DeVilbiss 525DS 5-Liter Oxygen Concentrator
DeVilbiss 525DS 5-Liter Oxygen Concentrator
WARNING
Under certain circumstances, oxygen therapy can be hazardous. Seek medical advice before using an oxygen concentrator.
Physician Information
- Physician Name: __________________________________________________________________
- Telephone: ______________________________________________________________________
- Address: ________________________________________________________________________
Prescription Information
- Name: _________________________________________________________________________
Oxygen liters per minute
- at rest:____________________ during activity:____________________ other: _____________
Oxygen use per day
- Hours:_____________________________________ Minutes: ___________________________
- Comments: ______________________________________________________________________
DeVilbiss 5-Liter Oxygen Concentrator w/OSD Serial Number: _____________________________
DeVilbiss Equipment Provider Information
- Set-Up Person: ___________________________________________________________________
This instruction guide was reviewed with me, and I have been instructed on the safe use and care of the DeVilbiss Oxygen Concentrator
- Signature:____________________________________________________ Date: ________________
SYMBOL DEFINITIONS
IMPORTANT SAFEGUARDS
Read this entire guide before using your DeVilbiss concentrator. Important safeguards are indicated throughout this guide. Pay special attention to all safety information Imminently and potentially hazardous information is highlighted by these terms:
- DANGER
Indicates an imminently hazardous situation that could result in death or serious injury to the user or operator if not avoided - WARNING
Indicates a potentially hazardous situation that could result in death or serious injury to the user or operator if not avoided - CAUTION
Indicates a potentially hazardous situation that could result in property damage, injury, or device damage if not avoided - IMPORTANT
Indicates important information you should know. - NOTE
Indicates notes, useful tips, recommendations, and information
DANGER
- Oxygen causes rapid burning. Do not smoke while your oxygen concentrator is operating, or when you are near a person utilizing oxygen therapy.
- Smoking during oxygen therapy is dangerous and is likely to result in facial burns or death. Do not allow smoking within the same room where the oxygen concentrator or any oxygen-carrying accessories are located.
- If you intend to smoke, you must always turn the oxygen concentrator off, remove the cannula and leave the room where either the cannula or mask or the oxygen concentrator is located. If unable to leave the room, you must wait 10 minutes after you have turned off the oxygen concentrator before smoking.
- Oxygen makes it easier for a fire to start and spread. Do not leave the nasal cannula or mask on bed coverings or chair cushions if the oxygen concentrator is turned on but not in use. The oxygen will make the materials flammable. Turn the oxygen concentrator off when not in use to prevent oxygen enrichment.
- Keep the oxygen concentrator and cannula at least 2 m (6.5 feet) from hot, sparking objects or naked sources of flame.
- Open flames during oxygen therapy are dangerous and are likely to result in fire or death. Do not allow open flames within 2 m (6.5 feet) of the oxygen concentrator or any oxygen-carrying accessories.
- DeVilbiss oxygen concentrators are equipped with a fire mitigating outlet fitting that prevents the propagation of fire into the unit.
WARNING
- To avoid electric shock, do not plug the concentrator into an AC outlet if the concentrator cabinet is broken. Do not remove the concentrator cabinet. The cabinet should only be removed by a qualified DeVilbiss technician. Do not apply liquid directly to the cabinet or utilize any petroleum-based solvents or cleaning agents.
- Improper use of the power cord and plugs can cause a burn, fire or other electric shock hazards. Do not use the unit if the power cord is damaged.
- Ensure the mains power cord is fully inserted into the concentrator connector (230-volt units) and the power cord plug is completely inserted into a fully functioning AC wall outlet. Failure to do so may cause an electrical safety hazard.
- The use of a fire stop device as part of the oxygen accessories is optional and is not contraindicated for use with this device. If used with the oxygen accessories, the fire stop device should be located as close to the patient as practicable.
- Locate oxygen tubing and power supply cords to prevent tripping hazards and reduce the possibility of entanglement or strangulation
- Do not lubricate fittings, connections, tubing or other accessories of the oxygen concentrator to avoid the risk of fire and burns.
- Do NOT use lubricants, oils or grease
- Before attempting any cleaning procedures, turn the unit “Off.”
- Use only water-based lotions or salves that are oxygen-compatible before and during oxygen therapy. Never use petroleum or oil-based lotions or salves to avoid the risk of fire and burns.
- Use only spare parts recommended by the manufacturer to ensure proper function and to avoid the risk of fire and burns.
- When using the Transfiller Caddy with a Transfill device, always keep the system on a flat surface. Disassemble the system prior to moving.
WARNING
- If you feel discomfort or are experiencing a medical emergency while undergoing oxygen therapy, seek medical assistance immediately to avoid harm
- Geriatric, pediatric or any other patient unable to communicate discomfort can require additional monitoring and/or a distributed alarm system to convey the information about the discomfort and/or the medical urgency to the responsible caregiver to avoid harm.
- Use of this device at an altitude above 13,123 feet (4000 meters) or above a temperature of 95°F (35°C) or greater than 93% relative humidity may affect the flow rate and the percentage of oxygen and consequently the quality of the therapy. Refer to specifications for details regarding parameters tested.
- To ensure you receive the therapeutic amount of oxygen delivery according to your medical condition, the Oxygen Concentrator must:
- be used only after one or more settings have been individually determined or prescribed for you at your specific activity levels.
- be used with the specific combination of parts and accessories that are in line with the specification of the concentrator manufacturer and that were used while your settings were determined.
- Your delivery settings of the oxygen concentrator should be periodically reassessed for the effectiveness of therapy.
- For your safety, the oxygen concentrator must be used according to the prescription determined by your physician
- Under certain circumstances, oxygen therapy can be hazardous. Seek medical advice before using an oxygen concentrator.
