Klinger E-st Ultrasonic Flowmeter User Manual

Klinger E-st Ultrasonic Flowmeter User Manual

KLINGER - logoE-ST Ultrasonic Flowmeter
User ManualKLINGER E-ST Ultrasonic Flowmeter

Notice Thank you for choosing Model E-ST
This instruction manual contains the important use and operation information of the flow meter. Please read carefully the reference manual before the operation to make your flow meter exert the best performance.
If you make a mistake there will be affected the meter’s working and reduce the meter’s life or cause some malfunctions.

Product component

Inspection should be made before installing the Flow meter. Check to see if the spare parts are in accordance with the packing list. Make sure that there is no potential damage to the enclosure due to a loose screw or loose wire, which occurred during transportation.
Any questions, please contact your representative as soon as possible.

KLINGER E-ST Ultrasonic Flowmeter - figure 1

Dimensions

ModelA
(mm)
B
(ma’)
C
(mm)
D(mm)
minmax
DN152510581/0187.5/022.3
DN202515581/0254/028
DN 2529.19.581/0324/035
DN 3230.24681/0389/045
DN403627781/0487/054
DN 504132911/5/19588/5/1964
DN 6547.401051/0657/074
DN8052.431191/07613/086

KLINGER E-ST Ultrasonic Flowmeter - figure 2

Installation and connect

E-ST needs to install the flow sensor and temperature sensor and clean the pipeline before Installation.
Make sure no dirt, paint, or other stains on the surface of the tube. Then put the bottom parts on the side of the pipe.

KLINGER E-ST Ultrasonic Flowmeter - figure 3When ‘E-ST is installed, the Energy meter is wired.
Connect the DC power and RS485 output.

Panel function

KLINGER E-ST Ultrasonic Flowmeter - figure 4

Powering on
As soon as the E-ST Energy meter is switched on, the self-diagnosis program will start to run.

KLINGER E-ST Ultrasonic Flowmeter - figure 5

Signal Quality (SQ value)

  • value is short for Signal Quality. It indicates the level of the signal detected.
  • value is indicated by numbers from 0-99 represents the minimum signal detected while 99 represent the maximum.

Normally, the transducer position should be adjusted repeatedly and the coupling compound should be checked frequently until the signal quality detected is as strong as possible.

Keypad Functions

Follow these guidelines when using the Flow meter keypad:

KLINGER E-ST Ultrasonic Flowmeter - figure 6

KLINGER E-ST Ultrasonic Flowmeter - icon 1 Setting or display mode, when it is setting mode, that can return to the previous menu, KLINGER E-ST Ultrasonic Flowmeter - icon 2andKLINGER E-ST Ultrasonic Flowmeter - icon 3 scroll up and down to select the menu, when press KLINGER E-ST Ultrasonic Flowmeter - icon 2 move to next digit, press()and the numbers scroll from 0 to 9, you can select the number. PressKLINGER E-ST Ultrasonic Flowmeter - icon 4 to confirm.

Window descriptions

Display Menu

  • When the power on, the meter will display Velocity/Net Totalize.
    KLINGER E-ST Ultrasonic Flowmeter - figure 7Display signal quality (SQ), time, heat power ( Eq) heat totalizer (EH), cold totalizer (EC)
  • Press KLINGER E-ST Ultrasonic Flowmeter - icon 2will display T1, T2, delta T, and press KLINGER E-ST Ultrasonic Flowmeter - icon 3 will return to the previous menu.
    KLINGER E-ST Ultrasonic Flowmeter - figure 8Display date, time, outlet temp. (T1), inlet temp.(T2), Delta temp. (DT)
  • Press KLINGER E-ST Ultrasonic Flowmeter - icon 2will display Eq, EH, press KLINGER E-ST Ultrasonic Flowmeter - icon 3 will return to the previous menu.
    KLINGER E-ST Ultrasonic Flowmeter - figure 9 Display signal quality (SQ), time, Heat power (GJIh), Heat totalizer (EH)
  • Press KLINGER E-ST Ultrasonic Flowmeter - icon 2will display Eq, EC, and pressKLINGER E-ST Ultrasonic Flowmeter - icon 3 will return to the previous menu.
    KLINGER E-ST Ultrasonic Flowmeter - figure 10
  • PressKLINGER E-ST Ultrasonic Flowmeter - icon 2 will display Flow rate, Net totalizer, press KLINGER E-ST Ultrasonic Flowmeter - icon 3 will return to the previous menu.
    KLINGER E-ST Ultrasonic Flowmeter - figure 11Display signal quality (SQ), time, flow rate, Net totalizer
  • PressKLINGER E-ST Ultrasonic Flowmeter - icon 2· will display the Unit runtime, pressKLINGER E-ST Ultrasonic Flowmeter - icon 3 will return to the previous menu.
    KLINGER E-ST Ultrasonic Flowmeter - figure 12Display Unit runtime, monthly heat totalizer (E HM), monthly energy totalizer (ECM), monthly flow totalizer (ETM)

