Itsensor E2638-co Carbon Monoxide Detector-transmitter User Manual

Itsensor E2638-co Carbon Monoxide Detector-transmitter User Manual

itsensor logoitsensor E2638-CO Carbon Monoxide Detector Transmitter - puloraCarbon Monoxide Detector-Transmitter
E2638-CO
User Manualitsensor E2638-CO Carbon Monoxide Detector Transmitter -

Carbon Monoxide

Carbon Monoxide is a colorless, odorless, and tasteless gas that is highly toxic.
Synonyms: Carbonous oxide, Carbon (II) oxide, Flue gas, Monoxide.

Chemical formulaCO
Molar weight28
Relative gas density (to air)0.97
Conversion1 ppm = 1.15 mg/m3
Boiling point-191 .5 °C
Low explosive limit (LEL), % vol. in air13.
Upper explosive limit (UEL), % vol. in air74
OdorOdorless
HazardsHighly toxic. Mild poisoning causes lightheadedness, confusion, headache, dizziness, and flu-like effects.
Larger exposures can lead to toxicity of the CNS and heart, and death. After acute poisoning, long-term problems may occur. CO also has negative effects on a baby if exposed during pregnancy. Chronic exposure to low levels can lead to depression, confusion, and memory loss.
Exposure limits (NIOSH)TWA40 mg/m3 /35 ppm
IDLH1380 mg/m3 /1200 ppm

Conversion of ppm to mg/m³ is calculated for 25°C and 1 atm.

Specifications

Sensor typeElectrochemical cell
Sampling methodDiffusion
Detection range0…300 ppmI 0…1000 ppm
Maximum overload2000 ppm
Resolution / digital unit1 ppm
Response time T90< 30 s
Sensor lifetime> 10 years
Calibration interval12 months
Repeatability< t 5%
Signal updateEvery 1 second
Power supply12…36
12…36 VDC (default),  or 230 VAC as options
Power consumption< 2 VA
Analog outputs2 x 4-20 mA / 0-10 V, user settable
Load resistanceFk< (Us- 2V)/ 22 mA for 4-20 mA R. > 250 kOhm for 0-10 V mode:
Digital interfaceRS485, Modbus RTU protocol; No galvanic isolation
EnclosureGrey ABS, wall mount, protection class IP65
DimensionsH90 x W145 x D55 mm (overall dimensions H140 x W145 x D55 mm)
CE markingAccording to 2014/30/EU and 2014/35/EU, EN 50491-41:2012
EN 61000-6-3:2020, EN 61326-1:2013(EMC, emissions)
EN 61000-6-1:2019, EN 61000-6-2:2019(EMC, Immunity)
EN 60079-29-1:2016, EN 60079-29-2:2015 and EN 60079-29-3:201 4
Outputs assignmentOUT1 gas; OUT2 gasNOTE for LCD version only output 1 is available
Output scale width> 10 x resolution
Recommended: 20-100% of the detection range
Operating conditions-20…+50 °C, 0…90% RH non-condensing, 0,9…1,1 atm
Explosion-safe areas,
Non-aggressive atmosphere
NOTEI The sensor exhibits considerable sensitivity to acetylene, ethylene, hydrogen and nitric oxide
Relay option
Output relays2 x SPST relays (closing contact), 250 VAC/ 30 VDC, 5A max
Default alarm set-pointsRE1 (LOW): set 25; release 20 ppm
RE2 (HIGH): set 35; release 28 ppm
Signalling options
VisualRed and green LEDs
Acoustic alarmBuzzer 85 dB
LCD indicator option
Operating temperature0…+50 °C
Display dimensions72 x 36 mm
Number of digits3.5 7-segment
Character height14mm
Other featuresBacklight
Other options
Remote probeProtection IP65, default cable length 3.0 m; max height 80 mm, max diameter 65 mm

Product description

E2638 series detectors-transmitters belong to the PluraSens® family of multifunctional measurement instruments. The instruments utilize gas sensors of various types with excellent repeatability, stability, and long lifetime.
E2638 series provides two independent analog outputs OUT1 and OUT2, user-selectable to 4-20 mA or 0-10 V, proportional to gas concentrations. RS485 Modbus RTU digital communication interface allows easy instrument configuration and integration into various automation systems.
The following features are available as an option:

  • Two relays RE1 and RE2 with closing contacts. Relay outputs can be used to switch alarm sirens, ventilation fans, shut-off valves, or other actuators,
  • Acoustic (85 dB buzzer) and visual (red and green LED) signaling,
  • 24 VAC or 230 VAC as options for power supply module,
  • LCD indicator,
  • Remote probe.