WARNING
MR Unsafe
- Do not bring the device or accessories into a Magnetic Resonance (MR) environment as it may cause unacceptable risk to the patient or damage to the oxygen concentrator or MR medical devices. The device and accessories have not been evaluated for safety in an MR environment.
- Do not use the device or accessories in an environment with electromagnetic equipment such as CT scanners, Diathermy, RFID and electromagnetic security systems (metal detectors) as it may cause unacceptable risk to the patient or damage to the oxygen concentrator. Some electromagnetic sources may not be apparent, if you notice any unexplained changes in the performance of this device, if it is making unusual or harsh sounds, disconnect the power cord and discontinue use. Contact your home care provider.
- This device is suitable for use in home and healthcare environments except for near active HF SURGICAL EQUIPMENT and the RF shielded room of an ME SYSTEM for magnetic resonance imaging, where the intensity of Electromagnetic DISTURBANCES is high.
- Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
- Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the oxygen concentrator, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
CAUTION
- It is very important to follow your oxygen prescription. Do not increase or decrease the flow of oxygen – consult your physician.
- The surface temperature of the exhaust vents on the bottom of the unit may exceed 105.8˚F (41˚C) under certain conditions.
- When device is used under extreme operating conditions, the temperature near the exhaust vents on the bottom of the unit may reach 145˚F (63˚C). Keep body parts a minimum of 32” (81.2 cm) away from this area.
- Use of harsh chemicals (including alcohol) is not recommended. If bactericidal cleaning is required, a non-alcohol-based product should be used to avoid inadvertent damage.
IMPORTANT
- It is recommended that the homecare provider lock the flow control knob to prevent inadvertent adjustment. A flow setting other than prescribed may affect the patient therapy.
- Do not service or clean this device while in use with a Patient
- Installation of 515LF-607 low output flow meter package will cause the low flow alarm to not work and will prevent the device from meeting the requirements of ISO-80601-2-69:2014 Section 201.13.2.101.
- The Device is classified as IP21 which means it is protected against finger access to hazardous parts and protected against vertically falling water drops.
- Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
- This device contains electrical and/or electronic equipment. Follow local governing ordinances and recycling plans regarding disposal of device components
INTRODUCTION
This instruction guide will acquaint you with your DeVilbiss oxygen concentrator. Make sure that you read and understand this guide before operating your unit. Important safeguards are indicated throughout this guide. Pay special attention to all safety information. Contact your DeVilbiss equipment provider should you have any questions.
Intended Use
The DeVilbiss 5 Liter Oxygen Concentrator intended use is to provide supplemental low flow oxygen therapy for patients suffering from COPD, cardiovascular disease, and lung disorders. The oxygen concentrator is used in home-type environments, homes, nursing homes, patient care facilities, etc.
Indications For Use
The DeVilbiss Oxygen Concentrator is intended for use as an oxygen concentrator to provide supplemental low flow oxygen therapy in the home, nursing homes, patient care facilities, etc.
Contraindications
The device is not intended to be life-supporting or life-sustaining
Essential Performance
Essential Performance of the Oxygen Concentrator is to deliver a continuous flow of oxygen-enriched gas. Visual and audible alarms indicate if the device is not meeting specifications or a failure has been detected.
Service Life
The expected service life of the 525 is 5 years of operation when used in accordance with all manufacturer guidance for safe use, maintenance, storage, handling and general operation. The expected service life of the unit, and in particular the sieve beds and compressor, may vary based on the operating environment, storage, handling and the frequency and intensity of use.
Why Your Physician Prescribed Supplemental Oxygen
Today, many people suffer from heart, lung and other respiratory diseases. Many of these people can benefit from supplemental oxygen therapy. Your body requires a steady supply of oxygen to function properly. Your physician prescribed supplemental oxygen for you because you are not getting enough oxygen from room air alone. Supplemental oxygen will increase the amount of oxygen that your body receives. Supplemental oxygen is not addictive. Your physician prescribed a specific oxygen flow to improve symptoms such as headaches, drowsiness, confusion, fatigue or increased irritability. If these symptoms persist after you begin your supplemental oxygen program, consult your physician.
The oxygen delivery setting has to be determined for each patient individually with the configuration of the equipment to be used, including accessories. The proper placement and positioning of the prongs of the nasal cannula in the nose is critical to the amount of oxygen delivered to the respiratory system of the patient. Your Delivery settings of the oxygen concentrator should be periodically reassessed for the effectiveness of therapy.
How Your DeVilbiss Oxygen Concentrator Works
Oxygen concentrators are the most reliable, efficient and convenient source of supplemental oxygen available today. The oxygen concentrator is electrically operated. The unit separates oxygen from room air which allows high-purity supplemental oxygen to be delivered to you through the oxygen outlet. Although the concentrator filters the oxygen in a room, it will not affect the normal amount of oxygen in your room.
IMPORTANT PARTS OF YOUR CONCENTRATOR
Please take time to familiarize yourself with your DeVilbiss oxygen concentrator before operating
Front View (Figure A)
- Operating instructions
- Power Switch | = ON O = OFF
- Flow meter knob
- Flow meter
- Circuit breaker – resets the unit after electrical overload shutdown
- Oxygen outlet – oxygen is dispersed through this port
- Normal Oxygen (green) light (see page 8)
- Low Oxygen (yellow) light (see page 8)
- Red Service Required (red) light – when illuminated contact your DeVilbiss provider
- Hour meter
- Handgrip
- Exhaust
CAUTION
When device is used under extreme operating conditions, the temperature near the exhaust vents on the bottom of the unit may reach 145˚F (63˚C). Keep body parts a minimum of 32” (81.2 cm) away from this area. - Power cord and/or IEC power connector
- Line cord strap
- Filter Door with venting and compartment for optional gross particle filter.
