Macherey-nagel Mn 91306 Quantofix Sulfitt Test Strips Instructions

Macherey-nagel Mn 91306 Quantofix Sulfitt Test Strips Instructions

MACHEREY-NAGEL MN 91306 Quantofix Sulfitt Test Strips Instructions
MACHEREY-NAGEL MN 91306 Quantofix Sulfitt Test Strips Instructions

Pack content

1 aluminium container with 100 test strips

Measuring range:

Visually
10-1000 mg/L S032-

Reflectometrically
10-500 mg/L S032-

Color gradation:
0 • 10 • 25 • 50 • 100 • 250 • 500 • 1000 mg/L S032-

General indications

Remove only as many test strips as are required. Close the container immediately after removing a strip. Do not touch the test field.

Instructions for use

Instructions for use

  1. Dip the test strip into the sample (pH 6-9) for 1 s.
  2. Shake off excess liquid.
  3. Wait 20 s.
  4. Compare the test field with the color scale.
    If sulfite ions are present, the test field turns salmon-pink.
    Interferences:
    Acidic test solutions (i.e. white wine) must be buffered with solid sodium acetate prior to testing because the test paper does not react in acidic media. Sulfide ions also lead to a red coloration of the test paper in neutral solutions. This interference can be eliminated by adding an excess of nickel (II) ions. The nickel sulphide formed can be removed by filtration. Larger amounts of reducing agents such as hydroxylammonium chloride or ascorbic acid interfere with the determination leading to falsely low test results.

Disposal: Used test strips can be placed in the normal household waste.

Storage

Avoid exposing the strips to sunlight and moisture. Keep container cool and dry (storage temperature not above +30 °C). If correctly stored, the test strips may be used until the use-by-date printed on the packaging.

MACHEREY-NAGEL GmbH & Co. KG • Neumann-Neander-Str. 6-8 52355 Duren Germany Tel.: +49 24 21 969-0 • Fax: +49 24 21 969-199 • [email protected]www.mn-net.com

SYMBOLS

item number Item number

warning Warning

temperature range Storage temperature range

read instructions Please read instructions for use

Batch identification Batch identification

Use by Use by

dry place Store in a dry place

Keep container Keep container closed