The version of your device is marked on the package. If the symbol ⚠ is marked on the equipment, consult the documentation for further information.
Safety requirements
Misuse will impair the protection of the product. Always adhere to the safety provisions applicable in the country of use.
Do not perform any maintenance operation with the power on. Do not let water or foreign objects inside the device.
Removal of the PCB from the enclosure voids the warranty. Do not touch the electronic components directly, as they are sensitive to static electricity.
Connection diagrams can be found in the electrical connections section. The device might not perform correctly or be damaged if the wrong power supply is connected.
External circuits connected to the equipment should have sufficient insulation rating according to the environmental conditions and equipment power.
A disconnecting device that is marked as such and easily accessible should be included in the installation of this product.

Operating conditions

The device should be used both in a non-hazardous area and in a basic electromagnetic environment, where the latter is defined in EN 61326-1. Avoid strong mechanical shock and vibrations. Avoid corrosive atmosphere and areas highly contaminated with dust, oil mist, etc. Keep the instrument away from direct sunlight. A sudden temperature or humidity change might affect the sensitivity of the sensor.

Installation guidelines

Before proceeding with the installation it is mandatory to read the Safety requirements section and make sure to comply with all listed instructions. Installation standards EN
60079-29-2 and EN 45544-4 are also recommended for further instructions and related information about the installation. During the installation of the detector-transmitter the following points must be considered:

  • Application (air quality control or leakage detection)
  • Properties of the space under investigation (room geometry, direction, and velocity of airflows, etc.),
  • Carbon monoxide has practically the same density as air and spreads evenly in the room. Install the sensor in proximity to possible leakage or generation source, taking into consideration the temperature of the gas, the air currents in the room, etc.
  • The device should be accessible for maintenance and repair.

The aforementioned conditions above will affect the coverage area of the device. however, the coverage area for a transmitter is usually between 2.5 to 5 meters radius.
For early leakage detection install the sensor as close as possible to the potential leakage sources (flanges, valves, pressure reducers, pumps, etc.), taking into consideration other points listed above.
For general area monitoring without definite leakage sources, the detectors should be distributed evenly in the room.
For personal safety control, the detectors are installed in the breathing zone (at the height of the head of people or animals). The recommended sensor position is vertical, pointing downwards.
For installing the device, attach the device to the wall using provided screws through the side mounting holes or key slots (see mountains dimensions).

Mounting dimensions

itsensor E2638-CO Carbon Monoxide Detector Transmitter - Mounting dimensionsFront view

itsensor E2638-CO Carbon Monoxide Detector Transmitter - Front viewBottom view

Sensor probe handling

The E2638 series devices are available with a remote probe. The remote probe is connected to the main unit with a shielded cable. The default remote probe cable length is 3 m.itsensor E2638-CO Carbon Monoxide Detector Transmitter - probe handling

A )Wall mount remote probe with fixing clamp (default version) ,
B) Remote probe with rubber flange and three self-tapping screws (on request)
The sensor probes of all types are equipped with a hydrophobic microporous PTFE filter to protect the sensor from dust, dirt, and water drops. The filter should be replaced if it gets strongly contaminated. To replace the PTFE filter, unscrew sensor head cap and remove the old filter. Place a new filter into the cap and tighten it again.
NOTE! Never stab or press the filter near its center where the sensor is located since this may damage the sensor. Do not remove the filter as it may cause the device to show incorrect values and/or break the sensor.
The recommended orientation of the sensor probe is vertical with the sensor tip pointing downwards. This prevents the possible accumulation of condensed water on the sensor protection filter.

Gas sensor replacement procedures

  1. Remove the sensor head cap from the device (or the remote probe),
  2. Remove the PTFE filter (if it is not removed within the cap),
  3. Remove the O-ring rubber,
  4. Detach the electrochemical sensor from the device,
  5. Insert the new electrochemical sensor inside,
  6. Put back the O-ring rubber, PTFE filter, and the head cap, respectively.