- Auxiliary Oxygen Port (Serial numbers starting with R, N, or B): Your concentrator is equipped with an auxiliary oxygen port that can be used to fill oxygen cylinders with an FDA-cleared cylinder filling device that is designed to use oxygen from a concentrator to fill a cylinder. The port is only for use with FDA-cleared filling devices with compatible oxygen input specifications. Refer to the cylinder filling device instruction guide for the oxygen input/output specifications, connection, and operating instructions.
Accessories
Transfer Caddy DeVilbiss 525DD-650 Bubble Humidifier Salter Labs 7600 or equivalent
There are many types of humidifiers, oxygen tubing and cannulas/masks that can be used with this device. Certain humidifiers and accessories may impair the device’s performance. A mask or any nasal cannula can be used with continuous flow delivery and may be sized according to your prescription as recommended by your homecare provider who should also give you advice on the proper usage, maintenance and cleaning.
WARNING
When using the Transfer Caddy with a Transfill device, always keep the system on a flat surface. Disassemble the system prior to moving.
NOTE – A maximum of 50 feet (15 meters) of crush-proof oxygen tubing, plus 7 feet (2.1 meters) of a cannula, plus a bubble humidifier is allowed between the concentrator and the patient
NOTE – Your healthcare provider should verify the compatibility of the oxygen concentrator and all of the parts used to connect to the patient before use.
SETTING UP YOUR OXYGEN CONCENTRATOR
- Position your unit near an electrical outlet in the room where you spend most of your time
NOTE – Do not connect to an electrical outlet controlled by a wall switch.
DANGER
Oxygen causes rapid burning. Do not smoke while your oxygen concentrator is operating, or when you are near a person utilizing oxygen therapy. Keep the oxygen concentrator and cannula at least 2 m (6.5 feet) from hot, sparking objects or naked sources of flame. - Position your unit on a flat surface at least 6 inches (16 cm) from walls, draperies or any other objects that might prevent the proper flow of air in and out of your oxygen concentrator. The oxygen concentrator should be located in a well-ventilated area to avoid pollutants or fumes.
NOTE – To move the unit, firmly grasp the handle located on the top of the unit, rolling and/or lifting the unit over pathway obstacles. - Before operating your unit, always check to be sure the filter door vents (located on the back of your unit) are clean. Proper cleaning is discussed in the Caring For Your Concentrator section on page 8.
- Attach the appropriate oxygen accessories to the oxygen outlet.
Oxygen Tubing Connection:
- Thread the supplied oxygen outlet connector onto the oxygen outlet
- b. Attach the oxygen tubing directly to the connector (Figure 1).
Oxygen Tubing Connection With Humidification:
If your physician has prescribed an oxygen humidifier as part of your therapy, follow these steps (If using a prefill, go to step b.):
- Fill the humidifier bottle as per the manufacturer’s instructions.
- Thread the wing nut located on the top of the humidifier bottle to the oxygen outlet so that it is suspended (Figure 2). Make sure it is securely tightened.
- Attach the oxygen tubing directly to the humidifier bottle outlet fitting (Figure 3).
NOTE – Your physician has prescribed either a nasal cannula or a face mask. In most cases, they are already attached to the oxygen tubing. If not, follow the manufacturer’s instructions for attachment.
NOTE – Your healthcare provider should verify the compatibility of the oxygen concentrator and all of the parts used to connect to the patient before use.
Remove the power cord completely from the line cord strap. Make sure the power switch is in the “Off” position and insert the plug into the wall outlet. The unit is double insulated to guard against electric shock
WARNING
Ensure the mains power cord is fully inserted into the concentrator connector (230 volt units) and the power cord plug is completely inserted into a fully functioning AC wall outlet. Failure to do so may cause an electrical safety hazard.
NOTE – (only 115-volt units) The plug on the DeVilbiss oxygen concentrator has one blade wider than the other. To reduce the risk of electric shock, this plug is intended to fit in a wall outlet only one way. Do not attempt to defeat this safety feature.
NOTE – To check your oxygen concentrator and accessories for proper operation; 1. Check the output flow by placing the end of the nasal cannula under the surface of a half-full cup of water and look for the bubbles. 2. Check the system for leaks by bending the nasal prongs over and squeezing tight to stop the flow of oxygen. Look at the flow meter to see that the indicator ball on the flow meter drops to zero. If the indicator ball does not drop to zero, check all connections for possible leaks. Parts to check for leaks are tubing connections, humidifier bottles, and other accessories like firebreaks. Repeat these steps until the flow meter ball drops to zero. Contact your provider or service supplier immediately if you encounter any problems.
WARNING
Improper use of the power cord and plugs can cause a burn, fire or other electric shock hazards. Do not use the unit if the power cord is damaged.
OPERATING YOUR DEVILBISS OXYGEN CONCENTRATOR
DANGER
Oxygen causes rapid burning. Do not smoke while your oxygen concentrator is operating, or when you are near a person utilizing oxygen therapy. Keep the oxygen concentrator and cannula at least 2 m (6.5 feet) from hot, sparking objects or naked sources of flame. DeVilbiss oxygen concentrators are equipped with a fire mitigating outlet fitting that prevents the propagation of the fire into the unit.
- Press the power switch to the “On” position. When the unit is turned “On,” all three lights (Service Required, Low Oxygen, and Normal Oxygen) on the front panel will illuminate briefly and an audible signal will briefly alarm confirming that the LEDs and audible signal are functioning properly. The unit will then operate in “start-up” mode with the Low Oxygen light lit until a normal oxygen level is achieved, at which time the Normal Oxygen light will remain lit. The “start-up” may take up to 15 minutes.
NOTE – DeVilbiss recommends for optimal service life that the DeVilbiss Oxygen Concentrator be operated for at least 30 minutes after it is powered on. Shorter periods of operation, operating in extreme temperature/humidity conditions or in the presence of contaminates, and/or handling and storage conditions outside those specified, may affect the long term reliable operation of the product.