Setup menu

PressKLINGER E-ST Ultrasonic Flowmeter - icon 1 will display Setup menu.
KLINGER E-ST Ultrasonic Flowmeter - figure 13

The following options are available (by KLINGER E-ST Ultrasonic Flowmeter - icon 2 or KLINGER E-ST Ultrasonic Flowmeter - icon 3 buttons)

0. Pipe parameter3. Output setting
1. System setting4. Energy setting
2. Calibration5. History Data

Setup Menu – pipe parameter

Press)
SelectKLINGER E-ST Ultrasonic Flowmeter - icon 2.Pipe parameter, then KLINGER E-ST Ultrasonic Flowmeter - icon 4 display:

KLINGER E-ST Ultrasonic Flowmeter - figure 14

The following options are available (by KLINGER E-ST Ultrasonic Flowmeter - icon 2·orKLINGER E-ST Ultrasonic Flowmeter - icon 2 buttons)
0. Outer diameter
1. Wall thickness
2. Material: Move ·KLINGER E-ST Ultrasonic Flowmeter - icon 2or KLINGER E-ST Ultrasonic Flowmeter - icon 2·can option PVC, Carbon steel, Steel, Copper pipe.
3. Fluid type: Move KLINGER E-ST Ultrasonic Flowmeter - icon 2or KLINGER E-ST Ultrasonic Flowmeter - icon 2an option Water, Sea Water, Oil…etc.

Setup Menu – System setting

PressKLINGER E-ST Ultrasonic Flowmeter - icon 2, Select 1. System setting, then KLINGER E-ST Ultrasonic Flowmeter - icon 4 display:

KLINGER E-ST Ultrasonic Flowmeter - figure 15

The following options are available (by(·)or(· buttons)
0. System unit: MoveKLINGER E-ST Ultrasonic Flowmeter - icon 2 orKLINGER E-ST Ultrasonic Flowmeter - icon 3 an option Metric, English.
1. Flow rate unit: MoveKLINGER E-ST Ultrasonic Flowmeter - icon 2 orKLINGER E-ST Ultrasonic Flowmeter - icon 3 can option m3Ih, LPM, GPM.
2. Total unit: Move·KLINGER E-ST Ultrasonic Flowmeter - icon 2or·can m3, L, GAL.
3. Totalize RESET: AII parameters are reset, PressKLINGER E-ST Ultrasonic Flowmeter - icon 4, move, KLINGER E-ST Ultrasonic Flowmeter - icon 2orKLINGER E-ST Ultrasonic Flowmeter - icon 3 arrow to select “YES” or “NO After “YES is selected.

Time set 

KLINGER E-ST Ultrasonic Flowmeter - figure 16

Generally, it is unnecessary to modify the date-time as the system is provided with a highly reliable perpetual calendar chip.

System lock

KLINGER E-ST Ultrasonic Flowmeter - figure 17

Once the system is locked, any modifications to the system are prohibited, but the parameter is readable. “Unlock” using your designated password. The password is composed of 1to 4 numbers.

System INFO

KLINGER E-ST Ultrasonic Flowmeter - figure 18

System INFO: Display the serial number (SN) of the meter. This SN is the only one assigned to each flow meter ready to leave the factory.
The factory uses it for file setup and for management by the user. Set zero: PressKLINGER E-ST Ultrasonic Flowmeter - icon 4; reset Zero Point” which was set by the user.
Manual Totalizer: The manual totalize is a separate totalize. Press KLINGER E-ST Ultrasonic Flowmeter - icon 4to start, and press KLINGER E-ST Ultrasonic Flowmeter - icon 4to stop it. It is used for flow measurement and calculation.

Display dir

KLINGER E-ST Ultrasonic Flowmeter - figure 19

Can choose the direction of the display, convenient to observe the measurement data.

Setup Calibration

PressKLINGER E-ST Ultrasonic Flowmeter - icon 2.Select 2. Calibration, and thenKLINGER E-ST Ultrasonic Flowmeter - icon 4 display:

KLINGER E-ST Ultrasonic Flowmeter - figure 20

0. Scale factor

Refers to the ratio between “actual value” and reading value”. For example, when the measurement Is 2.00, and It Is Indicated at 1.98 on the instrument, the scale factor reading is 2/1.98. This means that the best scale factor constant is1.01.

