Mini-7KS Mini Combination Rotor Centrifuge
Product Information
Mini Centrifuge Mini-7KS
Manual
Hangzhou Yooning Instrument Co., Ltd. Address 4 floor, #6
Building, No.3 Shengang Road, Yuhang District, Hangzhou, China
Customer Service: +86-571-86907383 / 87613616
Website: www.yooning.com
General Information
Thank you for purchasing our Product. This Manual for users
contains function and operation of the Instrument. In order to use
the instrument properly, please read this manual carefully before
operation.
Please check the instrument and accessories with the packing
list at the first time you open the instrument packing case.
Packing List
| No. | Item | Model/Type | Quantity | Confirm |
|---|---|---|---|---|
| 1 | Main Device | Mini-7KS | 1 | |
| 2 | Allen Wrench | 1 | ||
| 3 | Fuse | 2 | ||
| 4 | QC PASS Card | 1 | ||
| 5 | Manual | 1 |
Contents
Combination Rotors Centrifuge
1. Product Introduce
Centrifuge-7KS is a micro centrifuge with Composite Rotor, users
can complete different experiments without replacing the rotor to
save time. The centrifuge can open the transparent cover to suspend
operation, close the cover and continue to run in the original
time, and has an electronic timing function. It is designed with a
three-in-one composite rotor, which is suitable for:
- 6×1.5/2.0ml centrifuge tube
- 6×0.5ml centrifuge tube
- 2x8x0.2ml PCR tube strip
- 16×0.2ml centrifuge tube
1.1 Structure Description
- Transparent Cover
- Composite Rotor
- Cover Release button
- Face Panel
Performance Test
| No. | Content | Result |
|---|---|---|
| 1 | Speed | Qualified |
| 2 | Basic Function | Qualified |
| 3 | Appearance | Qualified |
| 4 | Label | Qualified |
| 5 | Continuous Running Test | Qualified (30min without problem) |
Troubles and Shootings
| Troubles | Causes | Shootings |
|---|---|---|
| The rotor doesn’t work when it powers on |
|
|
| High noise, abnormal noise, and strong vibration |
|
|
| Broken rotor cover | Attention to maintenance |
Product Usage Instructions
2. The Parameters
| Model | Speed | Max. RCF | Capacity | Timer Working | Noise | Dimension | Weight | Voltage |
|---|---|---|---|---|---|---|---|---|
| Mini-7KS | 4000rpm, 6000rpm & 7000rpm | 960g (4000rpm), 2000g (6000rpm), 2900g (7000rpm) | 6×1.5/2.0ml centrifuge tube, 6×0.5ml centrifuge tube, 2x8x0.2ml PCR tube strip, 16×0.2ml centrifuge tube | 1s~99m59s | 45 dB | 1.6kg | AC100~110V/220~240V 50~60HZ |
3. Safety Instruction
- Don’t run instrument without rotor.
- Only add stuff into tubes before loading into rotor.
- Don’t remove instrument during working.
- Place instrument on a flat and firm benchtop.
4. Operation Guide
Press the up and down key to set the speed and the time and
press for longer to increase the setting value. The adjustable
speed is 4000/6000/7000 rpm and the max. time is 99:59.
Mini Centrifuge Mini-7KS
Manual
Hangzhou Yooning Instrument Co., Ltd. Address 4 floor, #6 Building, No.3 Shengang Road,
Yuhang District, Hangzhou, China Customer Service+86-571-86907383 / 87613616 Website: www.yooning.com
Hangzhou Yooning Instrument Co., Ltd.
General Information
Thank you for purchasing our Product. This Manual for users contains function and operation of the Instrument. In order to use the instrument properly, please read this manual carefully before operation.
Please check the instrument and accessories with the packing list at the first time you open the instrument packing case.
Packing List
No.