DANGER
Oxygen makes it easier for a fire to start and spread. Do not leave the nasal cannula or mask on bed coverings or chair cushions if the oxygen concentrator is turned on but not in use, the oxygen will make the materials flammable. Turn the oxygen concentrator off when not in use to prevent oxygen enrichment.
NOTE – If the audible signal alarms but the unit is not operating, there is no power to the unit. Refer to the Minor Troubleshooting chart on page 9 and contact your DeVilbiss provider if necessary.
NOTE – If an audible low-frequency vibration sound is detected, the unit is not operating properly. Refer to the Minor Troubleshooting chart on page 9, and contact your DeVilbiss provider if necessary. - Check the flow meter to make sure that the flow meter ball is centered on the line next to the prescribed number of your flow rate.
CAUTION
It is very important to follow your oxygen prescription. Do not increase or decrease the flow of oxygen – consult your physician.
NOTE – Your DeVilbiss provider may have preset the flow meter so that it cannot be adjusted.
NOTE – If the flow meter knob is turned clockwise, the flow decreases (and eventually will shut off the oxygen flow). If the knob is turned counterclockwise, the flow increases.
NOTE – For prescriptions of 5 LPM, be sure the ball is centered on the 5-liter line. The ball should not touch the red line. Setting the flow higher than 5 may cause the oxygen purity level to drop.
NOTE – The low-flow alarm may activate if the flow meter ball is set at or below 0.2 lpm. The unit will continue to run; however, the Service Required light will come on accompanied by an audible alarm. Adjust the flow meter to your prescribed flow. - Your DeVilbiss concentrator is now ready for use, properly position the cannula with the nasal prongs facing upward, insert the prongs into nose. Wrap the cannula tubing over the ears and position in front of body (Figure 4). Allow 15 minutes for the oxygen concentrator to reach stated performance.
DeVilbiss OSD® Operation
The OSD (Oxygen Sensing Device) is a device within your concentrator that monitors the oxygen produced by your unit. The OSD lights on the top panel are defined as follows:
- Green Normal Oxygen light–acceptable oxygen level.
- Yellow Low Oxygen light– below an acceptable oxygen level.
If the oxygen purity falls below the acceptable level: The green Normal Oxygen light will shut off, the yellow Low Oxygen light will illuminate, and an intermittent audible signal will sound.
Refer to the Minor Troubleshooting section in this guide on page 9, and switch to your reserve oxygen system. Do not attempt any other maintenance. Contact your DeVilbiss provider immediately.
RESERVE OXYGEN SYSTEM
As a precaution, your DeVilbiss provider may supply you with a reserve oxygen system. If your unit loses electrical power or fails to operate correctly, the Patient Alert System will sound to signal you to switch to your reserve oxygen system (if provided) and contact your DeVilbiss provider.
CARING FOR YOUR DEVILBISS OXYGEN CONCENTRATOR|
DeVilbiss recommends using only original DeVilbiss parts and filters in order to guarantee the reliable operation of the product.
WARNING
Do NOT use lubricants, oils or grease. Before attempting any cleaning procedures, turn the unit “Off.”
Cannula/Mask, Tubing and Humidifier Bottle
Clean and replace the cannula/mask, tubing and humidifier bottle according to the manufacturer’s instructions.
Filter Door with Vents
Inspect the vents periodically, and wipe with a dry cloth as needed to remove dust
Exterior Cabinet
Clean the concentrator exterior cabinet weekly by using a damp cloth and wiping dry; the vents can also be wiped with a damp cloth
Cleaning
C | Recommended cleaning interval | Number of cleaning cycles * | Compatible cleaning method |
Outer Cabinet | 7 days | 260 | Water, use only a damp cloth |
Filter Door Vents | 7 days | 260 | Wipe with dry cloth, or a cloth dampened with water to remove dust . |
Oxygen Outlet Connector | 7 days | 104 | Mild dish soap (2 tbsp) and warm water (2 cups) |
number of cleaning cycles determined by recommended cleaning interval and expected service life
WARNING
To avoid electric shock, do not plug the concentrator into an AC outlet if the concentrator cabinet is broken. Do not remove the concentrator cabinet. The cabinet should only be removed by a qualified DeVilbiss technician. Do not apply liquid directly to the cabinet or utilize any petroleum-based solvents or cleaning agents.
CAUTION
Use of harsh chemicals (including alcohol) is not recommended. If bactericidal cleaning is required, a non-alcohol based product should be used to avoid inadvertent damage.
TROUBLESHOOTING
The following troubleshooting chart will help you analyze and correct minor oxygen concentrator malfunctions. If the suggested procedures do not help, switch to your reserve oxygen system and call your DeVilbiss homecare provider. Do not attempt any other maintenance.
WARNING
To avoid electric shock, do not plug the concentrator into an AC outlet if the concentrator cabinet is broken. Do not remove the concentrator cabinet. The cabinet should only be removed by a qualified DeVilbiss technician.
Minor Troubleshooting Chart
SYMPTOM | POSSIBLE CAUSE | REMEDY |
A . Unit does not operate . All lights are off when the power switch is “On .” Audible alert is pulsing . | 1 . Power cord not properly inserted into wall outlet . | 1 . Check power cord connection at the wall outlet . On 230 volt units, also check the mains connection on the back of the unit . |
2 . No power at wall outlet . | 2 . Check your home circuit breaker and reset if necessary . Use a different wall outlet if the situation occurs again . | |
3 . Oxygen concentrator circuit breaker activated . | 3 . Press the concentrator circuit breaker reset button located below the power switch . Use a different wall outlet if the situation occurs again .