Electrical connections

Unscrew four lid screws and detach the lid from the device. Use two M16 cable glands to let in the cables of the power supply and of the external devices. Without turning on the power, plug the power cable, and connect the analog/relay outputs and/or digital interface terminals to the necessary devices according to the relevant connection diagram.itsensor E2638-CO Carbon Monoxide Detector Transmitter - Electrical connections

Jumpers
J1OUT1 type (open: 4-20 mA; closed 0-10 V)
J2OUT2 type (open: 4-20 mA; closed 0-10 V)
X6Reset Modbus network parameters to default
X4 terminals
OUT14-20 mA / 0-10 V output
OUT24-20 mA / 0-10 V output
OV0 V / 24 VAC Neutral (optional)
ARS485 A / Data +
BRS485 B / Data –
+U+24 VDC / 24 VAC Phase (optional)
X5 terminals (optional)
L90.. 265 VAC Phase
N90…265 VAC Neutral
RE1 NORelay 1, normally open terminal
RE1 COMRelay 1, common terminal
RE2 NORelay 2, normally open terminal
RE2 COMRelay 2, common terminal

The screwless quick connect spring terminals on the E2638 series devices are suitable for a wide range of wires with a cross-section of 0.2…1.5 mm2. We recommend striping the wire end by 8…9 mm and using wire end sleeves.
To connect the wire, insert the wire end into the terminal hole. To disconnect, push the spring-loaded terminal lever, pull the wire out, and release the lever.
Use a twisted-pair cable, e.g. LiYY TP 2×2×0.5 mm2 or CAT 5, to connect the device to the RS485 network. A and B on the device represent DATA+ and DATA- respectively, polarity must be respected when connecting to an external RS485 network.
Both analog outputs can be independently changed between 4-20 mA and 0-10 V type using jumpers J1 (OUT1) and J2 (OUT2). By closing pins on a specific jumper the related output is 0-10 V, with an open jumper the output is 4-20 mA. Power restart is required after changing the position of the jumpers.

NOTE! The outputs are not galvanically isolated from the external power supply and share a common 0V. Allowed load resistance limits are stated in the Specifications table. To power the instrument from an external power source, connect terminals 0V and +U to the source.
If the integrated mains power supply module is used, connect terminals L and N to the mains.
NOTE! Actuator short-circuits should be avoided, to protect the instrument relays, use external fuses or safety switches.

Correct and incorrect cabling for 24 VACitsensor E2638-CO Carbon Monoxide Detector Transmitter - incorrect cabling

Operation

Turn on the power. The instrument warm-up time takes about 1 minute after switching on and the final sensor stabilization time to maximum accuracy takes about 30 minutes. The operating status is indicated by the LED on the PCB of the device. The control LED (red) response to different processes is presented in the following table:

ModeLED mode
During calibration mode or sensor heating period (if activated)0.5 Hz (50%  on, 50% off)
Relay 1 turned onBlinking 1 Hz (50% on, 50% off)
Relay 2 turned onBlinking 2 Hz (50% on, 50% off)
During Modbus communication cycleShort on-off pulses
Normal operating/measurementContinuously on or off

Make sure that the detector-transmitter is properly mounted, the external devices connected, power LED (green) on and control LED (red) constantly lit. Place the lid back and fix it with the screws. The device is ready to use.

Maintenance

Do not perform any maintenance operation with the power on.
Clean the device with a soft damp cloth. Do not use any abrasive cleaning agents. Do not immerse the device in water or any cleaning media.
Calibration
E2638-CO gas detector-transmitters have been calibrated by the Manufacturer with standard gas mixtures before delivery. Provided that the sensor is used under moderate conditions, the recommended interval for field recalibration is 12 months. Please contact your dealer for more information.

Delivery set

  • Detector-transmitter E2638
  • Mounting accessories:
    • 4 screws with plastic dowels
    •  Fixing clamp for remote probe version

itsensor E2638-CO Carbon Monoxide Detector Transmitter - Delivery set

Order code for E2638-CO options

E2638 optionsOrder code
Remote probe, 3 m cableE2638-CO-RP33-3
Remote probe, 10 m cableE2638-CO-RP33-10
Duct mount option, stem 035xL230 mmE2638-CO-DM
Integrated 24 VAC power supply moduleE2638-00-24VAC
Integrated 90…265 V mains power supply moduleE2638-00-230
Local LCD single value display 36 mm x 72 mmE2638-CO-LCD
Remote LCD single value display, wall mount box 115 x 65 x 40mm, 3 m cableE2638-CO-RLCD3
Two SPST relaysE2638-CO-R
Two LEDs (visual alarm)E2638-CO-LED
Buzzer (acoustic alarm)E2638-CO-B
Self-test button for relaysE2638-CO-TB
Integrated coreless micropumpE2638-CO-MP

