MACHEREY-NAGEL Iron 1 Performing Colorimetric Tests Kit

Test kit for performing colorimetric tests on iron ions in surface water and sewage
Method
Colorimetric determination of Fe(II) and Fe(III) ions with a triazine derivate
Measurement range
0.04–1.0 mg/L Fe
Contents of test kit (refill pack)
sufficient for 200 tests
- 30 mL Fe-1*
- 2 screw-plug measuring glasses
- 1 slide comparator
- 1 color chart
- 1 plastic syringe 5 mL
- 1 instruction for use
Hazard warning:
Information regarding safety can be found on the box label and in the safety data sheet. You can download the SDS from www.mn-net.com/SDS.
Instructions for use
Also, refer to the pictogram on the back of the color chart
- Pour a 5 mL water sample into each of the measuring glasses using the plastic syringe. Place a measuring glass on position A in the comparator.
Only add the reagent to measuring glass B. - Add 5 drops of Fe-1, seal the glass, and mix.
- Open the glass after 3 min and place it on position B in the comparator.
- Slide the comparator until the colors match in the inspection hole on top. Check the measurement reading in the recess on the comparator reed. Mid-values can be estimated.
- After use, rinse out both measuring glasses thoroughly and seal them.
The reagents can also be used for the photometric evaluation with photometers PF-12 and PF-12Plus.
This technique can be used also for analyzing seawater.
Disposing of the samples:
Information regarding disposal can be found in the safety data sheet. You can download the SDS from www.mn-net.com/SDS.
Interferences
The following quantities of ions will not interfere: ≤ 5 mg/L Co2+, Ni2+, MoO42–; ≤ 3 mg/L Cu2+.
Conversion table
mg/L Fe | mmol/m3 |
| 0.04 | 0.7 |
| 0.07 | 1.3 |
| 0.10 | 1.8 |
| 0.15 | 2.7 |
| 0.20 | 3.6 |
| 0.30 | 5.4 |
| 0.50 | 9.0 |
| 1.0 | 18 |
Storage:
Store the test kit in a cool (< 25 °C) and dry place.
MACHEREY-NAGEL GmbH & Co. KG·
Neumann-Neander-Str. 6–8 · 52355 Düren · Germany Tel.: +49 24 21 969-0 · [email protected]· www.mn-net.com.