  1. Set zero: PressKLINGER E-ST Ultrasonic Flowmeter - icon 4; reset ‘Zero Point which was set by the user.
    KLINGER E-ST Ultrasonic Flowmeter - figure 22
  2. Low flow cut: The flow rate falls below the low flow cutoff value.
    KLINGER E-ST Ultrasonic Flowmeter - figure 23The flow indication is driven to zero. This function can prevent the flow meter from reading flow after a pump as shut down but there is still liquid movement in the pipe, which will result in a totalization error.
    Generally, 0.03m/s is recommended to enter as the low flow cutoff point.
    The low flow cutoff value has no relation to the measurement results once the velocity increases over the low flow cutoff value.
  3. Manual zero
    KLINGER E-ST Ultrasonic Flowmeter - figure 24

The seldom-used calibration method is suitable for experienced operators to artificially input an offset superimposed on the measured value
in order to obtain the true value when other calibration methods cannot be used well.
For example:
Actual measured value =250 m3/h
The offset valve =10 m3lh
Meter display =240 m3/h
In general, this value should be set: to “0”.

Setup Menu – Output

PressKLINGER E-ST Ultrasonic Flowmeter - icon 2.Select 3. Output setting, and then KLINGER E-ST Ultrasonic Flowmeter - icon 4display:

KLINGER E-ST Ultrasonic Flowmeter - figure 25

0. RS485 setup

KLINGER E-ST Ultrasonic Flowmeter - figure 26

The window is used to set the serial port. It connects with the equipment of its serja! port set of parameters must match.
MoveKLINGER E-ST Ultrasonic Flowmeter - icon 2orKLINGER E-ST Ultrasonic Flowmeter - icon 3 can option baud rate: 2400, 4800, 9600, 19200.
Data length fixed: 8; Stop bit for 1.
Factory serial port parameters for the default “9600, 8, None, 1.

  1. Alarm value(Option)
    KLINGER E-ST Ultrasonic Flowmeter - figure 27

Enter the low alarm value; any of the measured flow, which is lower than the low value, will activate the alarm in the OCT hardware or relay output signal.
Enter the high alarm value; any of the measured flow, which is higher than the high value, will activate the alarm in the OCT hardware or relay output signal.

Setup Menu – Energy setting

PressKLINGER E-ST Ultrasonic Flowmeter - icon 2, Select 4. Energy Setting, and then KLINGER E-ST Ultrasonic Flowmeter - icon 4display:

KLINGER E-ST Ultrasonic Flowmeter - figure 28

The following options are available (by· KLINGER E-ST Ultrasonic Flowmeter - icon 2orKLINGER E-ST Ultrasonic Flowmeter - icon 3 buttons)
0. Energy unit: MoveKLINGER E-ST Ultrasonic Flowmeter - icon 2 orKLINGER E-ST Ultrasonic Flowmeter - icon 3 an option: GJ, MBtu, KWh, MWh.

  1. Temp unit: MoveKLINGER E-ST Ultrasonic Flowmeter - icon 2 orKLINGER E-ST Ultrasonic Flowmeter - icon 3 an option: C or F
  2. Flow position: Move (·or·can option: Inlet, Outlet
  3. DT sensitivity: MoveKLINGER E-ST Ultrasonic Flowmeter - icon 2 orKLINGER E-ST Ultrasonic Flowmeter - icon 3, You can change the value
  4. RTD Calib: Temperature sensor calibration
    KLINGER E-ST Ultrasonic Flowmeter - figure 29

Setup Menu – History Data

PressKLINGER E-ST Ultrasonic Flowmeter - icon 2. Select 5. History Data, and thenKLINGER E-ST Ultrasonic Flowmeter - icon 3 display:
KLINGER E-ST Ultrasonic Flowmeter - figure 30

0. By Day
Display: Daily heat totalizer (EHD), Daily cold totalizer(ECD), Daily Flow totalizer (ETD)
KLINGER E-ST Ultrasonic Flowmeter - figure 31

  1. By Month
    Display: Monthly heat totalizer(EHM), Monthkt cold totalizer(ECM), Monthly Flow totalizer (ETM)
    KLINGER E-ST Ultrasonic Flowmeter - figure 32
  2. By Year
    Display: Year heat totalizer (EHM), Year cold totalizer(ECM), Year Flow totalizer (ETM)
    KLINGER E-ST Ultrasonic Flowmeter - figure 33

KLINGER - logoDenmark

KL I NGER
Danmark A/ S
Nyager 1 2- 1 4
DK- 2605
www.kling

References

Documents / Resouces

Download manual
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.


Related Manuals