Configuring

A standard configuration kit includes a USB-RS485 converter, fixed flow regulator, gas tubing with applicators, and a software pack. Please contact your Seller for more information.
Detectors-transmitters E2638 share all functionalities of the PluraSens® multifunctional platform. The features and options include:

  • Digital output change rate-limiting filter
  • Digital integrating (averaging) filter
  • Free assignment of each analog output to the chosen parameter
  • Flexible setting of analog output scales for each output
  • Output zero and slope adjustment for calibration
  • Free assignment of each of two relays to the chosen parameter
  • Several relay control logic modes
  • Switch delays and minimum on/off state durations for each relay

The output scales can be changed by Modbus commands by using the configuration software and the standard configuration kit (see Modbus RTU Communication).
Return to default settings
To reset the device’s Slave ID, baud rate, and stop bit numbers to factory settings, proceed as follows:

  1. De-energize the device
  2. Connect the X6 jumper
  3. Turn on the device
  4. De-energize the device
  5. Disconnect the X6 jumper
  6. Turn on the device

Modbus RTU Communication

RS485 communication interface

Databits: 8
Parity: none / odd / even
Stop bits: 1 or 2
Protocol: Modbus RTU
Supported Modbus functions:
03 – Read multiple registers
06 – Write a single register

Communication parameters

ParameterPermitted valuesDefault
Supported baud rates1200, 2400, 4800, 9600, 19200, 38400, 576009600
Data bits88
Paritynone / odd / evennone
Stop bits1, 21
ProtocolModbus RTU
Modbus functions03 – Read multiple registers
06 – Write a single register
Error codes01 – Illegal function
02 – Illegal data address
03 – Illegal data value
04 – Slave device failure
(details of last error 04 can be read from register 0x0008)

Modbus holding registers

Register addresses are shown 0-based, Address in hexadecimal, Reg in decimal format.
Modbus holding register numbers MHR are shown in the decimal 1-based format and may be addressed either from 00001 or 40001 base.

AddressReg / MHRRWDescriptionSupported values (dec)Default
0x00011 / 40002RHardware version
0x00022 / 40003RSoftware version
0x00033 / 40004RProduct serial number1…65535
0x00044 / 40005RWSlave ID (net address) *1…247 **1
 

0x0005

 

5 / 40006

 

RW

 

Baud rate *

1200, 2400, 4800, 9600, 19200,

38400, 57600

 

9600

0x00066 / 40007RWResponse delay, ms1…25510
0x00077 / 40008RWStop bits, parity bit *1 – No parity bit, 1 stop bit (default after factory reset) 2 – No parity bit, 2 stop bits 3 – Odd parity, 1 stop bit
4 – Even parity, 1 stop bit
NOTE: 3 and 4 are available starting from the Software version 0x218 (dec. 536)
 

 

 

1

0x00088 /40009RLast error code1…255
0x001117 / 40018RWTechnological: age of last data in seconds (read) / restart (write)0…65535 s (read), writing 42330 restarts the instrument 

0x00A5165 /40166RWZero adjustments for gas data, ADC-32000…+32000 ADC units0
0x00A6166 /40167RWSlope adjustment for gas data1…65535512
0x00A7167 /40168RWChange rate limit for gas units/ s1…32000, 0 – no limit0
0x00A8168 / 40169RWIntegrating filter time constant, s1…32000 (seconds), 0 – no filter0
AddressReg / MHRRWDescriptionSupported values (dec)Default
0x00C9201 /40202RWParameter tied to analog output 10 – None
2 – Gas concentration
9 – Forced Modbus control, value set in MHR / 40204
2
0x00CA202 /40203RWParameter tied to analog output 20 – None
2 – Gas concentration
9 – Forced Modbus control, value set in MHR / 40205
2
0x00CB203 /40204RWForced value for analog output 1***0…1000 (0.0%. 100.0% of output scale) 

