Arrow Eu-25552-hp Pressure Injectable Two-lumen Picc Instruction Manual

Arrow Eu-25552-hp Pressure Injectable Two-lumen Picc Instruction Manual

ARROW logoEU-25552-HP Pressure Injectable Two-Lumen PICC
Instruction Manual

ARROW EU 25552 HP Pressure Injectable Two Lumen PICC - icon

Pressure Injectable Two-Lumen PICC

Contents:
1: Two Lumen Catheter: 5 Fr. (1.81 mm OD) x 55 cm, Pressure Injectable, TPort Connector, Blue FlexTip®, Contamination Guard and Placement Wire
1: GlideThruTM Peel-Away Sheath: 5 Fr. x 2-3/4″ (7 cm) Radiopaque over 5 Fr. Dilator
1: Spring-Wire Guide, Nitinol, Marked: .018″ (0.46 mm) dia. x 17-3/4″ (45 cm) (Straight Soft Tip on One End – Straight Stiff Tip on Other) with Arrow Advancer
1: Introducer Needle: Echogenic 21 Ga. x 2-3/4″ (7 cm) TW 1: Syringe: 10 mL Luer-Lock
1: SecondSiteTM Adjustable Hub: Fastener
1: SecondSiteTM Adjustable Hub: Catheter Clamp
1: SharpsAway® II Locking Disposal Cup
1: Catheter Trimmer
1: Safety Scalpel: #11
1: Patient ID Card
1: Patient Information Booklet
1: Chart Sticker
1: Paper Tape Measure
1: Tourniquet
1: Dressing: GRIP-LOK®1 Catheter Fixation Device
2: Valve: MicroCLAVE®2 Neutral Displacement Connector
1A registered trademark of TIDI.
2A registered trademark of ICU Medical, Inc.
All components are CE 2797 unless otherwise noted.
Warning:
Read all package insert warnings, precautions, and instructions prior to use. Failure to do so may result in severe patient injury or death. www.teleflex.com/IFU Not made with natural rubber latex. Fluid path components are non-pyrogenic.
Contraindications: The Pressure Injectable PICC is contraindicated wherever there is the presence of device-related infections or the presence of thrombosis in the intended insertion vessel or catheter pathway. A clinical assessment of the patient must be completed to ensure no contraindications exist.

ARROW EU 25552 HP Pressure Injectable Two Lumen PICC - icon 1LumenPriming Volume*
(mL)
Gravity Flow Rate
(mL/hr)
Pump Flow Rate
(mL/hr)
MAX Pressure Injection
Flow Rate** (mL/sec)
Distal (18 Ga.)
Proximal (18 Ga.)
0.48
0.51
356
356
2934
3317
4
4

* Priming volumes are approximate and are done without accessories. Flow rate values are approximate and are determined using deionized water at 100 cm head height. Pump flow rates are determined at a maximum pump pressure of 10 PSIG and represent approximate flow capabilities.
** Pressure injection flow rates are determined at the injector pressure setting of 300 psi maximum using media of 11.8 centipoise viscosity, with 152 cm pressure tubing.
ARROW EU 25552 HP Pressure Injectable Two Lumen PICC - icon 2ARROW EU 25552 HP Pressure Injectable Two Lumen PICC - icon 3ARROW EU 25552 HP Pressure Injectable Two Lumen PICC - icon 4Pressure  Injectable  Two-Lumen PICC

ARROW EU 25552 HP Pressure Injectable Two Lumen PICC - icon 5 EU Authorized Representative and Importer:
KEISER AIR250 Seated Leg Curl Machine - Icon 5  Teleflex Medical IDA Business and
CE SYMBOLTechnology Park Dublin Road, Athlone
2797 Co. Westmeath, Ireland
ICONArrow International LLC
A subsidiary of Teleflex Incorporated
3015 Carrington Mill Blvd.
Morrisville, NC 27560 USA
ARROW EU 25552 HP Pressure Injectable Two Lumen PICC - logo 1
www.teleflex.com
LBL056484 R00 (2021-06)ARROW logo