If the above remedies do not work, contact your DeVilbiss provider . | |
B . Unit operates . Red Service Required light is illuminated . Audible alert may be sounding . | 1 . Filter door vents are blocked . | 1. Check filter door vents and ensure that the openings are not blocked. |
2 . Exhaust is blocked . | 2 . Check the exhaust area and make sure there is nothing restricting the unit exhaust . | |
3 . Blocked or defective cannula, face mask, or oxygen tubing . | 3. Detach cannula or face mask. If proper flow is restored, clean or replace if necessary . Disconnect the oxygen tubing at the oxygen outlet . If proper flow is restored, check oxygen tubing for obstructions or kinks. Replace if necessary . | |
4. Blocked or defective humidifier bottle . | 4. Detach the humidifier from the oxygen outlet. If proper flow is obtained, clean or replace humidifier. | |
5 . Flow meter set too low . | 5. Set flow meter to prescribed flow rate.
If the above remedies do not work, contact your DeVilbiss provider . | |
C . Unit operates . Audible low-frequency vibration sound is detected . | 1 . Electronic Assembly Malfunction . | 1 . Turn your unit “Off .” Switch to your reserve oxygen system and contact your DeVilbiss provider immediately . |
D . Yellow Low Oxygen light is on . | 1 . Unit in “start up” mode . | 1 . Allow unit up to 15 minutes to complete start up period . |
E . The yellow Low Oxygen light is on and the intermittent audible signal is sounding . | 1 . Flow meter is not properly set . | 1. Ensure the flow meter is properly set to the prescribed number. (The maximum flow meter setting is 3 LPM when an oxygen bottle is being filled with oxygen from the auxiliary port.) |
2 . Filter door vents are blocked . | 2. Check filter door vents and ensure that the openings are not blocked. | |
3 . Exhaust is blocked. | 3 . Check the exhaust area and make sure there is nothing restricting the unit exhaust.
If the above remedies do not work, contact your DeVilbiss provider. | |
F Red Service Required light is on and an intermittent audible signal is sounding. | 1 . Flowmeter is not properly set . | 1. Ensure the flow meter is properly set to the prescribed number. (The maximum flow meter setting is 3 LPM when an oxygen bottle is being filled with oxygen from the auxiliary port.) |
2 . Filter door vents are blocked. | 2. Check filter door vents and ensure that the openings are not blocked. | |
3 . Exhaust is blocked. | 3 . Check the exhaust area and make sure there is nothing restricting the unit exhaust.
If the above remedies do not work, contact your DeVilbiss provider. | |
4. Electronic Assembly Malfunction. | 4. Turn your unit “Off.” Switch to your reserve oxygen system and contact your DeVilbiss provider immediately. | |
G . If any other problems occur with your oxygen concentrator. | 1 . Turn your unit “Off .” Switch to your reserve oxygen system and contact your DeVilbiss provider immediately. |
OVERVIEW OF ALARMS
This device contains an alarm system that monitors the state of the device and alerts of abnormal operation, loss of essential performance or failures. Alarm conditions are shown on the LED display. The alarm system functions are tested at power-up by lighting all visual alarm indicators and sounding the audible alarm (beep). All alarms are Low Priority Technical Alarms
Alarm Condition | LED Indicator | Visual Alarm Signal
Meaning | Audible Alarm Signal | Visual Alarm Signal
Cleared by | Action to take |
Start-up Period | YELLOW Low O2 LED ON | No | After startup period, O2 ≥82% | Wait for unit to finish startup period, up to 15 minutes | |
Low Oxygen Concentration at <82% | YELLOW Low O2 LED ON | Yes | Turn unit off | See Troubleshooting table | |
Malfunction | RED
Service Required LED ON |
Yes |
Turn unit off | Return unit to provider for service |
RETURN AND DISPOSAL
This device may not be disposed of with household waste. After use of the device, please return the device to the provider for disposal. This device contains electrical and/or electronic components that must be recycled per EU Directive 2012/19/EU-Waste Electrical and Electronic Equipment (WEEE). Non-infectious used accessories (e.g. nasal cannula) can be disposed of as residential waste. The disposal of infectious accessories (e.g. nasal cannula from an infected user) must be made via an approved waste disposal company. Names and addresses can be obtained from the local municipality.
PROVIDER’S NOTES – Cleaning and Disinfection When There is a Patient Change
DeVilbiss Healthcare recommends that at least the following procedures be carried out by the manufacturer or a qualified third party between uses by different patients.
NOTE – If the following described complete processing of the concentrator by an appropriately trained individual is not possible, the device should not be used by another patient.
NOTE – If preventive maintenance is due at this time, these procedures should be carried out in addition to the servicing procedures.
- Use disinfectants safely Always read the label and product information before use
- Always wear personal protective equipment when performing this procedure Use suitable gloves and safety glasses Cover exposed skin on arms to prevent accidental contact with a bleach solution that has been applied to the concentrator
- Dispose of all accessories that are not suitable for reuse This includes but may not be limited to the oxygen tubing, tubing connectors, nasal cannula and/or mask, oxygen outlet connector, and humidifier bottle.
- Clean the exterior of the concentrator with a clean lint-free cloth. Heavy soil should be removed with a clean lint-free cloth dampened with water. A soft-bristled brush dampened with water can be used to remove stubborn soil Dry the concentrator using a clean lint-free cloth if water was used to remove soil
- Use 5.25% chlorine bleach (Clorox Regular Liquid Bleach or equivalent). Mix one (1) part bleach with four (4) parts water in an appropriate clean container. This ratio produces a one (1) part bleach to five (5) total parts solution (1:5). The total volume (amount) of solution required is determined by the number of concentrators in need of disinfection NOTE – An alternate suitable disinfecting agent (e.g. Mikrobac® forte or Terralin® Protect) may also be used. Follow disinfectant manufacturer’s instructions.
