3
  • Electrolyte Soaking Solution: Not specified
  • Electrode Dimension: Not specified
  • Connector Cable: Not specified

Product Usage Instructions:

  1. Before measuring, unscrew the bottle cap, pull out the
    electrode, and rinse it with deionized water.
  2. After use, put the electrode back into the bottle and screw the
    cap tightly.
  3. Prior to measurement, remove the rubber plug to maintain
    pressure of the reference solution and ensure a consistent flow
    rate and stable potentials of the junction.
  4. When the reference solution level falls below half the height
    of the electrode, add Protelyte solution to the refilling hole
    using a syringe or pipette.
  5. Keep the electrode connector clean and dry. If it gets
    contaminated, clean it with medical cotton and absolute alcohol,
    then blow dry to prevent short circuits and slow reactions.
  6. Soak the measuring tip of the electrode in a soaking bottle
    containing Protelyte solution to keep the membrane hydrated and
    junction unblocked. Replace the storage solution if it becomes
    turbid and mildewed. Do not store the electrode in pure water or
    buffer solution for long periods.
  7. After 1 year of use, it is recommended to replace the electrode
    for optimal accuracy.

Warranty:

We provide a limited warranty for the LabSen 851-1 Viscous pH
Electrode. This warranty covers defects in material and workmanship
for a period of six months. APERA INSTRUMENTS will repair or
replace any malfunctioned or damaged product attributable to their
responsibility. The warranty does not cover damages caused by
transportation, storage, improper use, failure to follow
instructions, modifications, unauthorized repair, normal wear and
tear, or external causes beyond their control.

LabSen 851-1 Viscous pH Electrode User Manual
LabSen electrochemical sensors are premium pH electrode with manufacturing technology and key components imported from Switzerland. LabSen851-1 is suitable for cosmetics and viscous samples. This probe has following features: Impact-resist membrane (see the right picture), there is no danger of
electrode breakage during normal use. Electrode structure, which is suitable for viscous samples: conical low
resistance membrane, Protelyte reference solution, and junction of 3 ceramic cores. Blue gel inner solution, does not flow and will not cause bubble. Silver ion trap reference prevents the contamination of the junction when testing sample contains sulfides and proteins.

1 Technical Data Measuring Range Temperature Range Shaft Material Membrane Shape Reference Junction

(0 ~ 14) pH (0 ~ 100)°C Lead-free Glass
Conical Silver Ion Trap
Ceramic x 3

Electrolyte Soaking Solution Electrode Dimension
Connector Cable

Protelyte Protelyte (Ø12×120) mm
BNC Ø3×1m

2 Usage and Maintenance 2.1 When measuring, please unscrew the bottle cap, pull out the electrode and rinse it with
deionized water. After using, please put the electrode back into the bottle and screw tight the cap.

Before pulling out or putting back the electrode, make sure to fully loosen the blue cap on the storage sleeve so that the electrode can move in and out smoothly.

2.2 Prior to measurement, remove the rubber plug to maintain pressure of the reference solution, keep consistent flow rate of reference solution and stable potentials of junction.
2.3 After a period of usage, the reference solution will running low. Whenever the level falls below 1/2 height of the electrode, add Protelyte solution to the refilling hole by using a syringe or pipette.
2.4 The connector of the electrode should be clean and dry. If being contaminated, please clean it with medical cotton and absolute alcohol and blow dry to prevent the short circuit of the electrode and slow reaction of electrode.
2.5 The electrode’s measuring tip should be soaked in the soaking bottle containing Protelyte solution to keep the membrane hydrated and junction unblocked. Clean the bottle and replace the storage solution if the storage solution gets turbid and mildewed. The electrode should never be stored in pure water or buffer solution for long.
2.6 After 1-year of use, we recommend replacing the electrode to ensure the best accuracy.
3 Warranty We warrant this electrode free from defects in material and workmanship and agrees to repair or replace free of charge, at option of APERA INSTRUMENTS any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS for a period of six months. Warranty period is the time limit to provide free service for the products purchased by customers, not the service life of the tester or electrodes.
This limited warranty does not cover any damages due to: transportation; storage; improper use; failure to follow the product instructions or to perform any preventive maintenance; modifications; combination or use with any products, materials, processes, systems or other matter not provided or authorized in writing by us; unauthorized repair; normal wear and tear; or external causes such as accidents, abuse, or other actions or events beyond our reasonable control.
APERA INSTRUMENTS (Europe) GmbH Address: Wilhelm-Muthmann-Str.18, 42329 Wuppertal, Germany Phone: +49 202 51988998 Website: www.aperainst.de Email: [email protected]

References

Documents / Resouces

Download manual
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.


Related Manuals