Item
Model/Type
Quantity
Confirm
1
Main Device
2
Allen Wrench
3
Fuse
4
QC PASS Card
5
Manual
QC:
Mini-7KS
1 1 2 1 1 Date:
Contents
1. Product Introduce································································1 1.1 Structure Description·····················································1
2. The Parameters····································································2 3. Safety Instruction·································································3 4. Operation Guide···································································3
4.1 Face Panel·······································································3 4.2 Functions of buttons·······················································3 4.3 Time and Speed·······························································3 4.4 Operation and Stop·························································4 5. Clean·····················································································4 6. Troubles and Shootings························································5 Performance Test·····································································6 Packing List ··············································································7
Combination Rotors Centrifuge
1. Product Introduce
Centrifuge-7KS is a micro centrifuge with Composite Rotor, users can complete different experiments without replacing the rotor to save the time. The centrifuge can open the transparent cover to suspend operation, close the cover and continue to run in the original time, and has electronic timing function. It is designed with “three in one” composite rotor, which is suitable for 6×1.5/2.0ml centrifuge tube 6×0.5ml centrifuge tube 2x8x0.2ml PCR tube strip 16×0.2ml centrifuge tube
1.1 Structure Description
Transparent Cover
Composite Rotor
Cover Release button
Face Panel
1
Performance Test
Item
Mini Centrifuge
Model
Date
No.
Content
1
Speed
2
Basic Function
3
Appearance
4
Label
5
Continuous Running Test
Methods Tachometer Visual check Visual check Visual check Visual check
Serial No.
Standards 4000/6000/ 7000±500rpm
Valid
Pass Pass 30min without problem
Result
Remarks:
Mini-7KS
Results Qualified Qualified Qualified Qualified Qualified
QC:
Confirmer:
6
6. Troubles and Shootings
Troubles
Causes
Shootings
The rotor doesn’t work when it powers on
High noise, abnormal noise and strong vibration
1. No power
2. The cover not closed
3. Fuse burned
1. The instrument not placed horizontally 2. The rotor not properly installed 3. Tubes not symmetrically distributed
4. Broken rotor
5. Different type tubes
Check the power, switch and power cord Close the cover again Exchange fuse/ Contact to the seller Place the instrument is horizontally
Installed the rotor properly
Distribute tubes symmetrically
Exchange the new rotor
Use the same type tubes
6. Broken rotor cover
Attention to maintenance
2. The Parameters
Model Speed
Max. RCF
Capacity
Timer Working
Noise Dimension
Weight Voltage
Mini-7KS
4000rpm,6000rpm&7000rpm
960g (4000rpm) 2000g (6000rpm) 2900g (7000rpm) 61.5ml/2.0ml centrifuge tube 60.5ml centrifuge tube 280.2ml PCR Tube Strips 160.2ml centrifuge tube
1s~99m59s
45 dB
160×176×135mm
1.6kg
AC100~ 110V/220~240V 50~60HZ
5
2
3. Safety Instruction
1) Don’t run instrument without rotor. 2) Only add stuff into tubes before loading into rotor. 3) Don’t remove instrument during working. 4) Place instrument on a flat and firm benchtop.
4 Operation Guide
4.1 Face Panel and buttons
4.2 Functions of buttons Press the up and down key to set the speed and the time and press for longer to increase setting value. The adjustable speed is 4000/6000/7000 rpm and the max. time is 99:59.
3
4.3. Time and Speed
1) Once Instrument turned on, the Instrument goes into the initial status with the sound of “de…”
2) About 1s later, the figure 06 is speed 6000rpm; 02:30 in the time display is the last set time.
3)
Press up and down key to set speed
4)
Press up and down key to set timer
5) When 00:00 is displayed , the time is unlimited.
4.4 Operation and Stop
1. Before turning on the power switch, check up the switch position , make it at ” 0″ , then connect instrument to power supply. 2. Turn on instrument, it will start initiate with alarm “de…” , open the lid, and load tubes into rotor. Set time according to steps 4.3, press down the lid to run the instrument. Rotor will stop once timer count down to 0. Only open the lid after the rotor stops by press down the lid button. If user needs to stop instrument during working, turn off the power switch.
Warning: 1) Do not open the lid before the rotor stops. 2) Centrifugal tubes should be placed symmetrically into rotor,
every tube contains stuff less than 3.0g 3) Stop the instrument in case of abnormal noisy 4) Don’t use multi tubes of 1.5ml0.5ml0.2ml in rotor 8×2ml.
5. Clean
Clean the instrument with a dry cloth and a non-corrosive cleaning solution with a pH of 7 ± 1. Do not immerse the instrument in a liquid or use a liquid to wet the instrument. Ensure that all parts of the instrument are completely dry before running.