- Apply the bleach solution in an even manner to the cabinet and power cord using a clean lint-free cloth The cloth should be dampened only and not dripping of solution Do not use a spray bottle to apply the solution Do not saturate the device with the solution Take care that no solution enters the vent areas on the concentrator base or the Auxiliary O2 fitting area on the back of the unit. Avoid over-saturating the cabinet seams so that no solution residue builds up in these areas. Avoid the caster wells located on the bottom of the unit
- The exposure time of the disinfectant solution should be 10 minutes minimum to 15 minutes maximum
- After the recommended exposure time, all surfaces of the concentrator should be wiped with a clean lint-free cloth dampened with drinking quality water no warmer than room temperature. Dry the unit with a dry, clean lint-free cloth. This is to remove residue that may stain or leave a film on the unit, especially after repeated disinfection.
- Check the cord, the plug on the back of the device, the power switch, the fuse holder, and the indicator lights for possible damage Replace all damaged or worn components
- Check the oxygen concentration. If the device is within specification, the extended life intake bacteria filter does not need to be replaced between patients. If the oxygen
- concentration is not within specification, the provider should refer to the service manual section on Troubleshooting.
NOTE – There is no portion of the gas pathways through the concentrator that could be contaminated with body fluids under normal conditions.
The device patient connection may unintentionally become contaminated with expired gases for a single fault condition i.e., a hose internal to the device becomes disconnected. This condition will cause no flow out of the device and/or an alarm condition. Should this occur, refer to the service manual for additional instructions.
Disinfection
NOTE – The disinfection process can only be completed by the manufacturer or by an appropriately trained individual.
Recommended disinfection
interval | Number of disinfection cycles | Compatible disinfection method | |
Cabinet, power cord | Between patients | 20 | 1:5 chlorine bleach (5.25%) and water solution, Mikrobac forte, Terralin Protect |
Oxygen tubing, tubing connectors, nasal cannula/mask, oxygen outlet connector, humidifier bottle | Do not clean, replace between patients | N/A | N/A |
SPECIFICATIONS
DEVILBISS 5-LITER SERIES | |||||
Catalog Number | 525DS, 525DS-Q | 525KS, 525KS-LT | 525PS | ||
Delivery Rate (Lower delivery rates available for low flow applications) | 0 .5 to 5 LPM | 0 .5 to 5 LPM | 0 .5 to 5 LPM | ||
Maximum Recommended Flow (@ nominal outlet pressures of zero & 7 kPa)** | 5 LPM | 5 LPM | 5 LPM | ||
Outlet Pressure | 8.5 ± 0.5 psig (58.6 ± 3.5 kPa) | 8.5 ± 0.5 psig (58.6 ± 3.5 kPa) | 8.5 ± 0.5 psig (58.6 ± 3.5 kPa) | ||
Auxiliary Oxygen Port ** | Outlet Pressure: <15 psi Outlet Flow: 2 LPM | Outlet Pressure: <15 psi Outlet Flow: 2 LPM | Outlet Pressure: <15 psi Outlet Flow: 2 LPM | ||
Electrical Rating | 115 V, 60 Hz, 3 .3 Amp | 220-230 V~, 50 Hz, 1 .55 Amp
230 V~, 60 Hz, 1 .9 Amp | 220-230 V~, 60 Hz, 1 .68 Amp | ||
Operating Voltage Range | 97-127 V~, 60 Hz | 187-253 V~, 50 Hz
195-253 V~, 60 Hz | 187-253 V~, 60 Hz | ||
Oxygen Percentage | 1-5 LPM=87%-96% | 1-5 LPM=93%±3% | 1-5 LPM=93%±3% | ||
Operating Altitude | |||||
(tested at 70˚F {21˚C} only) 0-1500 M (0-4921 ft) | Across the voltage range: No degradation of performance | Across the voltage range: No degradation of performance | Across the voltage range: No degradation of performance | ||
1500-4000 M (4921-13123 ft) | Tested at nominal voltage only: No degradation of performance | Tested at 230V/50Hz only: No degradation of performance | Tested at 230V/60Hz only: No degradation of performance | ||
Operating Environment Range* | |||||
41˚F (5˚C) to 95˚F (35˚C), humidity range of 10% to 95% | No degradation in performance across the operating voltage range . | No degradation in performance across the operating voltage range . | No degradation in performance across the operating voltage range . | ||
Power Consumption |
310 Watts Average 275 Watts @ 1.2 LPM & below | 230V / 50 Hz – 312 Watts Average 230V / 50 Hz – 296 Watts Average @ 1 .2 LPM & below
230V / 60 Hz – 387 Watts Average 230V / 60 Hz – 369 Watts Average @ 1 .2 LPM & below |
230V / 60 Hz – 334 Watts Average 230V / 60 Hz – 297 Watts Average @ 1 .2 LPM & below | ||
Weight | 36 lbs. (16.3 Kilograms) | 36 lbs. (16.3 Kilograms) | 36 lbs. (16.3 Kilograms) | ||
Safe Working Load | 53 lbs. (24 Kilograms) | 53 lbs. (24 Kilograms) | 53 lbs. (24 Kilograms) | ||
Sound Pressure Level at 3 and 5 LPM
(ISO 80601-2-69) | 525DS
50.9 dBA @ 3 LPM 50.7 dBA @ 5 LPM | 525DS-Q
46.7 dBA @ 3 LPM 46.7 dBA @ 5 LPM | 525KS
47.9 dBA @ 3 LPM 47.9 dBA @ 5 LPM | 525KS-LT
49.6 dBA @ 3 LPM 49.4 dBA @ 5 LPM | 45.4 dBA @ 3 LPM
45.3 dBA @ 5 LPM |
Sound Power Level at 3 and 5 LPM
(ISO 80601-2-69) | 525DS
54.7 dBA @ 3 LPM 54.5 dBA @ 5 LPM | 525DS-Q