0

0x00CC204 /40205RWForced value for analog output 2***0…1000 (0.0%. 100.0% of output scale)0
0x00D3211 /40212RWParameter tied to relay RE10 –none
2 – gas concentration 9 – control by Modbus control, state set in MHR / 40214
2
0x00D4212 /40213RWParameter tied to relay RE20 – none
2 – gas concentration 9- – control by Modbus control, state set in MHR / 40215
2
0x00D5213 /40214RWForced state for relay RE1***0 –off, 1 – on0
0x00D6214/ 40215RWForced state for relay RE2***0 – off, 1 – on0
0x00D7215 /40216RWSwitching delay for relay RE10…1000 (s)0
0x00D8216 /40217RWSwitching delay for relay RE20…1000 (s)0
0x00D9217 /40218RWMinimal on/off time for relay RE10…1000 (s)0
0x00DA218 /40219RWMinimal on/off time for relay RE20…1000 (s)0
AddressReg / MHRRWDescriptionSupported values (dec)Default
0x00DB219 /40220RWControl logic for relay RE1itsensor E2638-CO Carbon Monoxide Detector Transmitter - icon10 – none
1  – relay on at high values
2 – relay on at low values
3 – relay on at values within the range
4 – relay on for the values out of the range
0
0x00DC220 /40221RWControl logic for relay RE2itsensor E2638-CO Carbon Monoxide Detector Transmitter - icon20 – none
1 – relay on at high values
2 – relay on at low values
3 – relay on at values within the range
4 – relay on for the values out of the range
0
0x00DD221 /40222RWLOW setpoint for relay RE10…65535 (gas units)see Specifications
0x00DE222 /40223RWHIGH setpoint for relay RE1 

0…65535 (gas units)

 

see Specifications

0x00DF223 /40224RWLOW setpoint for relay RE20…65535 (gas units)see Specifications
0x00E0224 / 40225RWHIGH setpoint for relay RE20…65535 (gas units)see Specifications
0x00FF255 /40256RWSensor, analog outputs, LED and buzzer statusbit[0]=0/1 – sensor present/absent, RO bit[1]=0/1 – analog outputs deactivated/activated bit[2]=0/1 – in case the sensor is absent, turn signaling off/on analog output1 bit[3]=0/1 – in case the sensor is absent, turn on signaling with low current/high current on analog output1; if bit[2]=0 this bit will be ignored
bit[4]=0/1 – in case of sensor absent, turn signaling off/on analog output2
bit[5]=0/1 – in case of sensor absent, turn on signaling with low current/high current on analog output2; if bit[4]==0 this bit will be ignored bit[6]=0/1 – current/voltage output detected on output1, RO bit[7]=0/1 – current/voltage output detected on output2, RO bit[8]=0/1 – LED deactivated/activated
bit[9]=0/1 – buzzer deactivated/activated
bit[10]=0/1 – LED is on/off in normal condition bit[11]=0/1 – 1 Hz (50% on, 50% off) LED signal off/on if relay1 turned on
bit[12]=0/1 – 2 Hz (50% on, 50% off) LED signal off/on if relay2 turned on
 
0x0101257 /40258RRaw gas sensor data0…4095, ADC units 
0x0103259 /40260RGas concentration, gas units0…65535, gas units 
0x0105261 /40262RW0% value for analog output 1-32000…+32000, gas unitsUser defined
0x0106262 /40263RW100% value for analog output 1-32000…+32000 , gas unitsUser defined
0x0107263 /40264RW0% value for analog output 2-32000…+32000, gas unitsUser defined
0x0108264 /40265RW100% value for analog output 2-32000…+32000, gas unitsUser defined

* – The new value is applied after restart.
** – Broadcast slave ID 0 can be used to assign a new ID to the instrument with an unknown ID. When addressing by ID 0 the device shall be the only Modbus instrument in the network. The device will not respond to the Master command when addressed by ID 0.
*** – This value is dynamic and not kept in EEPROM after a restart.

Warranty

This product is warranted to be free from defects in material and workmanship for a period of one year from the date of the original sale. During this warranty period, the Manufacturer will, at its option, either repair or replace a product that proves to be defective. This warranty is void if the product has been operated in conditions outside ranges specified by the Manufacturer or damaged by customer error or negligence or if there has been an unauthorized modification.

PETTADORE AOE010 Air Eliminator Odour Remover - fig 11

References

Documents / Resouces

Download manual
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.


Related Manuals