Co-Dx COVID-R-001 Logic Smart COVID-19 RUO Product

Plate the Master Mix
Perform in Clean Area
- Load the Co-Dx Box software.

- In the Clean area, clean work surfaces with water > 10% bleach > water > 70% alcohol (in that order), or another equivalent method of cleaning that disinfects and degrades nucleic acids.

- Thaw the master mix (MM) and the nuclease-free water (NFW) (for the no-template control [NC]), on ice.

- Briefly vortex and centrifuge the MM and NFW reagents.

- Add 5 µL of MM into every well that will be used (1 well for NC, 1 well for positive control [PC], and 1 well for each sample).

- Pipet 5 µL of NFW into only the NC well and cap the NC well.

- Place reagents back into the freezer.

Plate Samples and Positive Control (PC)
- Load the Co-Dx Box software.

- In the Positive area, clean work surfaces with water > 10% bleach > water > and 70% alcohol (in that order).

- Thaw the PC on ice.

- Briefly vortex and centrifuge all reagents.

- Perform sample extraction, either using RNA extraction or lysis.

- Add 5 µL of the sample extract to the appropriate wells and cap the sample extract wells.

- Add 5 µL PC to the PC well and cap the PC well.

- Place capped tubes in the thermocycler, in appropriate locations and fill the rest of the carousel with blank tubes. (Optional): Cap all tubes and place samples and reagents back into the freezer.

Start the Run
- In the Co-Dx Box software, click New.

- Select an assay template and click OK.

- Under Run Setup, select Samples and type sample information in the appropriate order.

- Click Information and in the Notes field, type any needed information.

- Select the appropriate thermocycler and from the drop-down select Start Run.

- Verify the tube clamp is in place and the appropriate reaction volume is listed, and select Start. The run initializes and a progress bar gives an estimated time to completion.