50.4 dBA @ 3 LPM 50.4 dBA @ 5 LPM | 525KS
51.6 dBA @ 3 LPM 51.7 dBA @ 5 LPM | 525KS-LT
53.4 dBA @ 3 LPM 53.2 dBA @ 5 LPM | 49.2 dBA @ 3 LPM
49.1 dBA @ 5 LPM |
Sound Level (ISO 8359:1996) | 48 dBA (525DS)
46 dBA (525DS-Q) | 40 dBA (50 Hz (525KS)
48 dBA (50 Hz (525KS-LT) | — | ||
Alarm Sound Level | > = 62 dBA | > = 62 dBA | > = 62 dBA | ||
Dimensions | 24.5”H x 13.5”W x 12”D (62.2 x 34.2 x 30.4 cm) | 24.5”H x 13.5”W x 12”D (62.2 x 34.2 x 30.4 cm) | 24.5”H x 13.5”W x 12”D (62.2 x 34.2 x 30.4 cm) | ||
Maximum Limited Pressure | Normal Condition: 9 PSIG (62.0 kPa), Single Fault Condition: 27.6 PSIG (190.3 kPa) | Normal Condition: 9 PSIG (62.0 kPa), Single Fault Condition: 27.6 PSIG (190.3 kPa) | Normal Condition: 9 PSIG (62.0 kPa), Single Fault Condition: 27.6 PSIG (190.3 kPa) | ||
Operating System | Time Cycle / Pressure Swing | Time Cycle / Pressure Swing | Time Cycle / Pressure Swing | ||
Low Oxygen Indicator | <82% low oxygen
<60% very low oxygen | <82% low oxygen
<60% very low oxygen | <82% low oxygen
<60% very low oxygen | ||
Storage Conditions | -13˚F (-25˚C) to 158˚F (70˚C), humidity range of 15% to 93% non-condensing | -13˚F (-25˚C) to 158˚F (70˚C), humidity range of 15% to 93% non-condensing | -13˚F (-25˚C) to 158˚F (70˚C), humidity range of 15% to 93% non-condensing | ||
Equipment Class and Type | Class II Equipment Double Insulated; Type B Applied Part, IP21 | Class II Equipment Double Insulated; Type B Applied Part, IP21 | Class II Equipment Double Insulated; Type B Applied Part, IP21 | ||
Approval Body and Safety Standard | TUV
ANSI/AAMI ES60601-1:2005+A2 (R2012) +A1 IEC 60601-1-6:2010 IEC 60601-1-11:2015 ISO 80601-2-69:2014 *** CAN/CSA-C22.2 No. 60601-1:14 CAN/CSA-C22 .2 No . 60601-1-6:11 CAN/CSA-C22 .2 No . 60601-1-11:15 CAN/CSA-C22 .2 No . 80601-2-69:16 |
TUV approved for 50 Hz only to IEC 60601-1:2012 IEC 60601-1-6:2010+A1 IEC 60601-1-11:2015 EN ISO 80601-2-69:2014 |
TUV IEC 60601-1:2012 IEC 60601-1-6:2010+A1 IEC 60601-1-11:2015 EN ISO 80601-2-69:2014 | ||
CE mark | No | Yes | Yes | ||
EMC Compliance To | EN60601-1-2 | EN60601-1-2 | EN60601-1-2 |
NOTE – The OSD performance at 41˚F (5˚C) to 95˚F (35˚C), 93% R.H. through voltage range on the 525DS verified at 670m.
The maximum recommended flow is 3 LPM when an oxygen bottle is being filled with oxygen from the auxiliary oxygen port.
Use of the 515LF-607 low output flow meter package or other low output flow meter accessory will prevent the device from meeting the requirements of ISO-80601-2-69:2014 Section 201.13.2.101.
Specifications are subject to change without notice.
Oxygen Concentration vs Flow Rate (Across the listed voltage and environmental conditions.)
525DS, 525DS-Q | |
Flow L/m | %O2 |
5 | 87% – 96% |
4 | 87% – 96% |
3 | 87% – 96% |
2 | 87% – 96% |
1 | 87% – 96% |
.5 | 87% – 96% |
525KS, 525KS-LT, 525PS | |
Flow L/m | %O2 |
5 | 90% – 96% |
4 | 90% – 96% |
3 | 90% – 96% |
2 | 90% – 96% |
1 | 90% – 96% |
.5 | 90% – 96% |
ELECTROMAGNETIC COMPATIBILITY INFORMATION
WARNING
- MR Unsafe
Do not bring the device or accessories into a Magnetic Resonance (MR) environment as it may cause unacceptable risk to the patient or damage to the oxygen concentrator or MR medical devices. The device and accessories have not been evaluated for safety in an MR environment. Do not use the device or accessories in an environment with electromagnetic equipment such as CT scanners, Diathermy, RFID and electromagnetic security systems (metal detectors) as it may cause unacceptable risk to the patient or damage to the oxygen concentrator. Some electromagnetic sources may not be apparent, if you notice any unexplained changes in the performance of this device, if it is making unusual or harsh sounds, disconnect the power cord and discontinue use. Contact your home care provider.
This device is suitable for use in home and healthcare environments except for near active HF SURGICAL EQUIPMENT and the RF shielded room of an ME SYSTEM for magnetic resonance imaging, where the intensity of Electromagnetic DISTURBANCES is high.
WARNING
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally. Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the oxygen concentrator, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
WARRANTY
DeVilbiss Healthcare warrants the DeVilbiss 5 Liter Oxygen Concentrator under the conditions and limitations stated below. DeVilbiss warrants this equipment to be free from defects in workmanship and materials for three (3) years from date of factory shipment to the original purchaser, (typically the healthcare provider) unless contractually specified otherwise. This warranty is limited to the Buyer of new equipment purchased directly from Drive DeVilbiss, or one of its Providers, Distributors, or Agents. DeVilbiss’ obligation under this warranty is limited to product repair (parts and labor) at its factory or at an Authorized Service Center. Routine maintenance items, such as filters, are not covered under this warranty, nor does it cover normal wear and tear.
Warranty Claims Submissions
The original purchaser must submit any warranty claim to Drive DeVilbiss or to an Authorized Service Center. Upon verification of the warranty status, instructions will be issued. For all returns, the original purchaser must (1) properly package the unit in a DeVilbiss approved shipping container, (2) properly identify the claim with the Return Authorization Number, and (3) send the shipment freight prepaid. Service under this warranty must be performed by DeVilbiss and/or an Authorized Service Center.
NOTE – This warranty does not obligate DeVilbiss to provide a loaner unit during the time that an oxygen concentrator is undergoing repair.
NOTE – Replacement components are warranted for the unexpired portion of the original Limited Warranty. This warranty shall be voided, and DeVilbiss shall be relieved of any obligation or liability if:
- The device has been misused, abused, tampered with, or used improperly during this period
- Malfunction results from inadequate cleaning or failure to follow the instructions
- The equipment is operated or maintained outside the parameters indicated in the DeVilbiss operating and service instructions
- Unqualified service personnel conduct routine maintenance or servicing.
- Unauthorized parts or components (i.e., regenerated sieve material) are used to repair or alter the equipment.
- Unapproved filters are used with the unit.
THERE IS NO OTHER EXPRESS WARRANTY IMPLIED WARRANTIES, INCLUDING THOSE OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, ARE LIMITED TO THE DURATION OF THE EXPRESS LIMITED WARRANTY AND TO THE EXTENT PERMITTED BY LAW ANY AND ALL IMPLIED WARRANTIES ARE EXCLUDED THIS IS THE EXCLUSIVE REMEDY AND LIABILITY FOR CONSEQUENTIAL AND INCIDENTAL DAMAGES UNDER ANY AND ALL WARRANTIES ARE EXCLUDED TO THE EXTENT EXCLUSION IS PERMITTED BY LAW SOME STATES DO NOT ALLOW LIMITATIONS ON HOW LONG AN IMPLIED WARRANTY LASTS, OR THE LIMITATION OR EXCLUSION OF CONSEQUENTIAL OR INCIDENTAL DAMAGES, SO THE ABOVE LIMITATION OR EXCLUSION MAY NOT APPLY TO YOU
This warranty gives you specific legal rights, and you may also have other rights that vary from state to state.
NOTE – International warranties may vary.
ORDERING AND RETURNING PARTS
DeVilbiss Customer Service Contact Information
Customer Service (USA): 800-338-1988
International Department: 814-443-4881 / [email protected]
Ordering Non-Warranty Replacement Parts
Order non-warranty parts and literature from your DeVilbiss provider.
DOWNLOAD RESOURCES
- DeVilbiss 525DS 5-Liter Oxygen Concentrator [pdf] Instructions 525DS 5-Liter Oxygen Concentrator, 525DS, 5-Liter Oxygen Concentrator
- Read more: https://manualsee.com/devilbiss/525ds-5-liter-oxygen-concentrator-manual#ixzz7e8M975g8
FAQ’S
The DeVilbiss 525DS 5-Liter Oxygen Concentrator weighs about 28 pounds.
The DeVilbiss 525DS 5-Liter Oxygen Concentrator costs about $2,000.00 to $2,500.00.
The DeVilbiss 525DS 5-Liter Oxygen Concentrator has a one year warranty on parts and labor.
The wattage a stationary oxygen concentrator will use depends on its size, and the setting it’s on. For example, the small Inogen 5L Concentrator can use as much as 275 watts (2 LPM) while the Respironics Millenium 10 can use an average of 600 watts. Stationary models fall somewhere in between.
Remove the filter from the back of the device. Clean the filter with a mild detergent and water. Rinse in water and dry before reuse. You may notice if you do not clean your filters regularly that there may be reduced airflow, and you may also not be receiving the prescribed oxygen you need.
Home oxygen concentrators are like any machine, and they require regular maintenance and service to ensure they operate and perform correctly. Luckily, home units (concentrators that run only on AC/house power) are easy to maintain.
The general rule is to clean the oxygen concentrator’s inlet filter weekly, but you should check the unit’s manual for the manufacturer’s recommendations. You may want to give the filter an extra wash if you notice any buildup or have recently been to an especially dusty place. Locate and remove the filter.
Battery Life
You can expect portable oxygen concentrators to run on a single battery charge for up to four or five hours. The smaller, lighter units generally have shorter battery charge-spans, though many of the devices come with extra batteries that can double that off-the-grid lifespan and charging options in cars.
If you are using a humidifier bot- tle (optional for >4LPM), disinfect it with a solution of 3 parts water and 1 part white vinegar at least twice a week to prevent infection. Use distilled water in the humidifier bottle and change the water daily.
The ability of oxygen concentrators to continuously draw in and treat air ensures that, unlike oxygen tanks, the concentrator will never run out of oxygen.
Some people use oxygen all day and night, continuously. People who need to use continuous oxygen should use it for at least 15 hours or more a day. Some people only need oxygen at night, while they sleep (nocturnal oxygen).
Only fill to the maximum level marked on the bottle. Filling above this level may cause water to be pushed into the concentrator, causing damage to the unit. It is recommended to use distilled water wherever possible as this prevents potential mineral build up occurring if using tap water.
VIDEO
www://drivedevilbiss.co.uk/
Documents / Resources
DeVilbiss 525DS 5-Liter Oxygen Concentrator [pdf] Instructions 525DS 5-Liter Oxygen Concentrator, 525DS, 5-Liter Oxygen Concentrator |